2-8< 2. Installation >(6) Screw the protector to fix the sensor to the holder. In this case, remove the protective cap andsecure the protector so that the flange of the sensor compresses the O-ring firmly. about theinstallation method.<strong>pH</strong> <strong>Sensor</strong>O-ringProtectorF2.9.aiFigure 2.9 Installing <strong>Sensor</strong> in Submersion Holder(7) Close the waterproof cap, and attach the holder to the arm pipe, flange or cleaner holdercompletely.2.2.3 Installing <strong>Sensor</strong> in Flow-through HolderTo install the sensor in a flow-through holder, proceed as follows:(1) Connect the sensor cable to the associated instrument. First, remove the sensor fixing nut andpass the sensor cable through the nut.<strong>pH</strong> <strong>Sensor</strong><strong>Sensor</strong> Fixing NutFigure 2.10 Preparation for <strong>Sensor</strong> Cable ConnectionProperly connect the sensor cable by referring to Section 2.3.(2) Mount the liquid junction on the sensor body. Peel off the seal covering the liquid junctionmounting hole in the sensor body. Screw the liquid junction softly into the hole by turning it fortwo or three turns.IM 12B7J1-01E