Parameter/Equipment Range CMC 2, 3 () CommentsDC Resistance –Generate & MeasureFixed Points1 Ω10 Ω100 Ω1 k Ω10 k Ω62 µΩ/Ω33 µΩ/Ω32 µΩ/Ω20 µΩ/Ω19 µΩ/ΩFluke-742A-1IET-SRX-10IET-SRX-100Fluke-742A-1KFluke-742A-10K1 Ω(>1 <strong>to</strong> 100) Ω(>0.10 <strong>to</strong> 1.0) kΩ(>1.0 <strong>to</strong> 10) kΩ(>10 <strong>to</strong> 100) kΩ(>0.10 <strong>to</strong> 1.0) MΩ(1 <strong>to</strong> 100) MΩ(100 <strong>to</strong> 300) MΩ0.047 %49 µΩ/Ω74 µΩ/Ω20 µΩ/Ω21 µΩ/Ω39 µΩ/Ω0.083 %0.87 %Fluke 5720AAGILENT 3458A-02(0.30 <strong>to</strong> 1.0) GΩ0.018 %HTE-1 GOhmFluke 5522AAgilent 3458A-02Parameter/Range Frequency CMC 2, 3 () CommentsAC Voltage –Generate & Measure(22 <strong>to</strong> 220) mV(0.22 <strong>to</strong> 2.2) V(2.2 <strong>to</strong> 22) V(2.2 <strong>to</strong> 22) V1 kHz20 kHz20 kHz100 kHz0.028 %0.017 %33 µV/V0.082 %Fluke 5720A-03WB,Agilent 3458A-02(10) V100Hz <strong>to</strong> 1MHz0.09 %(700) V30 kHz0.19 %Fluke 5720A-03WB,Fluke 5725A amplifier(1 <strong>to</strong> 15) kV60 Hz0.13 %Fluke 289Ross Eng. HV divider(A2LA Cert. No. 3104.01) 06/03/2013 Page 2 <strong>of</strong> 6