Cleaning, Disinfection&SterilizingInstructionsAutomateddisinfectionDisinfectant: Depend on RDG andVario-TD-program you can selectone disinfectant which has beenapproved and released forautomated disinfection.Procedure:If automated disinfecting isemployed, a final rinse at 194°F(90°C) for 5 minutes may be usedto effect thermal disinfection. Anyassembly should be dis-assembledinto its individual componentsbefore, so that an optimaldisinfection is guaranteed.Individual partsOnly for: Titanium skull pins, screwfor Ball-Joint, nut for Ball-Joint,screws for C-Arc, screws forclamps, clamps, pin retain screw,adjustment nuts, square retainer,pin holders, LEYLA-Retractorsockets, screws for Miyabi Shell,Two-Point-Fixation, fixation, coilholders, multipurpose screwsNot allowed for: <strong>Head</strong> coils (bottom & top array) All cables Base plate C-Arc C-Arc holderAdviceSome substances contained in disinfectants are known to behazardous to your health. Their concentration in thesurrounding air must be below a certain legal limit. Followmanufacturer’s instructions.AdviceUse commercially available liquid disinfection solutions.For surface disinfection, all disinfectants, which were testedand approved by the Robert-Koch-Institute in Germany(www.rki.de) for 5 minutes disinfection are suitable.Page 8-67

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!