Operating<br />
important to mention that the disturbing effect is reduced, if the direction of the<br />
spurious AC field is perpendicular to the sensor optical axis. It has been<br />
experimentally determined that in this case the tested magnetometer operated<br />
satisfactory in presence of the disturbing field with magnitude 3500nT peak at<br />
the frequency 3kHz.<br />
The operating range of the tumble angle, in the plane perpendicular to the direction<br />
of the spurious field, was reduced from 10°-85° (without the AC field) to 25°-65° (with<br />
the AC field).<br />