<strong>Technical</strong> <strong>Parameters</strong> <strong>of</strong> <strong>RY</strong><strong>IIB+</strong> <strong>Anesthesia</strong> Workstation<br />
<strong>RY</strong>-<strong>IIB+</strong> is a new anesthesia workstation with modern computer control technology. It is made from<br />
engineering plastics for its main body and shaped from one mold and therefore appropriate for<br />
modern anesthesia operation. It brings high cost-effectiveness to customers with its novel<br />
appearance, convenient connection, excellent performance and fair price. Customers are allowed to<br />
configure the HJ-2001 CO2 patient monitor if needed.<br />
<strong>Anesthesia</strong> Machine<br />
Gas supply: supplying gas via central tube or gas cylinder with rated pressure <strong>of</strong> 0.4MPa<br />
fluctuating between 0.32MPa-0.60MPa.<br />
Flowmeter: four-tube flowmeter with thin and thick tubes, equipped with O2/N2O auto interlock<br />
and N2O auto cut<strong>of</strong>f(when O2 is insufficient) device. Its flow ranges from 0.1L/min to 10L/min.<br />
Low O2 supply alarm and auto cut-<strong>of</strong>f <strong>of</strong> N2O: triggering over 7s pneumatic sound alarm and<br />
cutting <strong>of</strong>f N2O supply when O2 supply pressure is too low.<br />
O2 flush flow: 35L/min-75L/min.<br />
Integrated <strong>Anesthesia</strong> Ventilation System<br />
With integrated breathing circuit, the system is compact in structure, convenient in installation and<br />
easy in connection. Full close, half close and half open are available to choose. Two CO2 absorbent<br />
canisters can be used all day long without replacing the sodalime. A sodalime quick-replace device<br />
is provided, with the flowmeter equipped inside able to monitor the airway pressure in both manual<br />
and mechanical modes.
Control <strong>Parameters</strong>:<br />
Exhaust valve: 0.5kPa-7.5kPa.<br />
Leakage: not greater than 150mL/min (3kPa).<br />
Inspiratory and expiratory impedance: not greater than 0.6kPa (60mL/min).<br />
Size <strong>of</strong> inspiratory and expiratory valve connectors: outside diameter <strong>of</strong> 22mm and 1:40 degree<br />
<strong>of</strong> taper.<br />
High-precision Anesthetic Agent Vaporizer<br />
The high-precision anesthetic agent vaporizer is capable <strong>of</strong> automatically compensating the flow,<br />
temperature and pressure. The vaporizing concentration <strong>of</strong> anesthetic agent can be set up precisely.<br />
The system is equipped with one vaporizer for standard configuration, or two according to<br />
customer's demands. The interlock device is employed to ensure that the two vaporizers will not be<br />
opened simultaneously. Customers are allowed to configure ENF vaporizer, ISO vaporizer and SEV<br />
vaporizer, and also import vaporizers.<br />
Concentration range: 0vol%-5vol%, flow range: 0.25L/min-15L/min, temperature range: 15℃-35℃.<br />
<strong>Anesthesia</strong> Ventilator<br />
The new anesthesia ventilator, with a 5.1’’ high-definition LCD, is computer-controlled and easily<br />
configured through a coder. It displays multiple respiration parameters and waveforms. Its friendly<br />
interface and Chinese menus enable anesthesia doctors to operate more conveniently and safely.<br />
Ventilation mode: IPPV and MANU. Applicable to adult and pediatric patients.<br />
Operating mode: pneumatically and electrically controlled.<br />
Control parameters:<br />
Tidal volume: at least 50ml-1,500ml; inspiratory flow: 3L/min-70L/min.<br />
Frequency: 3-99 times/min; airway pressure high alarm limit: 1kPa-9.9kPa.<br />
Expiration-inspiration ratio: 3:1-1:9; airway pressure low alarm limit: at least 0.4kPa-4kPa,<br />
with delay time <strong>of</strong> 10s-15s.<br />
Alarm silence: no greater than 120s; safety exhaust valve: ≤12.5kPa.<br />
Monitoring <strong>Parameters</strong>:<br />
Expiratory tidal volume (TVe), minute ventilation (MV), frequency, peak pressure, mean pressure,<br />
minimum pressure, O2 concentration(optional), expiratory flow waveform, airway pressure<br />
waveform and multi-parameter 24-hour graphic trends.<br />
Alarm <strong>Parameters</strong>:<br />
Airway pressure high/low limits alarm, minute ventilation (MV) alarm, O2 concentration high/low<br />
limits alarm, sustained high airway pressure alarm, apnea alarm, ventilation system integrity alarm,<br />
power failure alarm, gas supply failure alarm, etc.
Others<br />
Power supply: ~220V/50Hz; pressure range: 0.32MPa-0.60MPa.<br />
Gas supplies: O2 and N2O.<br />
This product is configured with a large-capacity built-in power supply to keep running for 3h when<br />