Retractable <strong>housing</strong>s <strong>InTrac</strong> <strong>796</strong><br />
<strong>InTrac</strong> <strong>796</strong> <strong>–</strong> <strong>the</strong> <strong>original</strong> <strong>retractable</strong> <strong>housing</strong> <strong>for</strong><br />
sterilization of electrodes in biotech processes.<br />
<strong>InTrac</strong> <strong>796</strong> is designed to meet <strong>the</strong><br />
specific application requirements of<br />
biotechnology processes.<br />
The <strong>InTrac</strong> <strong>796</strong> is a state-of-<strong>the</strong>art<br />
<strong>retractable</strong> <strong>housing</strong> specifically<br />
designed <strong>for</strong> on-line measuring<br />
applications in <strong>the</strong> biotechnology<br />
industry.<br />
It enables safe insertion and withdrawal<br />
of sensors during an ongoing<br />
process without compromising<br />
sterility.<br />
<strong>–</strong> Twin chamber lock enables easy<br />
electrode maintenance or replacement<br />
and steam-sterilization<br />
in-situ, without process interruption<br />
or compromise on sterility<br />
<strong>–</strong> Enhanced safety in combination<br />
with safety weld-in socket<br />
<strong>–</strong> Side-mounting in bioreactors and<br />
vessels through 25 mm standard<br />
(Ingold) or safety weld-in sockets<br />
<strong>–</strong> Suited <strong>for</strong> high-precision, liquidelectrolyte<br />
ph and redox electrodes<br />
<strong>–</strong> Versions <strong>for</strong> manual or pneumatic<br />
operation<br />
<strong>–</strong> Pressure resistant up to 6 bar<br />
(85 psi), allowing application in<br />
large-scale industrial fermenters<br />
and process vessels
∅ 68<br />
∅ 75<br />
∅ 50<br />
R1 1/4"<br />
∅ 25<br />
43<br />
H = 75<br />
1. «OUT» position approx. 575<br />
2. «IN» position approx. 527<br />
3. approx. 565<br />
a = 250 1<br />
2 3<br />
INTERNET<br />
http://www.mtpro.com<br />
ISO<br />
9001<br />
certified<br />
ISO<br />
14001<br />
certified<br />
Management System<br />
certified according to<br />
ISO 9001 / ISO 14001<br />
Retractable <strong>housing</strong>s <strong>InTrac</strong> <strong>796</strong><br />
Retractable <strong>housing</strong> <strong>for</strong> use in biotech.<br />
The electrode can be sterilized inside <strong>the</strong> <strong>housing</strong> when in retracted position, using <strong>the</strong><br />
controlled environment of <strong>the</strong> twin-chamber lock. This double lock effectively protects <strong>the</strong><br />
process from any intrusion of contamination after insertion of <strong>the</strong> electrode. The <strong>housing</strong><br />
also enables <strong>the</strong> electrode to be withdrawn from <strong>the</strong> process <strong>for</strong> maintenance purposes<br />
under sterile conditions while <strong>the</strong> process is still running. After servicing or exchange, <strong>the</strong><br />
electrode can be steam-sterilized within <strong>the</strong> <strong>housing</strong> be<strong>for</strong>e reinsertion into <strong>the</strong> process<br />
medium, without any fear of contamination.<br />
The <strong>InTrac</strong> <strong>796</strong> is easily installed through <strong>the</strong> side of <strong>the</strong> reactor or vessel, using a 25 mm<br />
weld-in socket (Ingold type or METTLER TOLEDO safety socket). It is available ei<strong>the</strong>r as a<br />
manually operated version or with pneumatic drive.<br />
Specifications<br />
Pressure in reactor:<br />
Temperature ranges<br />
<strong>InTrac</strong> <strong>796</strong>-M/75, <strong>InTrac</strong> <strong>796</strong>-P/75<br />
dependant on electrode specification, but not more <strong>the</strong>n 6 bar (85 psi)<br />
Retractable member: 0...130°C (32...266°F), steam-sterilizable<br />
Headpart: 0...80°C (32...176°F)<br />
Weld-in socket: ID = 25 mm, L = 40 mm, 15° chamfer<br />
Insertion length:<br />
Materials<br />
H = 75 mm<br />
Process wetted: Stainless steel, DIN 1.4435 with 3.1b material certificate<br />
Headpart: POM (polyoxymethylene) and PP (polypropylene)<br />
Seals process wetted: Viton ® FDA<br />
Flushing connections: G 1 ⁄8 " thread<br />
Ordering In<strong>for</strong>mation<br />
Designation Order no.<br />
<strong>InTrac</strong> <strong>796</strong>-M/75 (manual) 00 <strong>796</strong> 3002<br />
<strong>InTrac</strong> <strong>796</strong>-P/75 (pneumatic) 00 <strong>796</strong> 3010<br />
(Fitting sets <strong>for</strong> steam lines see «spare parts and accessories»)<br />
Standard-Electrodes<br />
Designation Order no.<br />
pH electrode: InPro 2000/250/Pt100/9848 52 001 436<br />
pH electrode: InPro 2000/250/Pt1000/9848 52 001 437<br />
pH electrode: 465-50-SC-P-S7/250/9848 10 465 4503<br />
Redox electrode: Pt 4865-50-SC-S7/250/9848 10 565 3140<br />
For <strong>the</strong> <strong>retractable</strong> installation of low maintenance gel-filled electrodes and 12 mm 02 or turbidity sensors,<br />
we recommend <strong>the</strong> <strong>InTrac</strong> 797 <strong>housing</strong>.<br />
Spare parts and accessories<br />
Description Order no.<br />
Service kit, small 00 <strong>796</strong> 2003<br />
Service kit, large 00 <strong>796</strong> 2002<br />
Weld-in socket, inclined 15°, L = 40 mm, stainless steel DIN 1.4435 00 764 1014<br />
Safety weld-in socket, inclined 15°, L = 40 mm, stainless steel DIN 1.4435 52 400 462<br />
Fitting set <strong>for</strong> <strong>InTrac</strong> <strong>796</strong>/797 Serto, Material: 1.4571 00 797 2006<br />
Fitting set <strong>for</strong> <strong>InTrac</strong> <strong>796</strong>/797 Gyrolok, Material: 1.4401 00 797 2008<br />
Fitting set <strong>for</strong> <strong>InTrac</strong> <strong>796</strong>/797 Swagelok, Material: 1.4401 00 797 2007<br />
Conversion kit, manual to pneumatic operation 00 777 2004<br />
Gasket 20 305 1004<br />
Temperature sensor Pt100-764/5m 10 100 3104<br />
The <strong>retractable</strong> <strong>housing</strong> <strong>InTrac</strong> <strong>796</strong> toge<strong>the</strong>r with <strong>the</strong> appropriate electrode and a transmitter<br />
from METTLER TOLEDO offer a complete system <strong>for</strong> <strong>the</strong> most accurate and reproducible<br />
measurement of pH and redox values in biotechnology.<br />
<strong>Mettler</strong>-<strong>Toledo</strong> GmbH, Process Analytics,<br />
Industrie Nord, CH-8902 Urdorf,<br />
Phone +41-1-736 22 11, Fax +41-1-736 26 36<br />
Subject to technical changes.<br />
© <strong>Mettler</strong>-<strong>Toledo</strong> GmbH 02/01 (Version 1.0)<br />
Printed in Switzerland. 52 400 084<br />
Sales and service: