DRAFT 2/24/05<br />
2.2 Shutdown<br />
2-2<br />
<strong>VHS</strong>-4, <strong>VHS</strong>-6 <strong>and</strong> <strong>VHS</strong>-<strong>250</strong> <strong>Diffusion</strong> <strong>Pumps</strong><br />
5. Monitor inlet <strong>and</strong> foreline pressures.<br />
During operation of the diffusion pump:<br />
❑ The gas load at the inlet should not exceed the maximum throughput capability of<br />
the pump.<br />
❑ The forepressure should not exceed the specified tolerable forepressure.<br />
1. Turn off the power to the diffusion pump <strong>and</strong> continue to back the diffusion pump with<br />
the appropriate mechanical pump.<br />
2. Allow cooling water to flow through the diffusion pump until the pump body<br />
temperature, located just above the boiler plate, has cooled to a temperature of<br />
approximately 130 °F.<br />
After isolating the backing pump, the diffusion pump can be vented to atmosphere.<br />
If faster cooling is desired, the pump can be cooled using the Quick Cool Coil at the<br />
bottom of the diffusion pump.