GAW - Station Transfer Calibration LAB WCC - Empa
GAW - Station Transfer Calibration LAB WCC - Empa
GAW - Station Transfer Calibration LAB WCC - Empa
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Field audit regional <strong>GAW</strong> station Jungfraujoch, Switzerland, January 1999 26 / 31<br />
Comment<br />
The results of the CO intercomparison measurements (figure 14) deviate only by about 1 to 3%<br />
from the conventional true value with a measurement uncertainty of approximately 2%. Regarding<br />
the relevant range (99 and 193 ppb, 5- and 95-percentile respectivley) and the methode in<br />
operation this is a very good result.<br />
WMO Global Atmosphere Watch World <strong>Calibration</strong> Centre for Surface Ozone and Carbon Monoxide