Section V Model 3300A<br />
5-14. Repeat 5-13 on CHANNEL B. Limit should<br />
remain > 15 volts peak-to-peak.<br />
5-15. Repeat 5-13 and 5-14 except load the<br />
instrument with the 50 ohm resistor. CHANNEL A and<br />
CHANNEL B voltage output should be > 2 volts peak-topeak.<br />
5-16. SQUARE WAVE RESPONSE.<br />
a. Test equipment required: Oscilloscope (-hp-<br />
Model 175A/1750B) and 10:1 Probe (-hp-<br />
Model 10001A).<br />
b. Connect CHANNEL A OUTPUT without a<br />
load to the oscilloscope using the 10:1<br />
Probe, and set the 3300A controls as<br />
follows:<br />
CHANNEL A function .................SQUARE<br />
FREQUENCY dial ......................10<br />
RANGE switch............................X10K<br />
c. Verify: Rise and fall time