Troubleshooting Guide<br />

4<br />

4.1 Display Messages<br />

4.1.1 Start Up Message<br />

GIBCO BRL<br />

press STOP to resume<br />

This is the start up message when the power supply is first turned on. An alarm will<br />

also sound. Press the start/stop keypad to display the last used program.<br />

4.1.2 End of <strong>Electrophoresis</strong> Display<br />

regulation END<br />

press STOP to resume<br />

<strong>Power</strong> output has stopped at the end of the last programmed step. Change the step<br />

or program to continue.<br />

4.2 Fault Detection Messages<br />

WARNING! This notice and the messages below alert you to potentially dangerous<br />

situations.<br />

CAUTION! This notice means serious damage may occur to your power supply or<br />

chamber.<br />

4.2.1 No Load Detection Message<br />

LOAD fault<br />

press STOP to resume<br />

The LOAD Fault Detection message and the continuous tone alarm indicate that the<br />

circuit between the power supply and the electrophoresis apparatus is not closed.<br />

Check that all of the leads are connected properly and press the start/stop keypad<br />

to resume.<br />

An open connection could be caused by any of the following conditions:<br />

1. a broken electrode within the electrophoretic chamber,<br />

2. a broken wire within the connecting cord,<br />

3. insufficient buffer levels in the electrophoretic chamber, or<br />

4. electrochemical energy stored within the electrophoretic chamber.<br />