HA420 Activate / deactivate methods option<br />

Command HA420 Inquiry of methods option<br />

Response HA420/A/x1 x1: Set methods option<br />

Command HA420/x1 Set methods option<br />

x1: Methods<br />

0 off<br />

1 on, factory setting<br />

Response HA420/A A Methods option set<br />

HA420/L Command understood, parameter wrong, (number,<br />

value range, …)<br />

Comments<br />

• Setting the menu parameters terminates a drying.<br />

• The command HA420 is not possible with HR83.<br />

HA421 Activate / deactivate statistics and journal function<br />

Command HA421 Inquiry of statistics and journal function<br />

Response HA421/A/x1 x1: Set statistics and journal function (see<br />

below)<br />

Command HA421/x1 Set statistics and journal function<br />

x1: 0 off<br />

1 on, factory setting<br />

Response HA421/A A Statistic and journal function activated<br />

HA421/L Command understood, parameter wrong, (number,<br />

value range, …)<br />

Comments<br />

• Setting the menu parameters terminates a drying.<br />

• Switching off statistics and journal deletes all the statistical and journal entries without<br />

warning!<br />

48 METTLER TOLEDO <strong>MT</strong>-<strong>SICS</strong> HR83/HG63 <strong>Reference</strong> <strong>Manual</strong>