3.<br />

4.<br />

17<br />

A Guide To <strong>Biosafety</strong> And <strong>Biosafety</strong> <strong>Cabinet</strong>s<br />

The burner should be placed as far into the cabinet working space as<br />

possible to minimize its effect on cabinet performance.<br />

A solenoid valve can be interlocked with the gas supply to automatically<br />

shut off the burner<br />

4. Using BSCs: Shutdown<br />

Note: operate the cabinet continuously whenever possible to maximize<br />

protection- Surface decontaminate all items<br />

� Seal biohazard bags if used<br />

� Surface decontaminate the cabinet innerside walls, back wall, work surface,<br />

drain pan and the inner side of the sliding sash / hinged window<br />

� Allow work zone air to purge<br />

� When available: install front closure or close sash and activate UV lamp<br />

(if desired)<br />

5. User maintenance<br />

� Pre-filter replacements once every 3 months<br />

� UV lamp replacement once a year<br />

�<br />

Operational verification if appropriate (e.g. Alarm checks, basic airflow<br />

measurements etc.)