14.12.2012 Views

proposal biomechanics gait analysis lab - Biomedical Engineering ...

proposal biomechanics gait analysis lab - Biomedical Engineering ...

proposal biomechanics gait analysis lab - Biomedical Engineering ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

PROPOSAL<br />

BIOMECHANICS GAIT ANALYSIS LAB<br />

BY:<br />

Kimberly Carr, Omar Chawiche, Angela Ensor<br />

CLIENT CONTACT:<br />

Dr. John D. Enderle<br />

University of Connecticut<br />

<strong>Biomedical</strong> <strong>Engineering</strong> Department<br />

Program Director & Professor for <strong>Biomedical</strong> <strong>Engineering</strong><br />

Bronwell Building, Room 217C<br />

260 Glenbrook Road<br />

Storrs, Connecticut 06269-2247<br />

Voice: (860) 486-5521<br />

FAX: (860) 486-2500<br />

Email: jenderle@bme.uconn.edu<br />

Website: www.eng2.uconn.edu/~jenderle<br />

BME Program Homepage: www.bme.uconn.edu<br />

EMB Magazine Homepage: EMB-Magazine.bme.uconn.edu<br />

1


TABLE OF CONTENTS<br />

Section Page Number<br />

1. Introduction 1 – 6<br />

1.1 Background 1<br />

1.2 Project Purpose 1<br />

1.3 Previous Work Done by Others 1 - 4<br />

1.3.1 Products 5<br />

1.3.2 Patent Search Results 6<br />

2. Project Description 7 – 14<br />

2.1 Objective 8 – 9<br />

2.2 Methods 9 – 13<br />

3. Budget 14<br />

4. Conclusion 15 - 16


1. Introduction<br />

1.1 Background<br />

Dr. John D. Enderle is the Editor-in-Chief of the EMB Magazine,<br />

<strong>Biomedical</strong> <strong>Engineering</strong> Book Series Editor for Morgan and<br />

Claypool Publishers, and the Program Director & Professor for<br />

<strong>Biomedical</strong> <strong>Engineering</strong> at the University of Connecticut. Dr.<br />

Enderle has been looking to improve the <strong>gait</strong> <strong>analysis</strong> <strong>lab</strong> for<br />

the undergraduate <strong>Biomedical</strong> <strong>Engineering</strong> program’s Biomechanics<br />

<strong>lab</strong>oratory. Currently, the <strong>biomechanics</strong> <strong>gait</strong> <strong>analysis</strong> <strong>lab</strong><br />

essentially consists of the students participating in the<br />

process of acquiring data through software applications in the<br />

forward motion only, which limits the students to only obtain or<br />

calculate measurements of distance, acceleration, speed, angles,<br />

and joint moments using the assumption that the ground reaction<br />

forces exerted is simply the subject’s weight. The missing<br />

component to the <strong>gait</strong> <strong>analysis</strong> is a true-life measure of ground<br />

reaction forces exerted on the foot strike, which are applied in<br />

multiple directions.<br />

1.2 Project Purpose<br />

Dr. Enderle has requested an expansion of the <strong>gait</strong> <strong>analysis</strong> <strong>lab</strong><br />

by utilizing a force measurement system. This improvement will<br />

afford students the opportunity to obtain an accurate<br />

measurement of the affects of motion in more than one direction.<br />

The new system will record and provide data on the affects of<br />

foot strike. The addition of a force measurement system will<br />

add value to the <strong>lab</strong> and the <strong>Biomedical</strong> <strong>Engineering</strong> program by<br />

allowing for a more “real world” <strong>biomechanics</strong> <strong>lab</strong>oratory<br />

experience. Many <strong>gait</strong> <strong>analysis</strong> <strong>lab</strong>oratories across the world<br />

have a force measurement system, along with a motion recording<br />

system.<br />

1.3 Previous Work Done by Others<br />

There are three categories for the collection of <strong>gait</strong> <strong>analysis</strong><br />

data that will be considered in this section; foot pressure,<br />

force plates, and motion systems. The acquired data from these<br />

devices or systems are sent to a computer software program, of<br />

which the type of program is dependent upon the types of <strong>gait</strong><br />

<strong>analysis</strong> devices used and the desired program features, to<br />

synthesize and display the results. Two of the foot pressure<br />

options consist of pressure mats and insoles. The force exerted<br />

on the ground can be collected by force plates and the motion<br />

Page 1


systems include electromyography (EMG) and a video motion<br />

system.<br />

Foot Pressure<br />

Tekscan, Inc., located in South Boston, MA, has designed and<br />

manufactured many <strong>gait</strong> <strong>analysis</strong> products, including a few models<br />

of pressure insoles and mats. (1c)Pressure insoles can be used,<br />

in a clinical setting, for evaluating patients before and after<br />

surgery, screen for neuropathic disorders such as diabetes, and<br />

measure any length discrepancies in lower limbs, which could<br />

potentially be useful in assessing patients prior to hip<br />

replacement surgery for determining implant specifications. The<br />

insoles work much the same as pressure mats, which consist of<br />

capacitive or force sensitive resistor (FSR) switches placed in<br />

between two pieces of material that is either placed on the<br />

floor as a mat or on the inside or outside of a shoe.<br />

Compression closing switches have two sheets of brass shim rock,<br />

(1b) a thin sheet of brass that has been cold rolled, with a<br />

compressible, non-conductive foam rubber insole, which has holes<br />

containing conductive rubber cylinders, in between the shim<br />

rock. The conductive rubber cylinders make contact with both<br />

sheets of sham rock and close an electric circuit upon<br />

application of pressure. Force sensitive resistor (FSR)<br />

switches have a similar sandwich approach that uses flexible<br />

plastic for the outer layers, with circuits printed on the<br />

inside of the two plastic sheets. In between the plastic<br />

sheets, there is a slender layer of double-sided adhesive, with<br />

holes that allow for areas of contact. Similar to the<br />

compression closing switch, pressure causes the electric circuit<br />

to close, except this device has a resistive electric circuit,<br />

so that when pressure is increased, the resistance decreases.<br />

Force Plate Device<br />

Several models of force plates have been designed and produced<br />

by Bertec Corporation, based in Columbus, OH, Advanced Medical<br />

Technology, Inc., based in Watertown, MA, and Kistler<br />

Instrumente AG, based in Winterthur, Switzerland. Ground<br />

reaction forces can be measured by using a force plate. These<br />

devices include a top plate and a bottom plate or frame, which<br />

are separated by force transducers attached to each corner of<br />

the rectangular or square plates. The entire device is<br />

installed into a platform so that the top plate is flush with<br />

the surface of the platform to prevent changes in a subject’s<br />

<strong>gait</strong>. As the subject walks along the platform and strikes the<br />

Page 2


top plate, the exerted force is transferred to the transducers.<br />

There are two types of force plates called piezoelectric and<br />

strain gauge. The piezoelectric force plate uses quartz<br />

transducers that create an electric charge when force is exerted<br />

on the top plate. The strain gauge force plates have load cells<br />

that measure stress when a force is exerted on the top plate.<br />

In both types, the devices provide vertical and shear forces, as<br />

well as the (1a)“center of pressure” during <strong>gait</strong>.<br />

Motion Systems<br />

Electromyography (EMG) utilizes electrogoniometers that are<br />

attached to the proximal and distal ends of the limb segments<br />

that are to be evaluated. One manufacturer of electrogoniometer<br />

is Biometrics Ltd., located in the United Kingdom and the United<br />

States, and another is Noraxon, located in the United States and<br />

Germany. These devices are electro-mechanical devices that (1a)<br />

“provide an output voltage proportional to the angular change<br />

between the two attachment surfaces.” This system for motion<br />

<strong>analysis</strong> makes the assumption that the devices move in sync with<br />

the midline of the segment being measured and therefore, give an<br />

accurate representation of the angular change at the joints of<br />

the segments under consideration. The disadvantage is the<br />

tracking (or in sync motion with the midline) is affected when<br />

the person is not a lean individual, so that any type of bulkyness<br />

will affect the accuracy of the results due to skin or<br />

muscle movement.<br />

ViconPeak is a manufacturer and supplier of motion systems and<br />

is based in the United Kingdom, with two locations in the United<br />

States. Video motion systems, also referred to as<br />

optoelectronic systems or stereophotogrammetry, use one or more<br />

digital video cameras that track markers placed on the body of a<br />

subject. The types of markers used include infrared (IR) lightemitting<br />

diodes (LEDs) and reflective markers. The infrared<br />

(IR) light-emitting diodes (LEDs)require the use of a battery,<br />

circuits, and cables that can be cumbersome to use, but<br />

alleviate the issue of a camera losing sight of the marker. The<br />

reflective markers are much simpler to use and can be just as<br />

effective, even under the restraints of the user maintaining<br />

marker visibility by the camera.<br />

Complete Analysis<br />

The foot pressure devices and the force plate device can be used<br />

in conjunction with the motion systems to offer a more<br />

comprehensive <strong>analysis</strong> of <strong>gait</strong>, which is the focus of this <strong>gait</strong><br />

Page 3


<strong>analysis</strong> project, and are used clinically. There a numerous<br />

<strong>lab</strong>oratories that utilize <strong>gait</strong> <strong>analysis</strong>, some examples include<br />

the Derby Gait and Movement Laboratory (www.<strong>gait</strong>.com), the Hugh<br />

Williamson Gait Analysis Laboratory<br />

(www.rch.org.au/<strong>gait</strong>/index.cfm?doc_id=1595) as part of the Royal<br />

Children’s Hospital in Melbourne, Australia, and the Shriner’s<br />

Hospital for Children. One company that provides entire <strong>gait</strong><br />

<strong>analysis</strong> systems is Ariel Dynamics, at www.arielnet.com.<br />

1.3.1 Products<br />

Tekscan, Inc. Products for Foot Pressure Devices:<br />

F-Scan ® System<br />

Bipedal In-Shoe Plantar<br />

Pressure/Force Measurement<br />

www.tekscan.com<br />

Walkway TM System<br />

Floor Mat-based Pressure/Force Measurement<br />

www.tekscan.com<br />

Bertec Corporation Products for Force<br />

Plates:<br />

4060 Force Plate Series<br />

www.bertec.com<br />

Dimensions in mm(342A X 552B X 29C X 24D)<br />

4060-08 model is made of solid aluminum<br />

4060-NC Force Plate Series<br />

Dimensions in mm(342A X 552B X 29C X 24D)<br />

Non-conductive and made of resin<br />

impregnated wood.<br />

www.bertec.com<br />

Page 4


NorAngle Electrogoniometer System (for<br />

Electromyography EMG)<br />

www.noraxon.com<br />

ViconPeak High Speed Video Camera<br />

Model HSC-200 PM<br />

www.vicon.com<br />

1.3.2 Patent Search Results<br />

Searching for United States patents prior to undertaking an<br />

engineering project or design is obligatory. A patent restricts<br />

the use of any part of a patented invention, in that, another<br />

company or person cannot make, use, or sell the invention in the<br />

United States for a period of time that depends on the type of<br />

invention. For a design patent, the term period is 14 years<br />

from the grant date of the patent.<br />

A quick search through the US Patent database of patents since<br />

1975 can be done at www.uspto.gov/patft/index.html. Using the<br />

term ‘<strong>gait</strong> <strong>analysis</strong>’ in the quick search brought up 110 filed<br />

patents with that term in the text of the patent form. Some of<br />

the patents include:<br />

6,997,882 6-DOF subject-monitoring device and method – February<br />

14, 2006 – Parker, et al.<br />

This patent covers an invention that includes methods and<br />

devices for monitoring and acquiring specific data from a<br />

subject’s movement and physiological measurements, which, in<br />

part, involves the use of an accelerometer.<br />

6,836,744 Portable system for analyzing human <strong>gait</strong> – December<br />

28, 2004 – Asphahani, et al.<br />

The patent describes an invention for a portable <strong>gait</strong> analyzer<br />

which uses a type of pressure insole with detachable parts.<br />

5,408,873 Foot force sensor – April 25, 1995 – Schmidt, et al.<br />

Page 5


This patent is for another type of pressure insole that consists<br />

of outer layers that are conductive and an inner layer that is<br />

nonconductive.<br />

4,631,676 Computerized video <strong>gait</strong> and motion <strong>analysis</strong> system and<br />

method – December 23, 1986 – Pugh; James W.<br />

This patent describes a computerized video <strong>gait</strong> and motion<br />

<strong>analysis</strong> system and method that use reflective markers, Sony<br />

Video Motion Analyzers, video recorders and displays, and<br />

computer system for displaying the results obtained from the<br />

data.<br />

2. Project Description<br />

2.1 Objective<br />

The objective of the project is to design a <strong>gait</strong> <strong>analysis</strong> <strong>lab</strong><br />

for a <strong>biomechanics</strong> class, so that students may learn about <strong>gait</strong><br />

<strong>analysis</strong> and how the data is analyzed through computer software<br />

applications. In general, <strong>gait</strong> <strong>analysis</strong> is used to provide<br />

objective measurements concerning a person’s movement patterns.<br />

The set-up will utilize the LabVIEW® software program to analyze<br />

analog data, and hardware parts such as a force measurement<br />

device to measure the forces and moments generated during<br />

footfall. In addition, National Instruments equipment, which is<br />

already avai<strong>lab</strong>le in the <strong>lab</strong>, will be used for the purpose of<br />

image acquisition and a digital camera to record the <strong>gait</strong> cycle.<br />

To record the <strong>gait</strong> cycle using the digital camera, external<br />

markers are placed laterally on the hip, knee, and ankle of the<br />

right or left leg depending on the camera set up, which reflect<br />

light to the camera and allow the segment of motion to be<br />

monitored. To properly set up the force measurement device in<br />

the <strong>biomechanics</strong> <strong>lab</strong>oratory, it must be placed so that the<br />

person can complete one full <strong>gait</strong> cycle and step on the<br />

measuring device, with the digital camera placed so that it will<br />

receive the reflected light from the markers. The camera should<br />

be connected to the National Instrument PXI-1031 which contains<br />

the National Instrument 1411 series, consisting of PCI and PXI<br />

plug-in image acquisition cards that receive analog signals from<br />

a standard color or monochrome camera. The BNC-2100 series<br />

connector block is added to the set-up for ease of connectivity<br />

and control of the analog input, analog output, and digital I/O.<br />

Page 6


In the <strong>lab</strong>oratory, the students will connect the digital<br />

recorder camera to the National Instrument PXI-1031 device,<br />

followed by connection to the BNC-2100 series connector block,<br />

to regulate the process, and connect BNC-2100 to the computer.<br />

Additionally, students may be required to build a portion of the<br />

LabVIEW® software program to analyze the data. Students will<br />

then place the markers on the different segments of the<br />

subject’s leg. The subject will walk towards the force<br />

measurement device, stepping on the device during the cycle,<br />

while the digital camera records the action. Throughout the<br />

entire process, the National Instrument devices and LabVIEW®<br />

program transform the digital data into analog data, which can<br />

be used to determine distance, speed, acceleration, angles, and<br />

forces acting on the body. The results will provide information<br />

that characterizes an individual’s foot strike pattern.<br />

In the end, the students will learn how the LabVIEW® program<br />

operates, how to set up a <strong>lab</strong>oratory environment, and how the<br />

entire set-up aggregates the information gathered to give<br />

meaningful results.<br />

2.2 Methods<br />

The Clinical Gait Model and Kinematic Data Analysis<br />

The clinical <strong>gait</strong> model is an algorithm that connects the data<br />

which is collected during the subject’s walking cycles to the<br />

information necessary for scientific <strong>analysis</strong>. For kinematic<br />

data <strong>analysis</strong>, three noncolinear reference points or markers are<br />

placed on a subject’s body segment to permit the tracking of the<br />

spatial orientation of that segment through space. From these<br />

reference points, a plane is formed which allows the derivation<br />

of a segmentally fixed coordinate system for the subject. In<br />

order to accurately quantify the subject’s kinematics, markers<br />

have to be referenced to their anatomy and placed either<br />

directly over apparent bony regions on the segment or at<br />

convenient positions along the segment which are visible to the<br />

measurement camera(s) being utilized. For example, markers<br />

placed over the right and left anterior-superior-iliac-spine<br />

(ASIS) and either the right or left posterior-superior-iliacspine<br />

(PSIS), for the pelvis, will allow for the calculation of<br />

an anatomically oriented segmental coordinate system (Figure 1).<br />

Page 7


Figure 1. Kinematic markers<br />

used to define pelvis and thigh<br />

coordinate systems. For the<br />

pelvis, PSIS denotes posteriorsuperior-iliac-spine,<br />

H is hip<br />

center, and RASIS and LASIS<br />

denote right and left anteriorsuperior-iliac-spine<br />

markers,<br />

respectively. For the thigh,<br />

TW is thigh wand, K is knee<br />

center, and MK and LK are<br />

medial and lateral knee<br />

(femoral condyle) markers,<br />

respectively.<br />

Kinematic data can also be combined with ground reaction data<br />

such as forces, torque, and their points of application also<br />

referred to as the centers of pressure. Also, the net joint<br />

reactions including joint forces and moments can be computed by<br />

combining the ground reactions with estimates of segment mass<br />

and mass moments of inertia. Consider the following patient<br />

with a mass of 25.2 kg to demonstrate the calculation of the<br />

reactions at the ankle (Figure 2); data for one instant in the<br />

<strong>gait</strong> cycle are as follows (2a):<br />

Page 8


Force Plates<br />

Figure 2. Ankle A and toe T<br />

marker data are combined with<br />

ground reaction force data Fg<br />

and segment mass and mass<br />

moment of inertia estimates to<br />

compute the net joint forces<br />

and moments.<br />

A force plate is a device that measures the ground reaction<br />

forces exerted by a subject as they step on it during <strong>gait</strong>.<br />

Force plates consist of a top plate, which is mounted level with<br />

the surrounding floor (platform), separated from the bottom<br />

frame by force transducers at each corner; the forces exerted on<br />

the top surface are transmitted through the force transducers.<br />

Force plates allocate the measurement of both the vertical and<br />

shear forces, as well as the center of pressure for the subject<br />

throughout <strong>gait</strong>. At the moment, there are two types of force<br />

plates commercially avai<strong>lab</strong>le: piezoelectric and strain gauge.<br />

The type of force plate employed does not necessarily make a<br />

significant difference for clinical <strong>gait</strong> applications.<br />

Piezoelectric force plates (PEFPs) utilize quartz transducers<br />

that produce an electric charge when stressed. These PEFPs do<br />

not require a power supply, but do however, require special<br />

charge amplifiers and low noise coaxial cables in order to<br />

convert the charge to a voltage proportional to the applied<br />

load. Strain gauge force plates (SGFPs) on the other hand, use<br />

strain gauges to measure the stress in specially machined<br />

aluminum transducer bodies referred to as load cells, when a<br />

subject’s load is applied. Unlike the PEFPs, the SGFPs do not<br />

require special cabling or charge amplifiers, but do require<br />

excitation of the strain gauge bridge circuit. Generally, the<br />

PEFPs are more sensitive and have a greater force range than the<br />

SGFPs.<br />

Page 9


Pressure Mats<br />

A pressure mat is a quick and easy way of obtaining a plantar<br />

pressure picture of a subject as they walk on it during <strong>gait</strong>.<br />

Today, there are two types of transducers in use for plantar<br />

measurement: capacitive and force sensitive resistor (FSR). The<br />

capacitor transducers consist of two capacitor plates separated<br />

by a compressible rubber dielectric material. When a pressure<br />

is applied to the transducer, the capacitor plates are pushed<br />

closer together, resulting in a capacitance that is calibrated<br />

in units of pressure. The FSR transducers, on the other hand,<br />

consist of two thin layers of flexible plastic with printed<br />

circuits on the inner surfaces, separated by thin layer of<br />

double-sided adhesive. When a pressure is applied to the<br />

transducer, carbon on one surface contacts a metal pattern on<br />

the other surface, creating a resistive electrical circuit; as<br />

pressure is applied, the resistance drops. Since the<br />

transducers tend to be nonlinear, the precision of these systems<br />

relies on the ability to consistently calibrate them. Commonly,<br />

pneumatic pressure bladder calibration systems are used to<br />

calibrate the transducers, and because the area of the<br />

transducers is known, the applied force may be determined by<br />

summing the force computed from each active sensor for a<br />

particular instance. Pressure mats provide a valuable and quick<br />

way of revealing the areas of high pressure on the plantar<br />

surface of the foot. The following table (Table 1) compares<br />

three different pressure mat systems that are factory calibrated<br />

and provide software that includes color pressure pictures, <strong>gait</strong><br />

lines, force and pressure versus time, force and pressure/time<br />

integrals, and masks for detailed <strong>analysis</strong> of selected areas of<br />

the foot:<br />

Manufacturer<br />

Musgrave<br />

Systems<br />

Novel<br />

Electronics<br />

(EMED)<br />

Tekscan<br />

(F-Mat)<br />

Sensor<br />

Type<br />

Table 1: Pressure Mat Features<br />

Size (mm)<br />

No. of<br />

Sensors<br />

Sensor<br />

Density<br />

(per cm 2 )<br />

FSR 194 x 394 x 38 2,048 2.7 55.6<br />

Capacitive 225 x 445 x 6 2,016 2 70<br />

FSR 320 x 470 x 6 2,128 1.4 120<br />

Sample<br />

Rate (Hz) Calibration<br />

Dynamic<br />

Force<br />

Static<br />

Pressure<br />

Bladder<br />

Static Force<br />

Special<br />

features<br />

Double Plate<br />

System<br />

Avai<strong>lab</strong>le<br />

Podometry<br />

Software<br />

Provided<br />

Real Time<br />

Display<br />

Page 10


Gait Mats<br />

A <strong>gait</strong> mat is a fairly new system that presents both temporal<br />

and spatial <strong>gait</strong> considerations. Gait mats usually consist of<br />

an array of switches embedded along the entire length of a long<br />

strip of walking surface, like a carpet (Figure 3). As a<br />

subject walks along the mat, the switches close under the feet,<br />

allowing the computer to compute the timing of each switch<br />

closure. The spatial parameters of <strong>gait</strong> can be determined using<br />

the geometry of the mat. The major advantages of <strong>gait</strong> mats are<br />

the step length measurements, removal of any <strong>gait</strong> hindering<br />

attachments, relatively low cost, and portability.<br />

Unfortunately, some disadvantages to using the <strong>gait</strong> mat are the<br />

temporal resolution due to limitations in the scan rate and the<br />

spatial resolution due to the finite size of the switches. The<br />

following table (Table 2) compares two different <strong>gait</strong> mat<br />

systems that both provide an extensive database and have<br />

provisions for editing the raw data file:<br />

Manufacturer Type<br />

CIR Systems<br />

(GAITRite)<br />

EQ, Inc.<br />

(Gait Mat)<br />

Portable<br />

(1)<br />

Transportable<br />

(2)<br />

Active Area<br />

(cm)<br />

Table 2: Gait Mat Features<br />

Thickness<br />

(mm)<br />

Switch<br />

Spacing<br />

(mm)<br />

Temporal<br />

Resolution<br />

(msec)<br />

61 x 3.66 4 12.7 11<br />

61 x 4.17 32 15 10<br />

Special Features/<br />

Considerations<br />

Can handle walking aid<br />

patterns, Computes<br />

FAP score (3)<br />

Needs 32 mm thick<br />

runways at each end<br />

for pre & post walk<br />

area (4)<br />

(1) Can be rolled up and carried in a convenient plastic golf case; (2) Folds into four 99 × 41 × 1 cm pieces that fit in a storage case; (3) The<br />

Functional Ambulation Performance (FAP) score is a single numeric representation of a person's <strong>gait</strong>, based upon temporal and spatial <strong>gait</strong> data<br />

as well as the person's physical measurements; (4) Manufacturer does not provide runways.<br />

a b<br />

Figure 3. The GAITRite portable <strong>gait</strong> <strong>analysis</strong> system. a)<br />

demonstrates the portability of the mat and b) shows a subject<br />

walking along the mat.<br />

Page 11


Electrogoniometers<br />

An electrogoniometer is an electro-mechanical device that spans<br />

the desired measured joint from the proximal and distal body<br />

segments (Figure 4). Electrogoniometers supply an output<br />

voltage comparative to the angular change between the two<br />

attachment surfaces. These devices measure the actual angular<br />

change at the specified joint, under the assumption that the<br />

attachment surfaces move along with the midline of the body<br />

segment. There are two main advantages when using<br />

electrogoniometers for <strong>gait</strong> <strong>analysis</strong>, they are quite inexpensive<br />

and they are easy to operate. Unfortunately, a major<br />

disadvantage is that skin and muscle movement may cause<br />

inaccurate readings in the subject’s angular change.<br />

Figure 4. A Penny & Giles strain<br />

gauge electrogoniometer applied at<br />

the knee. The strain gauge in the<br />

small spring measures the angle<br />

between the plastic endblocks that<br />

are attached to the leg with<br />

double-sided adhesive tape.<br />

Optoelectronic Systems - Stereophotogrammetry<br />

One or more video cameras can be used in a video system in order<br />

to track the bright markers that are placed on the subject along<br />

specific locations. For a passive marker system, the markers<br />

are usually solid shapes covered with retroreflective tape.<br />

Furthermore, for an active marker system, the markers are<br />

generally infrared (IR) light-emitting diodes (LEDs). Both of<br />

these systems keep track of the horizontal and vertical<br />

coordinates from all of the markers from each camera. For a<br />

three-dimensional (3D) system, the computer software program<br />

calculates the 3D coordinates for each of the specified markers<br />

using the two-dimensional (2D) data obtained from two or more<br />

video cameras. When a single camera is used for <strong>gait</strong> <strong>analysis</strong>,<br />

one is assuming that all of the motion is occurring in a single<br />

plane that is perpendicular to the camera axis. This is rarely<br />

the case upon examining a subject’s <strong>gait</strong>, and as a result, 2D<br />

systems are not recommended for use in <strong>gait</strong> <strong>analysis</strong> (2b).<br />

Page 12


3. Budget<br />

The cost of building the project is considerable lower compared<br />

to projects that have no components already avai<strong>lab</strong>le and<br />

required to use. This is due to the fact that we already have<br />

much of the equipment that will be used to build the project,<br />

with the exception of the force measuring device. The only<br />

exception may come from the need to use more advance equipment<br />

to obtain accurate data or the current equipment is found to be<br />

incompatible with other components.<br />

The digital camera and tripod that is used for the current<br />

<strong>biomechanics</strong> <strong>lab</strong> set-up will be considered for use in the new<br />

<strong>lab</strong>. Additionally, the National Instrument equipment already in<br />

stock will be used in the new <strong>lab</strong> and was not previously used.<br />

These devices including the National Instruments PXI-1031 and<br />

the BNC-2100 series connector block.<br />

The LabVIEW® software program is already avai<strong>lab</strong>le on the<br />

computers in the <strong>biomechanics</strong> <strong>lab</strong> and in use for other<br />

<strong>Biomedical</strong> <strong>Engineering</strong> <strong>lab</strong>s as well, so this project will not<br />

require any additional purchase of software as the customer has<br />

requested the use of this product. A complete system bought<br />

through retailers starts at around $22,000.<br />

Item Retail<br />

Digital Camera(s) $350-$10,000 $0<br />

Camera Tripod(s) $275-$825 $0<br />

Est.<br />

Project<br />

Reflective Ball Markers (set of 30 passive) $300 $105<br />

Computer Software and Computer $2,500-$5,000 $0<br />

Force Measurement Device Components $200-$30,000 $150-$350<br />

Electrical Components $100-$500 $50-$150<br />

National Instruments PXI-1031 $1,000 $0<br />

National Instruments BNC-2100 $295 $0<br />

Miscellaneous $100 $145<br />

Total $5,120-$48,295 $405-$750<br />

Page 13


4. Conclusion<br />

The <strong>gait</strong> <strong>analysis</strong> <strong>lab</strong> is important for the Biomechanics class<br />

because the students are able to gain a deeper understanding of<br />

mechanical applications and how <strong>gait</strong> <strong>analysis</strong> is used<br />

clinically. Additionally, they will gain further crossfunctional<br />

engineering skills by creating the LabVIEW® software<br />

program necessary to run the <strong>lab</strong>. The goal of our project is to<br />

build a different <strong>gait</strong> <strong>analysis</strong> <strong>lab</strong> where the students can learn<br />

more about the entire set-up, while keeping the project within a<br />

reasonable budget compared to the cost of a commercially built<br />

<strong>lab</strong>.<br />

The design of our project is considered to be unique for two<br />

reasons. First it has an added feature that the current <strong>lab</strong><br />

does not, which will provide the students with more realistic<br />

results. Missing from the current <strong>lab</strong>, a force measurement<br />

device will be introduced that will require the students to<br />

determine what forces are applied on the plate and how the plate<br />

operates. The force measurement device will give accurate force<br />

measurements, which were previously assumed. Second, as part of<br />

the new <strong>lab</strong>, students will be able to gain more knowledge of<br />

<strong>gait</strong> <strong>analysis</strong> by learning how to use the software as well as the<br />

hardware aspects of the <strong>lab</strong>.<br />

Comparing the new <strong>lab</strong> with the original <strong>lab</strong>, where students<br />

usually collect the data and use a program to display the<br />

results, in the new <strong>lab</strong>, creation of the LabVIEW® software<br />

program will provide a more challenging and educational<br />

environment that allows the students to increase their knowledge<br />

of writing programs and applying the program to a real<br />

application. As with the current <strong>lab</strong>, motion devices will be<br />

used during the <strong>lab</strong>, so that the students learn how the images<br />

are captured and used to obtain useful results. Finally<br />

students will gain more knowledge about the hardware used<br />

through setting up the <strong>lab</strong> and connecting all of the devices and<br />

assuring that it works properly.<br />

In the market, the cost of a <strong>gait</strong> <strong>analysis</strong> <strong>lab</strong> with all the<br />

equipment needed starts at around twenty-two thousand United<br />

States dollars, whereas our project will provide a <strong>lab</strong> that<br />

should fall under 5 percent of market cost. Our most costly<br />

expense is expected to come from the plate measurement device,<br />

which will also be produced considerably less than market value.<br />

Page 14


Overall, this project will provide the added components that<br />

have been delayed due to the time constraints of Dr. Enderle’s<br />

growing department. The final results will provide all of the<br />

elements required to make a more robust and successful<br />

educational experience for the students.<br />

References and Bibliography for Section 1.3:<br />

(1a) Instrumented Gait Analysis Systems by Ernest L. Bontrager,<br />

MS.<br />

www.<strong>lab</strong>oratorium.dist.unige.it/~piero/Teaching/Gait/B<br />

NTRAER%20Instrumented%20Gait%20Analysis%20Systems.htm<br />

(1b) www.toolsandsupplies.com/shimstockinfo.htm<br />

(1c) Tekscan, Inc.<br />

www.tekscan.com<br />

www.<strong>lab</strong>oratorium.dist.unige.it/~piero/Teaching/Gait/BONTRAG<br />

ER%20Instrumented%20Gait%20Analysis%20Systems.htm, was used<br />

as a source for the information provided in this section,<br />

with direct quotes noted in the text and with the exception<br />

of reference to (1b).<br />

Mentioned companies can be found at:<br />

www.teckscan.com<br />

www.bertec.com<br />

www.amti.biz<br />

www.kistler.com<br />

www.biometricsltd.com<br />

www.vicon.com<br />

References and Bibliography for Section 2.2:<br />

(2a) Enderle, J. D.(2000). Introduction to <strong>Biomedical</strong><br />

<strong>Engineering</strong>. San Diego, CA. P. 451-457.<br />

(2b) Instrumented Gait Analysis Systems by Ernest L. Bontrager,<br />

MS.<br />

www.<strong>lab</strong>oratorium.dist.unige.it/~piero/Teaching/Gait/BO<br />

NTRAGER%20Instrumented%20Gait%20Analysis%20Systems.htm<br />

Page 15

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!