NELOTEST<br />

N<br />

300<br />

<strong>Rohmann</strong> <strong>GmbH</strong><br />

• 2-Channel Eddy Current System Featuring The Performance<br />

And Precision Of A Lab Instrument<br />

• Static And Dynamic Inspection Without Trade-Offs


N<br />

ELOTEST<br />

N<br />

300<br />

Technical Data<br />

General<br />

ELOTEST N300 – The high-performance alternative<br />

to the PC-plug-in card. A MUST for all those who<br />

require the highest eddy curent performance data<br />

and want to work with a PC-interconnection such<br />

as high-resolution crack detection on rail heads. Also<br />

ideally suited for labs, schools, universities and<br />

training facilities.<br />

Small, compact with up to 2 independent test channels<br />

for universal and flexible use.<br />

Independent of the PC-hardware due to the<br />

connection via an RJ45 ethernet interface.<br />

x<br />

Technical Data<br />

Eddy Current Features<br />

• Test frequency 10 Hz – 10 MHz<br />

• Transmitting driver: 15 Vss, max. 250 mA, short-circuitproof<br />

• Preamplification 6 – 72 dB (6 dB to 52 dB for more than<br />

100 kHz) in 0.5-dB steps<br />

• Gain 0 dB to 60 dB in 0.5-dB steps<br />

• Y-axis spread 0 dB to 30 dB in 1-dB steps<br />

• Phasing 0° - 359.5° in 0.5° steps<br />

• Signal filter LP/HP: 1.8 Hz to 10 kHz in 40 steps separately<br />

adjustable; band-pass with variable bandwidth; HP may<br />

be disabled.<br />

Probe Connection<br />

• All probe types may be used<br />

• Probe connection via 15-pin D-Sub-connector<br />

Universal Scanner Interface<br />

• Multifunction connector (44-pin HD-Sub-connector) for<br />

rotary pulse encoder, analog encoder, analog output and<br />

TTL I/O featuring the following counter modes:<br />

- Adjustable timebase<br />

- Up-/down-counter with simple pulse, double pulse and<br />

quadruple pulse<br />

- Analog input for linear encoders<br />

Additional Interfaces<br />

• Ethernet port (10/100 Mbit/s; RJ45, 100 BaseT)<br />

• USB-port (type 1)<br />

• VGA-port for an external monitor<br />

Housing<br />

• Aluminum profile housing with base<br />

• All connections on one front<br />

• Forced ventilation with a fan<br />

Dimensions<br />

• Depth approx. 375 mm (14.76”)<br />

• Height approx. 217 mm (8.54”)<br />

• Width approx. 56 mm (2.2”)<br />

<strong>Rohmann</strong> <strong>GmbH</strong> • Rudolf-Diesel-Str. 13 • 67227 Frankenthal • GERMANY • Tel. +49(0)6233 - 3789-0 • Fax +49(0)6233 - 3789-77<br />

Internet: www.<strong>Rohmann</strong>.de • E-Mail: info@rohmann.de<br />

<strong>Rohmann</strong> <strong>GmbH</strong><br />

Weight<br />

Weight without external power pack 2.2 kg (4.85 lbs)<br />

Power Supply<br />

• Wide-range power pack, external<br />

• Input 120 – 240 V/AC; output 24 V/DC; 100 W<br />

Instrument Versions<br />

• ELOTEST N310: 1-channel test instrument<br />

• ELOTEST N320: 2-channel version; with two independent<br />

test channels may be configured as master-/slave channel<br />

Operating Features<br />

• Box instrument without operating elements of its own<br />

• Operation via Ethernet-interface (TCP/IP/protocol) and a<br />

suitable software<br />

• Win-Client operating software with parameter handling,<br />

signal display and recording for Windows PC-systems<br />

included in the scope of delivery<br />

• Description of the interface for customized software available<br />

upon request<br />

OPTIONS<br />

Probe Array Multiplex<br />

• Max. 32 probes<br />

Conductivity Measurement<br />

• Measurement in % IACS or MS/m between 1 % IACS and<br />

110 % IACS;<br />

• Test frequency 60 kHz, 120 kHz, 240 kHz or 480 kHz<br />

respectively<br />

• Calibration to 2 individually adjustable calibration marks.<br />

Layer-Thickness Measurement<br />

• Measures non-conductive layers on conductive non-ferritic<br />

materials<br />

• Measuring range up to 1000 µm or 40 mils<br />

Doc. N300, 4/04 All rights reserved. Patents pending. Registered design MR 1.140. The right for changes of technical data representing the progress is reserved.