General Catalog - emitec-industrial.ch
General Catalog - emitec-industrial.ch
General Catalog - emitec-industrial.ch
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Laboratory Grade Instruments<br />
Model 4028 Power Sensor<br />
BIRD ® Model 4028<br />
High Power, Precision Power Sensors for Precision Process<br />
Applications<br />
A<strong>ch</strong>ieves tighter, more consistent RF power measurements<br />
for improved yields.<br />
• ± 2% at calibration frequencies and power levels.<br />
± 4% at other frequencies and power levels. Add 2% to<br />
uncertainty outside 25 ± 10 °C.<br />
• Direct, plug-in operation with the industry standard<br />
Bird ® Model 4421 RF Power Meter.<br />
• Multiple sensors provide for measurement over a 250 kHz to<br />
30 MHz frequency range<br />
• Capable of power measurements up to 50kW.<br />
With the expanding surface areas in the LCD and Solar panel industry, higher<br />
power levels are necessary for the industries processes. The Bird 4028 series<br />
sensor is designed to meet the growing RF power measurement <strong>ch</strong>allenge with<br />
accuracy, repeatability and versatility to meet current and future applications. Bird<br />
continues to innovate useful sensor te<strong>ch</strong>nology to meet these fast paced markets.<br />
4028 SERIES — Power Sensor<br />
Model Power Range Frequency<br />
4028A250K 1.0 kW-20 kW 250-400 kHz<br />
4028A400K 1.0 kW-20 kW 400-550 kHz<br />
4028A2M 1.0 kW-25 kW 1.5-2.5 MHz<br />
4028A3M 1.0 kW-25 kW 2.5-3.5 MHz<br />
4028A4M 1.0 kW-25 kW 3.5-4.5 MHz<br />
4028A10M 1.0 kW-25 kW 10-15 MHz<br />
4028A25M 1.0 kW-25 kW 25-30 MHz<br />
4028B3M 1.0 kW-25 kW 2.5-4 MHz<br />
4028B10M 1.0 kW-25 kW 10-15 MHz<br />
4028C10M 500 W-50 kW 10-15 MHz<br />
POWER MEASUREMENT<br />
Accuracy ±2% (1�) at calibration frequencies and power levels;<br />
±4% at other frequencies and power levels. Add 2% to<br />
uncertainty outside 25 ± 10°C<br />
Calibration Power Level 250 and 400kHz units: 1.7kW<br />
VSWR Range 1.0 to 2.0 (40.00 to 9.5 dB return loss)<br />
Directivity 28 dB<br />
Insertion Loss < 0.05dB<br />
Uniformity 2% maximum unit to unit, at calibration frequency<br />
and power levels.<br />
Speed 2 readings per second.<br />
Maximum Power 4028AxxK: 20kW<br />
4028AxxM: 25kW<br />
4028B: 25kW<br />
4028C: 50kW<br />
CONNECTORS<br />
4028A Series Customer Specified<br />
4028B Series 1-5/8 EIA Flanged<br />
4028C Series 3-1/8 EIA Flanged<br />
Sensor Interface Lat<strong>ch</strong>-n-Lock<br />
POWER REQUIREMENTS<br />
External DC 12 VDC, supplied from Bird 4421 Power Meter<br />
15 Bird ® Te<strong>ch</strong>nologies Group 1-866-695-4569 www.bird-te<strong>ch</strong>nologies.com<br />
PHYSICAL SPECIFICATIONS<br />
Dimensions 4.7" L x 3.2" W x 4.0" H (4028A Series)<br />
6.8" L x 3.5" W x 4.8" H (4028B Series)<br />
8.0" L x 5.2" W x 6.4" H (4028C Series)<br />
Weight 3.3 lbs. (4028A Series)<br />
5.2 lbs. (4028B Series)<br />
7.3 lbs. (4028C Series))<br />
ENVIRONMENTAL SPECIFICATIONS<br />
Operating Temperature 0°C to 50°C (32°F to 122°F)<br />
(derate accuracy outside 25 ± 10°C)<br />
Storage Temperature -20°C to 70°C (-4°F to 158°F)<br />
Humidity 95% non-condensing<br />
Altitude 10,000 feet (3,000 m)<br />
<strong>General</strong> EMC Designed to carry CE mark (with immunity<br />
exception noted below)<br />
Emissions EN-55011, 1991, Class B<br />
Immunity EN-50082-1, 1995<br />
Safety EN-61010, 1993 in accordance with Council<br />
Directives 73/23/EEC and 93/68/EEC<br />
Calibration Cycle 6 months