<strong>of</strong> the size distribution in the SMPS window. It is preferable to leave the system<br />

running with SMPSDAT as the active window.<br />

On power up, almost all operating parameters <strong>of</strong> the electrostatic classifier ESC should<br />

be automatically set to their proper values. The two exceptions to this are the sheath<br />

flow rate and the source <strong>of</strong> the DMA voltage control. These should be properly set by<br />

commands from the SMPSDAT program when it is started but the operator should<br />

verify the correct settings on the front panel display <strong>of</strong> the ESC. Refer to Chapter 5 <strong>of</strong><br />

the Model 3080 manual for detailed information on the front panel display and control<br />

knob.<br />

The box in the lower left corner <strong>of</strong> the ESC display indicates the source <strong>of</strong> the DMA<br />

voltage control. On power up this is set to “Panel Ctrl”. If the SMPSDAT program has<br />

not done so, change this setting to “Analog Ctrl”. Above this is a large box labeled<br />

“Sheath Flowrate”. Normally this box indicates the measured sheath flowrate, not the<br />

sheath flowrate setting. Note that it takes several seconds for the active flow control to<br />

stabilize at a new level after a setting change. If the stable flow reading is not “7.0<br />

lpm,” check the setting by highlighting the box and pressing the control knob. If<br />

necessary, adjust the setting to “7.0 lpm” and press the knob again.<br />

A menu <strong>of</strong> other ESC operating parameters may be accessed through the box at the<br />

bottom center <strong>of</strong> the display. <strong>Table</strong> 5 shows the standard settings for the operating<br />

parameters shown on this menu. All except for sheath flowrate should be correct on<br />

power up. If necessary, the sheath flowrate may be set here but it is usually more<br />

convenient to set it as described above. Note that only the flowrate setting is shown on<br />

the menu, never the measured flowrate.<br />

<strong>Table</strong> 5. Standard TSI ESC Menu Parameters<br />

PARAMETER NORMAL DISPLAY ALLOWABLE VARIATIONS<br />

Sheath Flow Mode Dual Blower<br />

Sheath Flowrate 7.0 lpm<br />

Bypass Flowrate Disabled<br />

DMA Model Model 3081<br />

Impactor .0508 cm<br />

Cabinet Temperature 22.0 C near room temperature<br />

Display Contrast 4 as desired<br />

Display Brightness 7 as desired<br />

Flow Calibration for rezeroing flow xducers only<br />

Diagnostic Off<br />

Firmware Version 1.28<br />

27