HA24 Inquiry of temperature<br />

Command HA24 Inquiry of current temperature<br />

Response HA24/A/x1 x1: Current temperature in °C<br />

Example<br />

Command HA24 Inquiry of current temperature<br />

Response HA24/A/105 The temperature is 105 °C.<br />

HA25 Inquiry of drying weights<br />

Command HA25 Inquiry of drying weight of the last or current drying<br />

Response HA25/A/x1/x2/x3/x4<br />

x1: Drying status<br />

0 = No drying exists<br />

1 = Drying running<br />

2 = Drying ended<br />

3 = Drying terminated<br />

x2: Wet weight in grams<br />

x3: Current weight or dry weight in grams<br />

x4: Drying time (seconds)<br />

Example 1<br />

Command HA25 Inquiry of drying weights<br />

Response HA25/A/2/12.345/7.890/180<br />

Drying has been ended regularly, wet weight<br />

12.345 g, dry weight 7.890 g,<br />

drying time 180 seconds<br />

Example 2<br />

Command HA25 Inquiry of drying weights<br />

Response HA25/A/0/0.000/0.000/0<br />

No drying exists, e.g. as the battery was discharged<br />

Comment<br />

Together with the command HA07 – Report instrument status change – dryings can be<br />

shown in parallel on the host.<br />