Field audit regional <strong>GAW</strong> station Jungfraujoch, Switzerland, January 1999 16 / 31<br />

station station MOHp<br />

Jungfraujoch<br />

TEI 49C-PS<br />

transfer TEI 49C standard #306<br />

analog signals<br />

Figure 7: Experimental set up, ozone<br />

analog signals<br />

TEI 49C-PS<br />

transfer standard<br />

<strong>Station</strong> Ozone<br />

Generator<br />

Module<br />

<strong>Station</strong> MOHp-data-acquisition-facility<br />

data-acquisiton facility<br />

EMPA MOHp<br />

ADC<br />

UPK TEI 49C 8002<br />

RS-232 analog connection signals<br />

interface ADC<br />

final data<br />

PC<br />

EMPA-data <strong>WCC</strong> data- acquisition<br />

facility acquisiton facility<br />

PC<br />

data review<br />

data processing *<br />

The <strong>WCC</strong> acquisition system, which was used for the audit, consists of a 16 channel ADC circuit<br />

board and a PC with the corresponding software. Hooked up to the analog output of all the<br />

instruments involved the data was collected by the data acquisition system of the <strong>WCC</strong>. In<br />

advance, it was checked that the readings of the two acquisition systems were equal at zero<br />

(ozone) and at 200 ppb ozone. For data interpretation the <strong>WCC</strong> data is used.<br />

WMO Global Atmosphere Watch World <strong>Calibration</strong> Centre for Surface Ozone and Carbon Monoxide