Analysis: Overall uncertainties<br />
Radionuclid<br />
e<br />
<strong>TDCR</strong> <strong>CIEMAT</strong>/<strong>NIST</strong> E β,max in<br />
keV<br />
u(a)/a in %<br />
90 Y 0.12 0.16 2280<br />
32 P 0.23 0.25 1711<br />
89 Sr 0.25 0.26 1495<br />
99 Tc 0.27 0.45 294<br />
45 Ca 0.25 0.27 256<br />
33 P 0.24 0.30 249<br />
147 Pm 0.35 0.35 225<br />
35 S 0.33 0.29 167<br />
63 Ni 0.97 0.58 67<br />
O. Nähle <strong>and</strong> K. Kossert <strong>Comparison</strong> <strong>of</strong> <strong>TDCR</strong> <strong>and</strong> CN <strong>for</strong> β emitters