HA01 Reset application / escape<br />

Command HA01 Reset application / escape<br />

Response HA01/A A Application reset<br />

Comments<br />

This command has the same effect as the Reset key, see operating instructions of the<br />

<strong>Halogen</strong> <strong>Moisture</strong> Analyzer. It terminates all current commands and activities.<br />

HA02 Set factory settings<br />

Command HA02 Set factory setting of the menu and method<br />

parameters<br />

Response HA02/A A Menu and method parameters set to factory setting<br />

Comment<br />

This command terminates a drying.<br />

HA03 Switch keypad on/off<br />

Command HA03/x1 x1: 0 Keypad of <strong>Halogen</strong> <strong>Moisture</strong> Analyzer<br />

switched off<br />

1 Keypad of <strong>Halogen</strong> <strong>Moisture</strong> Analyzer<br />

switched on (factory setting)<br />

Response HA03/A A Command executed<br />

HA03/L Command understood, parameter wrong, (number,<br />

value range, …)<br />

Comments<br />

This command has no effect on the On/Off key. The user can thus switch the instrument on<br />