Rev. 2<br />

3<br />

Regulator outlet pressure, typical 2-3 bar or 30 to 40 psi (not to exceed 6.0 bar/90 psi).<br />

A Nitrogen “System Purge Gas” of 50- 75 cc is required. This gas should be clean <strong>and</strong><br />

dry having minimum purity of 99.99%. Regulator Outlet Pressure 2-3 bar or 30 to 40 psi<br />

(not to exceed 6.0 bar/90 psi).<br />

For Subambient Liquid Nitrogen operation:<br />

A low pressure Nitrogen source is required to pressurize the LN2 Liquid Nitrogen Tank.<br />

25-30 PSI is required.<br />

“Sample Purge Gas” such as dry argon, nitrogen, helium; minimum purity of 99.99% .<br />

(A gas having a dew point ≤ -180 °C is recommended)<br />

Regulator outlet pressure, typical 2-3 bar or 30 to 40 psi (not to exceed 6.0 bar/90 psi).<br />

A Nitrogen “System Purge Gas” of 50- 75 cc is required. This gas should be clean <strong>and</strong><br />

dry having minimum purity of 99.99%. Regulator Outlet Pressure 2-3 bar or 30 to 40 psi<br />

(not to exceed 6.0 bar/90 psi).<br />

For all Autosampler subambient operation:<br />

A dry box purge may be required depending on the relative humidity in the laboratory. A<br />

separate purge line is required to control the flow between 35 to 50 ml/min. This can be<br />

accomplished using a restrictor (provided with the analyzer).<br />

Gas Line Tubing:<br />

Provided with the <strong>Jade</strong> is approximately 10 feet (~3 meters) of 1/8” Teflon tubing <strong>and</strong><br />

fittings to be used to connect one sample purge <strong>and</strong> one system purge line. Additional<br />

lengths if required due to the location of the gas source or additional sample purge gases<br />

is the responsibility of the customer. Customer is to notify the Customer Care Agent if<br />

more purge/sample gas tubing is needed.<br />

Regulators:<br />

Customer must provide the necessary regulators to connect the required purge <strong>and</strong><br />

system gases. Regulators are available from <strong>Perkin</strong><strong>Elmer</strong> at an additional cost.<br />

N5190462 Gas Split Kit provides a convenient method of splitting a single source of gas.<br />

One kit required for each split in the source gas.<br />

Filter Driers:<br />

Moisture in purge gases will negatively affect instrument performance. Filter driers<br />

N5370103 are strongly recommended to maintain a clean dry purge gas.<br />

e) Coolant requirements:<br />

The <strong>Jade</strong> <strong>DSC</strong> requires a coolant of one of the following types:<br />

Circulating water: A source of water <strong>and</strong> a drain are required. Flow rate of 5 l/min<br />

required.<br />

Chiller: A liquid circulating device such as the PolyScience Model 9102. The 9102<br />

water circulator does not include the cooling liquid. 50/50 Ethylene Glycol (non<br />

automotive grade, Fisher P/N E1774), Distilled Water <strong>and</strong> Algaecide (N0776059 8 drops<br />

per gallon) is required as the cooling media to -20 °C. Minimum temperature control ±<br />

0.1 °C is required.<br />

Intercooler 6P Refrigeration based.