<strong>of</strong> the size distribution in the SMPS window. It is preferable to leave the system<br />
running with SMPSDAT as the active window.<br />
On power up, almost all operating parameters <strong>of</strong> the electrostatic classifier ESC should<br />
be automatically set to their proper values. The two exceptions to this are the sheath<br />
flow rate and the source <strong>of</strong> the DMA voltage control. These should be properly set by<br />
commands from the SMPSDAT program when it is started but the operator should<br />
verify the correct settings on the front panel display <strong>of</strong> the ESC. Refer to Chapter 5 <strong>of</strong><br />
the Model 3080 manual for detailed information on the front panel display and control<br />
knob.<br />
The box in the lower left corner <strong>of</strong> the ESC display indicates the source <strong>of</strong> the DMA<br />
voltage control. On power up this is set to “Panel Ctrl”. If the SMPSDAT program has<br />
not done so, change this setting to “Analog Ctrl”. Above this is a large box labeled<br />
“Sheath Flowrate”. Normally this box indicates the measured sheath flowrate, not the<br />
sheath flowrate setting. Note that it takes several seconds for the active flow control to<br />
stabilize at a new level after a setting change. If the stable flow reading is not “7.0<br />
lpm,” check the setting by highlighting the box and pressing the control knob. If<br />
necessary, adjust the setting to “7.0 lpm” and press the knob again.<br />
A menu <strong>of</strong> other ESC operating parameters may be accessed through the box at the<br />
bottom center <strong>of</strong> the display. <strong>Table</strong> 5 shows the standard settings for the operating<br />
parameters shown on this menu. All except for sheath flowrate should be correct on<br />
power up. If necessary, the sheath flowrate may be set here but it is usually more<br />
convenient to set it as described above. Note that only the flowrate setting is shown on<br />
the menu, never the measured flowrate.<br />
<strong>Table</strong> 5. Standard TSI ESC Menu Parameters<br />
PARAMETER NORMAL DISPLAY ALLOWABLE VARIATIONS<br />
Sheath Flow Mode Dual Blower<br />
Sheath Flowrate 7.0 lpm<br />
Bypass Flowrate Disabled<br />
DMA Model Model 3081<br />
Impactor .0508 cm<br />
Cabinet Temperature 22.0 C near room temperature<br />
Display Contrast 4 as desired<br />
Display Brightness 7 as desired<br />
Flow Calibration for rezeroing flow xducers only<br />
Diagnostic Off<br />
Firmware Version 1.28<br />
27