Saes Monotorr PS8 MGS 10 120 Spec T - Teesing
Saes Monotorr PS8 MGS 10 120 Spec T - Teesing
Saes Monotorr PS8 MGS 10 120 Spec T - Teesing
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Facility <strong>Spec</strong>ifications<br />
Purifier Model <strong>PS8</strong>-<strong>MGS</strong><strong>10</strong>-N/R <strong>PS8</strong>-<strong>MGS</strong>20-N/R <strong>PS8</strong>-<strong>MGS</strong>30-N/R <strong>PS8</strong>-<strong>MGS</strong>40-N/R<br />
Dimensions<br />
(If internal bypass column is<br />
ordered, see <strong>MGS</strong>50~60<br />
dimensions.)<br />
Purifier Clearance<br />
Mechanical Connections<br />
Height: 1375 mm (~54 in)<br />
Width: 800 mm (~31 in)<br />
Depth: 400 mm (~15 in)<br />
Weight: < 1<strong>10</strong> Kg (< 242 lbs.)<br />
1 meter ( ~3 feet) in front; 150 mm (~6 inches) left and right sides<br />
Feed Inlet: ½” FVCR<br />
Purified Outlet: ½” FVCR<br />
Regeneration/Relief Vent: ½” FVCR<br />
H 2 Inlet: ¼” Tube Stub<br />
Cabinet Vent: 4” duct<br />
Instrument Air Inlet: 3/8” FNPT<br />
Ambient air + 30°C (+ 54°F)<br />
Programmable<br />
Regeneration Vent Temp.<br />
Purify Time<br />
N 2 /R Consumed per Regen 15 m 3 15 m 3 30 m 3 30 m 3<br />
H 2 Consumed per Regen 0.03 m 3 @ 0.6 slpm 0.042 m 3 @ 0.6 slpm 0.071 m 3 @ 1 slpm 0.075 m 3 @ 1 slpm<br />
Main Power Installed Power: 675 W Installed Power: 1150 W<br />
208 or 230 VAC, 1 Phase, 50/60 Hz<br />
Power Consumption during Purification: 0.1 kW<br />
Power Consumption during Regeneration: 0.26 kW Power Consumption during Regeneration: 0.43 kW<br />
Separate Control Power<br />
(Option)<br />
1<strong>10</strong> or 230 VAC, 1 Phase, 50/60 Hz<br />
Power Consumption: 0.1 kW<br />
Cabinet Ventilation 170 Nm 3 /hr (<strong>10</strong>0 scfm) [@ Facilities Vacuum - 0.2 inches H 2 O]<br />
Installation<br />
Indoor, Ambient Temp (min – max) 5° – 35°C (41° – 95°F)<br />
Dimensions<br />
<strong>PS8</strong>-<strong>MGS</strong><strong>10</strong>-N/R (<strong>10</strong> Nm3/hr) through <strong>PS8</strong>-<strong>MGS</strong>40-N/R (40 Nm3/hr)<br />
S1<strong>10</strong>‐925_B DCN 4907 3 <strong>Spec</strong>ification subject to change<br />
WWW.TEESING.COM | +31 70 4130 750