Final View: Stability Blocking<br />
Screws<br />
Once stability blocking screw insertion is<br />
complete, remove the Blocking Screw<br />
Attachment and Anterior Drop from the<br />
Drill Guide and obtain both AP and lateral<br />
radiographic images to confirm accurate<br />
placement.<br />
TRIGEN <strong>Nail</strong> Cap and <strong>Nail</strong> Cap<br />
Set Screw Insertion: Optional<br />
Remove the Drill Guide/Anterior Drop assembly.<br />
Attach the selected <strong>Nail</strong> Cap or <strong>Nail</strong> Cap Set<br />
Screw to the Medium Hexdriver/T-Handle<br />
assembly and insert into the end of the nail<br />
until tight.<br />
Note: The TRIGEN <strong>Nail</strong> Cap does not engage<br />
with the most distal locking screws to create a<br />
fixed angle construct.<br />
Note: If cross-threading occurs, rotate the <strong>Nail</strong><br />
Cap or <strong>Nail</strong> Cap Set Screw counterclockwise<br />
until its threads line up with those of the nail.<br />
Proceed with insertion until tight.<br />