SCOPE OF ACCREDITATION TO ISO/IEC <strong>17025</strong>:<strong>2005</strong>& ANSI/NCSL Z540-1-1994 & ANSI/NCSL Z540.3-2006INSTRUMENT RENTAL LABS2100 W. 6th Ave.Broomfield, CO 80020Everett Hedrick Phone: 303 469 5335CALIBRATIONValid until: May 31, 2015 Certificate Number: 3104.01In recognition <strong>of</strong> the successful completion <strong>of</strong> the A2LA evaluation process, <strong>accreditation</strong> is granted <strong>to</strong>this labora<strong>to</strong>ry <strong>to</strong> perform the following calibrations 1 :I. Electrical – DC/Low FrequencyParameter/Equipment Range CMC 2, 3 () CommentsDC Voltage –Generate & MeasureUp <strong>to</strong> 220 mV(0.22 <strong>to</strong> 2.2) V(2.2 <strong>to</strong> 11) V(11 <strong>to</strong> 22) V(22 <strong>to</strong> 220) V(220 <strong>to</strong> 1100) V(1.0 <strong>to</strong> 40) kV34 µV/V21 µV/V21 µV/V9.3 µV/V10 µV/V14 µV/V0.18 %Fluke 5720A-03WBFluke 289Ross Eng. HV dividerDC Current –Generate & MeasureUp <strong>to</strong> 220 µA(0.22 <strong>to</strong> 2.2) mA(2.2 <strong>to</strong> 22) mA(22 <strong>to</strong> 220) mA(0.22 <strong>to</strong> 1.1) A0.011 %71 μA/A71 μA/A0.018 %0.019 %Fluke 5720A-03WB(1.1 <strong>to</strong> 20.0) A0.58 %Fluke 5520A& 50 A shunt(A2LA Cert. No. 3104.01) 06/03/2013 Page 1 <strong>of</strong> 6
Parameter/Equipment Range CMC 2, 3 () CommentsDC Resistance –Generate & MeasureFixed Points1 Ω10 Ω100 Ω1 k Ω10 k Ω62 µΩ/Ω33 µΩ/Ω32 µΩ/Ω20 µΩ/Ω19 µΩ/ΩFluke-742A-1IET-SRX-10IET-SRX-100Fluke-742A-1KFluke-742A-10K1 Ω(>1 <strong>to</strong> 100) Ω(>0.10 <strong>to</strong> 1.0) kΩ(>1.0 <strong>to</strong> 10) kΩ(>10 <strong>to</strong> 100) kΩ(>0.10 <strong>to</strong> 1.0) MΩ(1 <strong>to</strong> 100) MΩ(100 <strong>to</strong> 300) MΩ0.047 %49 µΩ/Ω74 µΩ/Ω20 µΩ/Ω21 µΩ/Ω39 µΩ/Ω0.083 %0.87 %Fluke 5720AAGILENT 3458A-02(0.30 <strong>to</strong> 1.0) GΩ0.018 %HTE-1 GOhmFluke 5522AAgilent 3458A-02Parameter/Range Frequency CMC 2, 3 () CommentsAC Voltage –Generate & Measure(22 <strong>to</strong> 220) mV(0.22 <strong>to</strong> 2.2) V(2.2 <strong>to</strong> 22) V(2.2 <strong>to</strong> 22) V1 kHz20 kHz20 kHz100 kHz0.028 %0.017 %33 µV/V0.082 %Fluke 5720A-03WB,Agilent 3458A-02(10) V100Hz <strong>to</strong> 1MHz0.09 %(700) V30 kHz0.19 %Fluke 5720A-03WB,Fluke 5725A amplifier(1 <strong>to</strong> 15) kV60 Hz0.13 %Fluke 289Ross Eng. HV divider(A2LA Cert. No. 3104.01) 06/03/2013 Page 2 <strong>of</strong> 6
Parameter/Range Frequency CMC 2, 3 () CommentsAC Current –Generate & MeasureUp <strong>to</strong> 0.220 mA(0.22 <strong>to</strong> 2.2) mA(2.2 <strong>to</strong> 22) mA(22 <strong>to</strong> 220) mA(0.22 <strong>to</strong> 2.2) A100 Hz <strong>to</strong> 10 kHz100 Hz <strong>to</strong> 10 kHz100 Hz <strong>to</strong> 10 kHz100 Hz <strong>to</strong> 10 kHz100 Hz <strong>to</strong> 10 kHz0.37 %0.21 %0.22 %0.41 %0.65 %Fluke 5720A-03WBAgilent 3458A-02Fluke 5520A, Agilent3458A-02(2.2 <strong>to</strong> 11 ) A40 Hz <strong>to</strong> 10 kHz0.15 %Fluke 5720A-03WB w/Fluke 5725A amplifier& Fluke 289 meter(11 <strong>to</strong> 3000 ) A60 Hz0.94 %Fluke 5520A, Agilent3458A-02HTE multi-turn coilsFluke Flexi i3000s ampRogowski air cored coilCapacitance –Generate & MeasureFixed Points(1.0, 10, 100,1000) pF1 kHz1.1 %Capaci<strong>to</strong>r Set, HP16380A with InstekLCR-81710 nF <strong>to</strong> 10 000 µFAt various frequenciesranging anywherefromDC <strong>to</strong> 1 kHz0.78 %Fluke 5520A calibra<strong>to</strong>rwith Instek LCR-817meterInductance –Generate & Measure1 mH100 mH1 H(0.1 <strong>to</strong> 10) kHz(0.1 <strong>to</strong> 10) kHz(0.1 <strong>to</strong> 1) kHz0.13 %0.13 %0.15 %Genrad 1482 seriesstandard induc<strong>to</strong>rsInstek 817 LCR meter(A2LA Cert. No. 3104.01) 06/03/2013 Page 3 <strong>of</strong> 6
Parameter/Equipment Range CMC 2, 3 () CommentsElectrical Calibration <strong>of</strong>Thermocouples –Thermocouple SimulationGenerate & MeasureJ-Type(-210 <strong>to</strong> -100) ºC(-100 <strong>to</strong> -30) ºC(-30 <strong>to</strong> 150) ºC(150 <strong>to</strong> 760) ºC(760 <strong>to</strong> 1200) ºC0.27 ºC0.18 ºC0.16 ºC0.19 ºC0.25 ºCFluke 5520AK-Type(-200 <strong>to</strong> -100) ºC(-100 <strong>to</strong> -25) ºC(-25<strong>to</strong> 120) ºC(120 <strong>to</strong> 1000) ºC(1000 <strong>to</strong> 1372) ºC0.35 ºC0.20 ºC0.18 ºC0.28 ºC0.42 ºCT-Type(-250 <strong>to</strong> -150) ºC(-150 <strong>to</strong> 0) ºC(0 <strong>to</strong> 120) ºC( 120 <strong>to</strong> 400) ºC0.65 ºC0.26 ºC0.18 ºC0.16 ºCESD Impulse – Measure Voltage: (1 <strong>to</strong> 30) kV 3.8 % Agilent 54845Aoscillo<strong>scope</strong> calibra<strong>to</strong>r,EMC partner ESDVeri-V, target 2-DN,Agilent 34401A DMMOscillo<strong>scope</strong> –Rise Time 150 ps 39 ps Fluke 9500B, activehead 9530 , Tek 694Coscillo<strong>scope</strong>II. Electrical – RF/MicrowaveParameter/Range Frequency CMC 2 () CommentsRF Power LowFrequency – Generate& Measure0.1 Hz <strong>to</strong> 100 kHz 0.023 V RMS Agilent 3325Bsynthesized signalgenera<strong>to</strong>r, Agilent3458A-02 DMM(A2LA Cert. No. 3104.01) 06/03/2013 Page 4 <strong>of</strong> 6
Parameter/Range Frequency CMC 2 () CommentsRF Power – Low <strong>to</strong>Medium FrequencyGenerate & Measure(-56 <strong>to</strong> +24) dBm(1mV <strong>to</strong> 10 V)(0.1 <strong>to</strong> 20) MHz 0.095 V RMS Agilent 3325Bsynthesized signalgenera<strong>to</strong>r,HP8902A-050measuring receiver,HP 11722A sensorRF Power –Generate & Measure(-5 <strong>to</strong> -85) dBm1 GHz0.69 dBmHP 8902A-050measuring receiver;11793A microwaveconverter; HP 8340Bsynthesizer-assetHTE0333 only)w/ 11722A sensor(-5 <strong>to</strong> -85) dBm7 GHz1.2 dBmw/ 11792A sensor(-5 <strong>to</strong> -85) dBm12.4 GHz1.4 dBmw/ 11792A sensor(-5 <strong>to</strong> -85) dBm20 GHz & 26 GHz2.1 dBmw/ 11792A sensorRF Attenuation TunedRF Level – Generate &Measure(10 <strong>to</strong> 100 dB) 10 MHz50 MHz1 GHz4 GHz10 GHz15 GHz18 GHz0.53 dB0.55 dB0.63 dB0.50 dB0.94 dB1.4 dB0.51 dBAgilent 8902A-050measuring receiverAgilent 11793Amicrowave converter,HP 8496B / 8494Bstep attenua<strong>to</strong>rsw/ Agilent 11722Asensorw/ 11792A powersensor(A2LA Cert. No. 3104.01) 06/03/2013 Page 5 <strong>of</strong> 6
Parameter/Range Frequency CMC 2 () CommentsPhase Noise –Measure10 MHz <strong>to</strong> 1 GHz(2 <strong>to</strong> 18) GHz2.7 dBc/Hz3.0 dBc/HzHP 3048A phasenoise test setHP/Agilent 11848A,11729C, 3561A,3585A, HP 8663AFluke 9640A-LPN& HP 8340BIII. Time & FrequencyParameter/Equipment Range CMC 2 () CommentsFrequency –Measuring Equipment (1.0 <strong>to</strong> 26.5) GHz 5.0 parts in 10 12 Agilent 5071A (cesiumbeam), Agilent 8340Bsynthesized sweeperFrequency –Measure10 Hz <strong>to</strong> 40 GHz 1 Hz Agilent 5071A (cesiumbeam), HP 5352B-010________________________________1 This labora<strong>to</strong>ry <strong>of</strong>fers commercial calibration service.2 Calibration and Measurement Capability (CMC) is the smallest uncertainty <strong>of</strong> measurement that alabora<strong>to</strong>ry can achieve within its <strong>scope</strong> <strong>of</strong> <strong>accreditation</strong> when performing more or less routinecalibrations <strong>of</strong> nearly ideal measurement standards or nearly ideal measuring equipment. Calibrationand Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 %level <strong>of</strong> confidence, usually using a coverage fac<strong>to</strong>r <strong>of</strong> k = 2. The actual measurement uncertainty <strong>of</strong> aspecific calibration performed by the labora<strong>to</strong>ry may be greater than the CMC due <strong>to</strong> the behavior <strong>of</strong>the cus<strong>to</strong>mer’s device and <strong>to</strong> influences from the circumstances <strong>of</strong> the specific calibration.3Unless listed otherwise stated, all CMCs stated in percent are in percent <strong>of</strong> reading(A2LA Cert. No. 3104.01) 06/03/2013 Page 6 <strong>of</strong> 6
A2LA has accreditedINSTRUMENT RENTAL LABSBroomfield, COfor technical competence in the field <strong>of</strong>CalibrationThis labora<strong>to</strong>ry is accredited in accordance with the recognized International Standard ISO/IEC <strong>17025</strong>:<strong>2005</strong> General Requirements forthe Competence <strong>of</strong> Testing and Calibration Labora<strong>to</strong>ries. This labora<strong>to</strong>ry also meets the requirements <strong>of</strong> ANSI/NCSLI Z540-1-1994 and therequirements <strong>of</strong> ANSI/NCSLI Z540.3-2006 and any additional program requirements in the field <strong>of</strong> calibration. This <strong>accreditation</strong>demonstrates technical competence for a defined <strong>scope</strong> and the operation <strong>of</strong> a labora<strong>to</strong>ry quality management system(refer <strong>to</strong> joint ISO-ILAC-IAF Communiqué dated 8 January 2009).Presented this 3 rd day <strong>of</strong> June 2013._______________________President & CEOFor the Accreditation CouncilCertificate Number 3104.01Valid <strong>to</strong> May 31, 2015For the calibrations <strong>to</strong> which this <strong>accreditation</strong> applies, please refer <strong>to</strong> the labora<strong>to</strong>ry’s Calibration Scope <strong>of</strong> Accreditation.