32 FONIX <strong>7000</strong> Hearing Aid Analyzer<br />
RESET FREQ: The frequency the analyzer returns to when RESET is pressed.<br />
AVG FREQS: The frequencies used for averaging in pure-tone measurements.<br />
Each frequency set is represented on the screen by the highest frequency in the<br />
set. The sets are:<br />
HFA (High Frequency Average) 1000, 1600, 2500<br />
SPA (Special Purpose Average) 800, 1250, 2000<br />
SPA 1250, 2000, 3150<br />
SPA 1600, 2500, 4000<br />
SPA 2000, 3150, 5000<br />
IEC–(HAIC) 500, 1000, 2000<br />
Coupler Default Settings<br />
BATTERY MEAS: Status of the battery current measurement. Choose ON or OFF.<br />
BATTERY SIZE: Size of battery used with the hearing aid being tested. Used in<br />
battery current measurements.<br />
NOISE RED (TONE): Amount of pure-tone noise reduction used in coupler measurements.<br />
See Section 2.3.1.1.<br />
NOISE RED (COMP): Amount of composite noise reduction used in coupler<br />
measurements. See Section 2.3.2.1.<br />
RESET SOURCE: The signal type selected in the Coupler screen when the<br />
[RESET] button is pushed.<br />
DISPLAY: Display of the Coupler Multicurve screen. Choose between Gain and<br />
SPL.<br />
CURVE LABEL: Label used for the two available “ears.” Choose between LEFT/<br />
RIGHT, and A/B.<br />
GRAPH SCALING: Scaling used when making measurements. NORMAL scales<br />
to the upper-most curve displayed. AUTO scales to the current selected curve.<br />
COUPLER TYPE: The type of coupler being used in making sound chamber<br />
measurements. Selections of CIC and MZ turn on corresponding software correction<br />
factors (see Sections 3.6 and 3.7). A selection of NONE or 2cc does not<br />
apply correction factors to the curve.<br />
STATIC TONE: Status of individual frequency measurement. Choose SINGLE<br />
to emit a static tone in between composite measurements or pure-tone sweeps.<br />
Choose AVG to emit a three frequency average between other measurements.<br />
Choose OFF to turn off this measurement.<br />
DISTORTION: Type of harmonic distortion used in a pure-tone measurement.<br />
Choose between 2nd, 3rd, and TOTAL. See Section 2.3.1.3.