10<br />

Preparing to Run <strong>UPLC</strong>; Injection and Detection Paramters<br />

Pressure Assist Injection Sequence<br />

Step 3 – Position the sample<br />

Summary of Partial Loop<br />

using Pressure Assist<br />

Shorter Cycle Time than PLUNO<br />

No sample cushion<br />

Sample volume is conserved<br />

Recommended for large sample loop injections<br />

Performance dependency on weak wash solvent matching to mobile<br />

phase<br />

Air Gaps injected onto Column<br />

Accuracy is generally lower compared to Needle Overfill<br />

Has lower injection range, within a given loop for partial loop<br />

injections. Injection range is from 10 – 50% of the loop volume<br />

Accuracy and Precision are lower than Full Loop Mode<br />

Dr. Shula Levin, Medtechnica<br />

Pressure Assist Injection Sequence<br />

Step 4 – Injection valve injects sample<br />

Sample Loop contains<br />

Sample, air gaps and weak wash<br />

Partial Loop Using Needle Overfill