Basic Sound Chamber Tests 95<br />

If there is a sequence of tests that you would like to perform frequently, you can<br />

save it into the <strong>7000</strong> Test System’s permanent memory using the SAVE GROUP<br />

function. Three groups of up to 10 curves can be saved into SAVE GROUP 1, 2,<br />

and 3, respectively. The settings for both ears are saved when SAVE GROUP is<br />

used, so it is possible to program one set of settings for the left ear and other set<br />

from the right ear.<br />

When you want to manually load a group that has previously been saved, you<br />

use the LOAD GROUP function. It is also possible to have the <strong>7000</strong> analyzer<br />

automatically load a saved group and put itself in Fully-Auto or Semi-Auto<br />

mode when the analyzer is first turned on or the [RESET] button is used. This<br />

is described in Section 3.8.2. When the "Factory" group is loaded, the factory<br />

default settings are loaded.<br />

3.8.1 Setting up a test sequence<br />

The first step in running your custom automated test sequence is deciding on<br />

the source types and levels of the frequency response curves. When finished,<br />

these selections can be stored into the analyzer’s permanent memory.<br />

1. Use [F4] to delete ALL CURVES & SETTINGS in the Coupler screen. (Press<br />

[F4]. Use [∨, ∧] to select ALL CURVES & SETTINGS. Use [>] to complete the<br />

selection and close the pop-up menu.)<br />

2. Use [F6] to select PROGRAM. This will open a pop-up window with helpful<br />

programming instructions.<br />

3. Use [F1] to select the ear.<br />

4. Use [F2] to select the curve number.<br />

5. Use [F5] to select the desired source type.<br />

6. Use [∨, ∧] to select the desired source level.<br />

7. Press [MENU] and use the arrow keys to change other menu settings, if<br />

desired. Press [EXIT] to return to the coupler screen.<br />

8. Save and advance to the next curve. This can be done two ways:<br />

• Press [START] if you want the analyzer to be programmed to automatically<br />

advance to the next curve and start measuring it.<br />

• Press [STOP] if you want the analyzer to be programmed to advance to<br />

the next curve, but then pause and wait for user intervention before taking<br />

the next measurement. Use this if you want to make an adjustment<br />

to the aid before taking the next measurement. A horizontal line will be<br />

drawn across the Curve Characteristics box when a pause is inserted in<br />

this fashion.<br />

9. Repeat steps 5-8 to select the desired source type and level for all the curves<br />

that you want to measure in the automated test sequence.