DRAFT 2/24/05<br />

2.2 Shutdown<br />

2-2<br />

<strong>VHS</strong>-4, <strong>VHS</strong>-6 <strong>and</strong> <strong>VHS</strong>-<strong>250</strong> <strong>Diffusion</strong> <strong>Pumps</strong><br />

5. Monitor inlet <strong>and</strong> foreline pressures.<br />

During operation of the diffusion pump:<br />

❑ The gas load at the inlet should not exceed the maximum throughput capability of<br />

the pump.<br />

❑ The forepressure should not exceed the specified tolerable forepressure.<br />

1. Turn off the power to the diffusion pump <strong>and</strong> continue to back the diffusion pump with<br />

the appropriate mechanical pump.<br />

2. Allow cooling water to flow through the diffusion pump until the pump body<br />

temperature, located just above the boiler plate, has cooled to a temperature of<br />

approximately 130 °F.<br />

After isolating the backing pump, the diffusion pump can be vented to atmosphere.<br />

If faster cooling is desired, the pump can be cooled using the Quick Cool Coil at the<br />

bottom of the diffusion pump.