Field audit regional <strong>GAW</strong> station Jungfraujoch, Switzerland, January 1999 26 / 31<br />

Comment<br />

The results of the CO intercomparison measurements (figure 14) deviate only by about 1 to 3%<br />

from the conventional true value with a measurement uncertainty of approximately 2%. Regarding<br />

the relevant range (99 and 193 ppb, 5- and 95-percentile respectivley) and the methode in<br />

operation this is a very good result.<br />

WMO Global Atmosphere Watch World <strong>Calibration</strong> Centre for Surface Ozone and Carbon Monoxide