MI-42-0005 Rev 8<br />
3.8.1 Full Body/Neonate Body/Helical<br />
Indications:<br />
Helical mode for the combination of bone and soft tissue studies, e.g. masses, trauma, disorders of the joint etc. A<br />
range of 150 mm will be covered in 7.5 sec.<br />
Full Body/Neonate BodyHelical<br />
Newborn<br />
kV 80<br />
mA 7<br />
Effective mAs 9.5<br />
Rotation Time<br />
1.92 s<br />
Slice Collimation 1.06 mm<br />
Slice Width<br />
1.0 mm<br />
Feed/Rotation<br />
48 mm<br />
Pitch Factor 1.415<br />
Kernel<br />
Standard<br />
CTDIvol<br />
0.5 mGy<br />
Weight Adapted No<br />
Weight<br />
3 kg<br />
Page 37 of 38