Error Messages<br />

Message Cause and effect<br />

SIGNAL X Encoder signal is too small,<br />

e.g. when an encoder is<br />

contaminated.<br />

PROB. ERROR Before touching off on the<br />

workpiece, the tool must move<br />

by a distance of at least 0.2 mm.<br />

ERR. REF. X The spacing of the reference<br />

marks as defined in P43 is not<br />

the same as the actual spacing.<br />

FRQ. ERR. X The input frequency for this<br />

encoder input is too high. This<br />

can occur when the scale is<br />

moved too fast.<br />

ERR. MEMORY Check sum error: Check the<br />

datum, operating parameters and<br />

compensation values for multipoint<br />

axis error compensation.<br />

If the error recurs, contact your<br />

service agency!<br />

To erase error messages:<br />

After you have removed the cause of error:<br />

➤ Press the CL key.<br />

Error Messages<br />

29