5.5 Intra-Operative ScanIntervention Workflow5.5.1 Removing Non MR-Compatible PartsPrior to each scan with the MR-System, you have to remove all non MR-compatibleparts. This could be clamps, scalpels, <strong>NORAS</strong> accessories such as the VarioGuideAdapter, Sugita ® Adapter or others.5.5.2 Mounting of the Top Array Coil1 2 3Cover all with one layerof sterile drape, use asecond sterile drape todouble – cover thesterile clampsMount right coil holder, besure that there are twolayers of sterile drapebetween the holder and thesterile clampsMount left coil holder, besure that there are twolayers of sterile drapebetween the holder andthe sterile clamps4 5 6Mount coil to the highestpositionAdjust coil to the lowestpossible positionMaximum position is 30 !Tighten all multipurposescrews by handPage 5-41