<strong>Oxi</strong> <strong>330i</strong>/<strong>340i</strong><br />

Technical data<br />

Accuracy<br />

(± 1 digit)<br />

mg/l<br />

Concentration<br />

%<br />

Saturation<br />

mbar<br />

Partial pressure<br />

Temperature<br />

compensation<br />

± 0.5 % of the measured value<br />

at an ambient temperature of 5 °C ... 30 °C<br />

± 0.5 % of the measured value<br />

when measuring in the range of<br />

± 10 K around the calibration temperature<br />

± 0.5 % of the measured value<br />

at an ambient temperature of 5 °C ... 30 °C<br />

< 2 % at 0 ... 40 °C<br />

T [°C] ± 0.1<br />

Correction<br />

functions<br />

Salinity correction<br />

Air pressure correction<br />

0 ... 70.0 SAL<br />

Automatic through installation of<br />

pressure sensor in the range<br />

500 ... 1100 mbar<br />

Analog output<br />

(only Cond <strong>340i</strong>)<br />

Automatic switchover when the recorder is connected AK 323.<br />

Voltage<br />

Resolution<br />

Saturation<br />

0 ... 200.0 %<br />

0 ... 600 %<br />

Concentration<br />

0...20.00 mg/l<br />

0...100.0 mg/l<br />

0 ... 2 V<br />

0 ... 600 mV<br />

0...2 V<br />

0...1 V<br />

0.1 %<br />

1 %<br />

0.01 mg/l<br />

0.1 mg/l<br />

Accuracy<br />

Internal resistance<br />

± 0.5 % of display value<br />

± 0.1(% saturation)<br />

or<br />

± 0.5 % of the display value ± 0.01 mg/l<br />

< 5 Ohm (current limited to max. 0.2 mA<br />

output current)<br />