PWR Power On/Off<br />
Command PWR/x1 Switch <strong>Halogen</strong> <strong>Moisture</strong> Analyzer On or Off<br />
x1: 0 Set <strong>Halogen</strong> <strong>Moisture</strong> Analyzer to<br />
standby mode<br />
1 Switch <strong>Halogen</strong> <strong>Moisture</strong> Analyzer on<br />
Response PWR/A A<br />
<strong>Halogen</strong> <strong>Moisture</strong> Analyzer has been switched off<br />
successfully<br />
PWR/A A<br />
<strong>Halogen</strong> <strong>Moisture</strong> Analyzer with the serial number<br />
I4/A/"text" according to text has been switched on succesfully<br />
(see also I4 command)<br />
PWR/L Command understood, parameter wrong, (number,<br />
value range, …)<br />
Comments<br />
• In the standby mode, the interface remains active; but all commands except PWR, HA07<br />
and HA20 are answered with EL.<br />
• On switching on, the <strong>Halogen</strong> <strong>Moisture</strong> Analyzer also sends the serial number (see also<br />
I4 command).<br />
• On switching off, all current commands are terminated (see also @ command).<br />
P100 Print out text on the strip printer<br />
Command P100/"x1" Print out "text" on the internal printer<br />
Response P100/A A Command executed<br />
P100/I Command can not be executed at present as there is<br />
no printer or the printer buffer is full<br />
P100/L Command understood, parameter wrong, (value<br />
range, …)<br />
Example<br />
Command P100/"HALLO" Print out HALLO on the strip printer<br />
Response P100/A A<br />
Printout has been started<br />
26 METTLER TOLEDO <strong>MT</strong>-<strong>SICS</strong> HR83/HG63 <strong>Reference</strong> <strong>Manual</strong>