Technical Data<br />

Performance<br />

CO 2<br />

measurement range 0 ... 5 %, 0 ... 10 % or 0 ... 20 %<br />

Accuracy (<strong>in</strong>clud<strong>in</strong>g repeatability,<br />

non-l<strong>in</strong>earity and calibration<br />

uncerta<strong>in</strong>ty)<br />

±(1.5% of range + 3 % of read<strong>in</strong>g)<br />

Long-term stability<br />

±5 % FS/2 years<br />

Response time T 90<br />

1 m<strong>in</strong> at 0.5 l/m<strong>in</strong> flow<br />

Flow rate dependence<br />

< 1 l/m<strong>in</strong> flow no effect<br />

1 ... 10 l/m<strong>in</strong> flow 4 % of read<strong>in</strong>g/ l/m<strong>in</strong><br />

Temperature dependence, typical<br />

-0.3 % of read<strong>in</strong>g/°C<br />

Pressure dependence, typical<br />

+0.15 % of read<strong>in</strong>g/hPa<br />

Warm-up time<br />

1 m<strong>in</strong>, 10 m<strong>in</strong> for full<br />

specifications<br />

Product lifetime<br />

> 10 years<br />

Dimensions<br />

Dimensions <strong>in</strong> mm<br />

TUBE I.D. Ø4<br />

(TUBE O.D. Ø6)<br />

30<br />

25<br />

65<br />

6<br />

60<br />

mA<br />

V<br />

0<br />

B<br />

A<br />

0<br />

24V<br />

Ø3.2<br />

mA<br />

V<br />

0<br />

B<br />

A<br />

0<br />

24V<br />

96<br />

Operat<strong>in</strong>g Environment<br />

Temperature<br />

+5 ... +55 °C (+41 ... +131 °F)<br />

Humidity<br />

0 ... 99 % RH non-condens<strong>in</strong>g<br />

Pressure<br />

700 ... 1200 hPa<br />

Gas flow<br />

operat<strong>in</strong>g range<br />

< 10 l/m<strong>in</strong><br />

recommended range<br />

0.2 ... 0.8 l/m<strong>in</strong><br />

Electromagnetic compatibility<br />

Complies with EMS Standard EN61326-1, Generic Environment<br />

Inputs and Outputs<br />

Outputs<br />

Operat<strong>in</strong>g voltage<br />

Power consumption<br />

4 ... 20 mA, 0 ... 10 V<br />

RS485, 2-wire, non-isolated<br />

24 V (±20 %) AC/DC<br />