Field audit regional <strong>GAW</strong> station Jungfraujoch, Switzerland, January 1999 28 / 31<br />

approx. concentration levels: 30 / 60 / 90 / 125 / 185 ppb<br />

sequence of concentration: random<br />

averaging interval per concentration: 5 minutes<br />

number of runs: 3 before and 3 after audit<br />

zero air supply: Pressurised air - zero air generator (CO catalyst,<br />

Purafil, charcoal)<br />

ozone generator: SRP's internal generator<br />

data acquisition system: SRP's ADC and acquisition<br />

ZERO AIR<br />

GENERATOR<br />

OZONE-<br />

GENERATOR<br />

Figure 16: Instruments set up SRP -TEI 49C-PS<br />

SRP#15<br />

2 cells (approx. 90 cm)<br />

TEI 49C-PS<br />

The stability of the transfer standard is thoroughly examined with respect to the uncertainties of the<br />

different components (systematic error and precision). For the <strong>GAW</strong> transfer standard of the <strong>WCC</strong>-<br />

O3 (TEI 49C-PS) the assessment criteria, taking into account the uncertainty of the SRP, are<br />

defined to (1 ppb + 0.7%).<br />

Figures 17 and 18 show the resulting linear regression and the corresponding 95% prediction<br />

interval for the comparisons of TEI 49C-PS vs. SRP#15. Clearly, the linear regression and the<br />

prediction interval remain within the recommended tolerance.<br />