13.11.2020 Views

catalog2020v16-update

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

RP<br />

Internal Hex - Regular Platform 3.75mm<br />

APPROVED<br />

DRILLING PROTOCOL<br />

Hex 2.42<br />

4.2<br />

RP<br />

L<br />

Drill sequence by bone type - An additional 1 mm must<br />

be added to the length of the drill to account for the angled<br />

cutting tip.<br />

D-I, D-II Dense bone considerations - Countersink<br />

should be used if the standard dense bone protocol is not<br />

sufficient to fully seat the implant without exceeding the<br />

recommended maximum insertion torque of max. 60Ncm<br />

for implant Ø3.75mm and above. Do not exceed max.<br />

45Ncm for implants diameter lower than 3.75mm.<br />

SP4206<br />

L 6.0mm<br />

SP4208<br />

L 8.0mm<br />

SP4210<br />

L 10.0mm<br />

SP4211<br />

L 11.5mm<br />

SP4213<br />

L 13.0mm<br />

SP4216<br />

L 16.0mm<br />

Soft Bone<br />

Dense Bone<br />

Drilling<br />

Speed - RPM<br />

1200-<br />

1500<br />

900-<br />

1,200<br />

700-<br />

900<br />

500-<br />

700<br />

400-<br />

600<br />

200-<br />

400<br />

Diameter<br />

Ø 1.9 Ø 2.0 Ø 2.8 Ø 3.2 Ø 3.65 Ø 3.5-4.2<br />

Optional<br />

*The procedure recommended by GP Implant LTD cannot replace the judgment and the experience of the surgeon.<br />