Q.<strong>raxx</strong> <strong>A107</strong>-<strong>16</strong> <strong>slimline</strong> <strong>RS</strong> / <strong>EC</strong> Universal Measurement System<br />
The Q.<strong>raxx</strong> <strong>slimline</strong> product is based on the standardized 19” technology, one rack unit (1 U) and is designed for measurements with a<br />
high level of flexibility, reliability and accuracy in the field of stationary testing and assembly.<br />
The <strong>EC</strong> version includes a Test Controller that offers the user a powerful solution with PAC functionality, synchronized data acquisition,<br />
sequencing, mathematics, combinations a Ethernet TCP/IP interface as well as a EtherCAT fieldbus. It is possible to connect three<br />
<strong>slimline</strong> basic units to the Test Controller of a <strong>slimline</strong> <strong>EC</strong> unit.<br />
Beside the pre-defined standard versions customized systems regarding numbers of channels, kind of measurement inputs/connectable<br />
sensors as well as the required connectors are configurable.<br />
This modularity permits to design an optimized individual solution for any application. Further it is possible to mix different product lines<br />
like Q.<strong>raxx</strong>, Q.bloxx within Q.series.<br />
Most important features:<br />
<strong>16</strong> universal analog input channels<br />
voltage, current, resistance, potentiometer, Pt100,<br />
Pt1000, thermocouples, measuring bridges<br />
Fast high accuracy digitalization<br />
24 bit ADC, 10 kHz sample rate per channel<br />
Signal conditioning<br />
Linearization, digital filter, average, scaling, min/max storage,<br />
RMS, arithmetic, alarm<br />
Galvanic isolation<br />
Kanal II Kanal II Versorgung II Schnittstelle, Viso 500 VDC<br />
Galvanic isolation<br />
of analog in/out, power supply and interface, Viso 500 VDC<br />
Power supply 10…30 VDC<br />
With embedded Test Controller (version <strong>EC</strong>):<br />
Optional fieldbus interface EtherCAT<br />
EtherCAT according specification ETG,<br />
256 read and 256 write variable with 10 kHz<br />
Ethernet interface for configuration and data output<br />
FTP, TCP/IP, UDP<br />
FTP Server and FTP Client functionality<br />
configurable function<br />
High data rate over Ethernet<br />
<strong>16</strong> real variables with 10 kHz (block transfer)<br />
64 real variables with 300 Hz (online<br />
Data buffer memory dyn. <strong>16</strong> MByte (RAM), stat. 128 MByte (flash)<br />
data buffer at block transfer of measurements<br />
Connection of additional racks to the Test Controller<br />
up to 3 <strong>slimline</strong> racks without Test Controller can be connected to a<br />
<strong>slimline</strong> <strong>EC</strong> system with embedded controller<br />
Gantner Instruments Test & Measurement GmbH www.gantner-instruments.com office@gantner-instruments.com<br />
Silvrettastraße 13 6780 Schruns / Austria T +43 (0) 5556 77463-0 F +43 (0) 5556 77463-300<br />
Heidelberger Landstraße 74 64297 Darmstadt / Germany T +49 (0) 6151 95136-0 F +49 (0) 6151 95136-26
Q.<strong>raxx</strong> <strong>A107</strong>-<strong>16</strong> <strong>slimline</strong> <strong>RS</strong> / <strong>EC</strong> Universal Measurement System<br />
Connection Diagram Q.<strong>raxx</strong> <strong>slimline</strong> <strong>EC</strong><br />
Analog Inputs<br />
Number <strong>16</strong><br />
Accuracy 0.01 % typical<br />
0.02 % in controlled environment¹<br />
0.05 % in industrial area²<br />
Linearity error 0.01 % of the final value typical<br />
Repeatability 0.003 % typical (within 24 h)<br />
Isolation voltage 500 VDC channel to channel to power supply to interface 3<br />
Measurement Voltage Range max. Deviation Resolution<br />
Input resistance >100 MΩ<br />
±10 V ±2 mV 1.2 µV<br />
±1 V ±0.2 mV 120 nV<br />
±100 mV ±20 µV 12 nV<br />
Long term drift
Q.<strong>raxx</strong> <strong>A107</strong>-<strong>16</strong> <strong>slimline</strong> <strong>RS</strong> / <strong>EC</strong> Universal Measurement System<br />
Measurement Current<br />
(internal shunt 50 Ω)<br />
Range max. Deviation Resolution<br />
±25 mA ±5 µA 3.0 nA<br />
Long term drift
Q.<strong>raxx</strong> <strong>A107</strong>-<strong>16</strong> <strong>slimline</strong> <strong>RS</strong> / <strong>EC</strong> Universal Measurement System<br />
Analog/Digital-Conversion<br />
Digital Inputs (<strong>slimline</strong> <strong>EC</strong> only)<br />
Resolution 24 bit<br />
Sample rate 10 kHz, (measurement thermocouple 10 Hz)<br />
Conversion method Sigma-Delta (group delay time 600 µs)<br />
Anti-aliasing filter 2 kHz, 5 th order<br />
Digital filter IIR, low pass, high pass, band pass, 4 th order, 1 Hz up to 1 kHz in steps 1, 2, 5<br />
Averaging configurable or automated according the selected data rate<br />
Function fixed definition<br />
Input voltage max. 30 VDC<br />
Input current max. 1.5 mA<br />
Upper switching threshold >3.5 V (high)<br />
Lower switching threshold
Q.<strong>raxx</strong> <strong>A107</strong>-<strong>16</strong> <strong>slimline</strong> <strong>RS</strong> / <strong>EC</strong> Universal Measurement System<br />
Slave Interfaces <strong>RS</strong> 485 (<strong>slimline</strong> <strong>EC</strong> only) to connect further <strong>slimline</strong> <strong>RS</strong>, see Connection Diagram<br />
Number of interfaces 3<br />
Standard <strong>RS</strong> 485<br />
Data format 8E1<br />
Protocol Local Bus<br />
Baud rate 9.6 kbps up to 24 Mbps<br />
Connectable devices max. <strong>16</strong> modules at one UART line<br />
Isolation voltage 500 V<br />
Data Memory (<strong>slimline</strong> <strong>EC</strong> only)<br />
RAM <strong>16</strong> MByte (optional 90 MByte), cycle buffer<br />
Flash 128 MByte<br />
Synchronization of a Multi Device System (<strong>slimline</strong> <strong>EC</strong> only)<br />
Power Supply<br />
Mechanical<br />
Interface <strong>RS</strong>485 Standard<br />
Mode Master Slave principle, IRIG standard DCF77, AFNOR etc, GPS over IRIG standard<br />
GPS NMEA over <strong>RS</strong>232 SNTP over Ethernet<br />
Power supply 10 bis 30 VDC, Überspannungs- und Verpolungsschutz<br />
Power consumption <strong>slimline</strong> <strong>RS</strong> approx. 10 W<br />
Power consumption <strong>slimline</strong> <strong>EC</strong> approx. 15 W<br />
Environmental<br />
Influence of the voltage
Q.<strong>raxx</strong> <strong>A107</strong>-<strong>16</strong> <strong>slimline</strong> <strong>RS</strong> / <strong>EC</strong> Universal Measurement System<br />
Programming Tool test.con<br />
Using test.con for programming of the PAC-function in a graphical way:<br />
Warm Up Time<br />
All declarations are valid after a warm up time of 45 minutes.<br />
Valid from January 2011. Specification subject to change without notice<br />
DB_Q.<strong>raxx</strong>_<strong>A107</strong>-<strong>16</strong>SL_E_20.docx<br />
Gantner Instruments Test & Measurement GmbH www.gantner-instruments.com office@gantner-instruments.com<br />
Silvrettastraße 13 6780 Schruns / Austria T +43 (0) 5556 77463-0 F +43 (0) 5556 77463-300<br />
Heidelberger Landstraße 74 64297 Darmstadt / Germany T +49 (0) 6151 95136-0 F +49 (0) 6151 95136-26