5.11.3 The pump fail to achieve specified performance (failure to reach ultimate vacuum)<br />
PAGE<br />
34<br />
MAINTENANCE<br />
• Your measuring technique or gauge is unsuitable.<br />
• There is a leak in the external vacuum system.<br />
• The gas-ballast control is op<strong>en</strong>.<br />
• The oil level is too low.<br />
• You have filled the pump with the wrong type of oil.<br />
• The oil is contaminated.<br />
• The pump has not primed.<br />
• The vacuum fittings are dirty.<br />
• The inlet-filter is blocked.<br />
• The pump has not warmed up.<br />
• The motor shaft rotates in the wrong direction (three phase motors only).<br />
5.11.4 The pump is noisy<br />
• The motor fan-cover is damaged.<br />
• The coupling <strong>and</strong> fan assembly are poorly adjusted.<br />
• The motor coupling is loose.<br />
• The motor bearings are worn.<br />
• The oil is contaminated with solid particles.<br />
• One of the pump blades is sticking.<br />
5.11.5 The pump surface temperature is above 100 o C<br />
• The ambi<strong>en</strong>t temperature is too high.<br />
• The cooling-air supply is insuffici<strong>en</strong>t or is too hot.<br />
• The cooling-air supply is blocked.<br />
• The electrical supply voltage is too high.<br />
• The exhaust filter or exhaust line is blocked.<br />
• The oil level is too low.<br />
• The pump is filled with the wrong type of oil.<br />
• The oil is contaminated.<br />
• The process gas is too hot, or the throughput is too high.<br />
Issue K 34 Jul 05