6.1.6 Isocenter Height (nominal): 129.5 cm<br />

6.2 Collimator<br />

6.2.1 Extended Rotation Range: ±165°<br />

6.2.2 Mechanical Isocenter Accuracy:<br />

≤0.05 cm radius from isocenter<br />

6.2.3 Position Indicators (gantry and console)<br />

6.2.3.1 Digital Readouts:<br />

• Accuracy: ±0.5°<br />

• Resolution: 0.1°<br />

6.2.3.2 Mechanical Scales:<br />

• Accuracy: ±1.0°<br />

• Resolution: 1.0°<br />

6.3 Field Size Collimation<br />

6.3.1 Range: The field size is continuously<br />

variable from 0.5 x 0.5 cm2 to 40 x<br />

40 cm2 as measured at 100 cm TSD.<br />

Field sizes larger than 35 x 35 cm2 are<br />

limited to a 49.5 cm diagonal (the<br />

diameter of the circle defined by the<br />

primary collimator at 100 cm TSD).<br />

The field size is defined as the distance<br />

along the radial and transverse axes<br />

between the points of 50% density on<br />

an X-ray film taken at 100 cm TSD with<br />

minimum buildup. The SRS 6 MV<br />

beam field size is limited to a maximum<br />

of 15 x 15 cm2 .<br />

6.3.2 Position Indicators<br />

• Accuracy: ±0.2 cm<br />

• Resolution: 0.1 cm<br />

6.3.3 Light and X-ray Field Coincidence: The<br />

field-defining light coincides to within<br />

1.5 mm of the 50% isodensity line on<br />

an X-ray film. This is defined at 100 cm<br />

TSD with minimum buildup for any<br />

field size.<br />

6.4 Exact ® Couch with Indexed Immobilization 6.4.1 The Exact Couch is standard with<br />

<strong>Trilogy</strong> Tx. (<strong>Specification</strong>s and standard<br />

vs. optional accessories for the Exact<br />

Couch are provided on Exact Couch<br />

specification sheet, RAD 1951.)<br />

page 9<br />

6.4.2 Motion Controls<br />

• Two Hand Pendants that control the<br />

accelerator and Exact Couch<br />

• Two Couch Side Panels<br />

• Remote Couch Control<br />

6.4.3 Position Indicators<br />

• Corrective motions: small translations<br />

(in x, y, and z) and small rotation of<br />

the couch to fine tune patient setups<br />

• Planned motions: large rotations of the<br />

couch to sequence between noncoplanar<br />

fields and arcs<br />

6.4.3.1 Translation<br />

• Accuracy: ±0.1 cm<br />

• Resolution: 0.1 cm<br />

6.4.3.2 Rotation<br />

• Accuracy: ±0.5°<br />

• Resolution: 0.1°<br />

6.5 Compact stand assembly<br />

6.5.1 Single access and through-door viewing<br />

of all gas and water system status<br />

indicators<br />

6.5.2 Imager electronics (PortalVision MV<br />

imaging system and On-Board Imager kV imaging system) incorporated in<br />

reduced height stand<br />

7.0 Mechanical Accessories Options<br />

The following optional accessories are available with<br />

<strong>Trilogy</strong> Tx and the Exact Couch:<br />

7.1 Collimator Accessories:<br />

• Interface Mount<br />

• Accessory Mount<br />

• Port Film Graticule<br />

• 4-Way Wedge Set (four wedges 15°, 30°,<br />

45°, 60°)<br />

• Optional Electron Applicators: A set of<br />

five electron applicators can be provided, with<br />

selection from 6 sizes: 6 x 6 cm2 , 6 x 10 cm2 ,<br />

10 x 10 cm2 , 15 x 15 cm2 , 20 x 20 cm2 , and<br />

25 x 25 cm2 .<br />