4 Applications<br />

Fig. 20:<br />

<strong>pH</strong> <strong>measurement</strong> in a swimming pool<br />

(1) Swimming pool (3) Filter<br />

(2) Raw water (4) Purified water<br />

The <strong>pH</strong> electrode is hardly exposed to any contaminants in the pool water. Electrodes systems with<br />

electrolyte solution (possibly slightly thickened) and ceramic diaphragms have the best <strong>measurement</strong><br />

behavior here. The style used must be designed for an excess pressure, even if this is only<br />

slight, in the measured pipeline. Electrodes with a reference electrolyte gel age very rapidly. As well<br />

as the more frequent outlay for new electrodes, this also results in increased maintenance costs<br />

and poorer control results.<br />

36 4 Applications JUMO, FAS 622, Edition 04.07