GC COLUMNS | APPLICATION-SPECIFIC COLUMNS<br />

Petroleum & Petrochemical<br />

Aromatics & Oxygenates in Gasoline Analysis<br />

Rt®-TCEP Columns (fused silica)<br />

(highly polar phase; 1,2,3-tris[2-cyanoethoxy]propane—not bonded)<br />

• General purpose <strong>columns</strong>, ideal for aromatics and oxygenates in gasoline.<br />

• Temperature range: 0 °C to 135 °C.<br />

Most gasolines contain aliphatic hydrocarbons up to n-dodecane (C12). To improve<br />

identification of the aromatics and oxygenates, it is desirable to elute benzene after C11<br />

and toluene after C12. The extremely polar Rt®-TCEP stationary phase provides a<br />

retention index for benzene greater than 1100 and permits the separation of alcohols<br />

and aromatics from the aliphatic constituents in gasoline.<br />

Rt®-TCEP <strong>columns</strong> have the same high polarity as TCEP packed <strong>columns</strong> (pre<strong>columns</strong><br />

in ASTM Method D4815 for the analysis of petroleum oxygenates), with the efficiency<br />

of a capillary column. The result is a column that can separate a wide variety of compounds<br />

with an elution pattern unattainable using other high polarity siloxanes.<br />

similar phases<br />

SPB-TCEP, CP-TCEP<br />

The Rt®-TCEP column incorporates a nonbonded stationary phase coated on a surface<br />

specialized for enhanced polymer stability and extended column lifetime. Solvent rinsing<br />

should be avoided. Conditioning is necessary only if the column is to be used at<br />

temperatures near the maximum operating temperature.<br />

ID df temp. limits 30-Meter 60-Meter<br />

0.25mm 0.40µm 0 to 135°C 10998 10999<br />

Petroleum oxygenates on an Rt®-TCEP column.<br />

1<br />

2<br />

1. methyl tert-butyl ether<br />

2. n-undecane<br />

3. tert-butanol<br />

4. methanol<br />

5. isopropyl alcohol<br />

6. ethanol<br />

7. n-dodecane<br />

8. n-tridecane<br />

9. n-butanol<br />

3<br />

5 6 7<br />

8<br />

9<br />

4<br />

min. 2 4 6 8 10 12 14 16<br />

GC_PC00195<br />

Column: Rt ® -TCEP, 60m, 0.25mm ID, 0.4µm (cat.# 10999)<br />

Inj.:<br />

1.0µL split injection, components @ 500ppm.<br />

Oven temp.: 60°C (hold 5 min.) to 100°C @ 5°C/min. (hold 10 min.)<br />

Inj./det. temp.: 200°C<br />

Carrier gas: helium<br />

Linear velocity: 30cm/sec. set @ 80°C<br />

FID sensitivity: 6.4 x 10 -11 AFS<br />

Split flow: 46mL/min.<br />

Solutions For Your Petroleum & Petrochemical Analyses<br />

Improved best-in-class GC <strong>columns</strong> • Standards • Industry experts at your service.<br />

Visit us at www.restek.com/petro<br />

80 www.restek.com<br />