Esco Labculture® Plus Class II Type A2 Biosafety Cabinet - Comlibris
Esco Labculture® Plus Class II Type A2 Biosafety Cabinet - Comlibris
Esco Labculture® Plus Class II Type A2 Biosafety Cabinet - Comlibris
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
3.<br />
4.<br />
17<br />
A Guide To <strong>Biosafety</strong> And <strong>Biosafety</strong> <strong>Cabinet</strong>s<br />
The burner should be placed as far into the cabinet working space as<br />
possible to minimize its effect on cabinet performance.<br />
A solenoid valve can be interlocked with the gas supply to automatically<br />
shut off the burner<br />
4. Using BSCs: Shutdown<br />
Note: operate the cabinet continuously whenever possible to maximize<br />
protection- Surface decontaminate all items<br />
� Seal biohazard bags if used<br />
� Surface decontaminate the cabinet innerside walls, back wall, work surface,<br />
drain pan and the inner side of the sliding sash / hinged window<br />
� Allow work zone air to purge<br />
� When available: install front closure or close sash and activate UV lamp<br />
(if desired)<br />
5. User maintenance<br />
� Pre-filter replacements once every 3 months<br />
� UV lamp replacement once a year<br />
�<br />
Operational verification if appropriate (e.g. Alarm checks, basic airflow<br />
measurements etc.)