32 FONIX <strong>7000</strong> Hearing Aid Analyzer<br />

RESET FREQ: The frequency the analyzer returns to when RESET is pressed.<br />

AVG FREQS: The frequencies used for averaging in pure-tone measurements.<br />

Each frequency set is represented on the screen by the highest frequency in the<br />

set. The sets are:<br />

HFA (High Frequency Average) 1000, 1600, 2500<br />

SPA (Special Purpose Average) 800, 1250, 2000<br />

SPA 1250, 2000, 3150<br />

SPA 1600, 2500, 4000<br />

SPA 2000, 3150, 5000<br />

IEC–(HAIC) 500, 1000, 2000<br />

Coupler Default Settings<br />

BATTERY MEAS: Status of the battery current measurement. Choose ON or OFF.<br />

BATTERY SIZE: Size of battery used with the hearing aid being tested. Used in<br />

battery current measurements.<br />

NOISE RED (TONE): Amount of pure-tone noise reduction used in coupler measurements.<br />

See Section 2.3.1.1.<br />

NOISE RED (COMP): Amount of composite noise reduction used in coupler<br />

measurements. See Section 2.3.2.1.<br />

RESET SOURCE: The signal type selected in the Coupler screen when the<br />

[RESET] button is pushed.<br />

DISPLAY: Display of the Coupler Multicurve screen. Choose between Gain and<br />

SPL.<br />

CURVE LABEL: Label used for the two available “ears.” Choose between LEFT/<br />

RIGHT, and A/B.<br />

GRAPH SCALING: Scaling used when making measurements. NORMAL scales<br />

to the upper-most curve displayed. AUTO scales to the current selected curve.<br />

COUPLER TYPE: The type of coupler being used in making sound chamber<br />

measurements. Selections of CIC and MZ turn on corresponding software correction<br />

factors (see Sections 3.6 and 3.7). A selection of NONE or 2cc does not<br />

apply correction factors to the curve.<br />

STATIC TONE: Status of individual frequency measurement. Choose SINGLE<br />

to emit a static tone in between composite measurements or pure-tone sweeps.<br />

Choose AVG to emit a three frequency average between other measurements.<br />

Choose OFF to turn off this measurement.<br />

DISTORTION: Type of harmonic distortion used in a pure-tone measurement.<br />

Choose between 2nd, 3rd, and TOTAL. See Section 2.3.1.3.