Table of Contents - California Air Resources Board
Table of Contents - California Air Resources Board
Table of Contents - California Air Resources Board
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
• CRPAQS: <strong>California</strong> Regional Particulate <strong>Air</strong> Quality Study<br />
• DMA: Differential Mobility Analyzer<br />
• ESC: Electrostatic Classifier<br />
• LDMA: Long Differential Mobility Analyzer<br />
• OPC: Optical Particle Counter<br />
• PM10 inlet: A standard precut device to select particles <strong>of</strong> less than 10-µm diameter<br />
• PMS: Particle Measuring Systems<br />
• PSL: Polystyrene Latex<br />
• SMPS: Scanning Mobility Particle Sizer<br />
4. Health and Safety Warnings<br />
4.1. Laser<br />
The LASAIR contains a HeNe laser, and exposure to the direct or scattered laser light<br />
should be avoided.<br />
4.2. Radioactive Source<br />
The TSI SMPS contains a Kr85 radioactive source.<br />
5. Cautions<br />
The pump for the CPC should be exhausted outside, well away from any hydrocarbon<br />
canister samplers.<br />
None.<br />
6. Interferences<br />
7. Personnel Qualifications<br />
The system requires a technically experienced operator who understands the system, its<br />
operation and calibration. This operating procedure assumes the operator can properly<br />
use flow standards and is familiar with computer operations.<br />
Inlet consisting <strong>of</strong>:<br />
8. Apparatus and Materials<br />
• PM10 inlet<br />
• downtube<br />
• flow splitter, plumbing and electrically conducting flexible tubing<br />
• bypass flow line with rotameter and pump<br />
Climet Spectro 0.3<br />
Particle Measuring Systems LASAIR 1003<br />
TSI SMPS system consisting <strong>of</strong>:<br />
9