Field audit regional <strong>GAW</strong> station Jungfraujoch, Switzerland, January 1999 16 / 31<br />
station station MOHp<br />
Jungfraujoch<br />
TEI 49C-PS<br />
transfer TEI 49C standard #306<br />
analog signals<br />
Figure 7: Experimental set up, ozone<br />
analog signals<br />
TEI 49C-PS<br />
transfer standard<br />
<strong>Station</strong> Ozone<br />
Generator<br />
Module<br />
<strong>Station</strong> MOHp-data-acquisition-facility<br />
data-acquisiton facility<br />
EMPA MOHp<br />
ADC<br />
UPK TEI 49C 8002<br />
RS-232 analog connection signals<br />
interface ADC<br />
final data<br />
PC<br />
EMPA-data <strong>WCC</strong> data- acquisition<br />
facility acquisiton facility<br />
PC<br />
data review<br />
data processing *<br />
The <strong>WCC</strong> acquisition system, which was used for the audit, consists of a 16 channel ADC circuit<br />
board and a PC with the corresponding software. Hooked up to the analog output of all the<br />
instruments involved the data was collected by the data acquisition system of the <strong>WCC</strong>. In<br />
advance, it was checked that the readings of the two acquisition systems were equal at zero<br />
(ozone) and at 200 ppb ozone. For data interpretation the <strong>WCC</strong> data is used.<br />
WMO Global Atmosphere Watch World <strong>Calibration</strong> Centre for Surface Ozone and Carbon Monoxide