Specifications 1<br />

CONDUCTIVITY *<br />

TEMPERATURE<br />

PRESSURE<br />

Technical Specifications<br />

<strong>NEO</strong>-2 Meter <strong>User</strong>’s Guide 10<br />

RANGE RESOLUTION ACCURACY<br />

100.0 to 199.9 mS<br />

20.0 to 99.9 mS<br />

2.00 to 19.99 mS<br />

1.0 mS<br />

0.1 mS<br />

0.01mS<br />

± 8% of reading<br />

± 2.0 mS<br />

± 0.1 mS<br />

.020 to 1.999 mS<br />

0.001 mS<br />

± 0.01 mS<br />

10.0 o to 90.0 o C 0.1º C 10.0 o to 40.0 o C: ±0.2 o C<br />

41.0 o to 90.0 o C: ±1.0 o C<br />

Gauge:<br />

-700 to +1000 mmHg<br />

Differential:<br />

-1000 to +1500 mmHg<br />

1 mmHg<br />

1 mmHg<br />

± 5 mmHg<br />

± 1% of reading<br />

pH 1.0 to 12.0 pH units 0.1 pH unit ± 0.1 pH unit<br />

*Temperature-compensation: 15° to 45°C.<br />

Physical Specifications<br />

DIMENSIONS WEIGHT POWER BATTERY TYPE<br />

H 4.85” x W 4.1” x D 3.4”<br />

12.3 x 10.4 x 8.6 cm<br />

1 Specifications are subject to change without notice.<br />

1.16 lbs./525 g Battery 9V Alkaline