29.04.2014 Views

Analytical Measurement - Nuova Elva

Analytical Measurement - Nuova Elva

Analytical Measurement - Nuova Elva

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

JUMO GmbH & Co. KG<br />

Delivery address:Mackenrodtstraße 14,<br />

36039 Fulda, Germany<br />

Postal address: 36035 Fulda, Germany<br />

Phone: +49 661 60 03-0<br />

Fax: +49 661 60 03-6 07<br />

E-mail: mail@jumo.net<br />

Internet: www.jumo.net<br />

JUMO Instrument Co. Ltd.<br />

JUMO House<br />

Temple Bank, Riverway<br />

Harlow, Essex CM20 2TT, UK<br />

Phone: +44 12 79 63 55 33<br />

Fax: +44 12 79 63 52 62<br />

E-mail: sales@jumo.co.uk<br />

JUMO PROCESS CONTROL INC.<br />

885 Fox Chase, Suite 103<br />

Coatesville PA 19320, USA<br />

Phone: 610-380-8002<br />

1-800-554-JUMO<br />

Fax: 610-380-8009<br />

E-mail: info@JumoUSA.com<br />

Internet: www.JumoUSA.com<br />

Data Sheet 20.2820<br />

Page 1/7<br />

Immersion fittings<br />

Series 202820<br />

Brief description<br />

Immersion fittings are used for holding electrochemical sensors (e.g. pH and redox electrodes,<br />

glass conductivity cells, compensation thermometers etc.) with a Pg13.5 screw-in<br />

thread and a fitting length of 120mm.<br />

Fitting types for up to 3 sensors are available.<br />

The fittings are mounted in open sluices or containers. They protect the installed sensor<br />

from breaking and enable measurement in different immersion depths. Thanks to various<br />

options and accessories, the fittings can be optimally adjusted to the conditions on site.<br />

Two pipe clips for wall mounting are provided for the standard versions, but sliding flanges,<br />

which are available as an option, also permit installation in container lids, for example.<br />

The following points have to be taken into account:<br />

■ The fittings must be readily accessible, to ensure that the sensor can be cleaned and<br />

serviced at regular intervals.<br />

■ pH and redox electrodes must not be allowed to remain dry for a prolonged period -<br />

this can be prevented by using a wetting cup.<br />

■ The suitability of the materials (e.g. chemical compatibility) has to be tested by the<br />

system designer.<br />

202820/063...<br />

202820/040...<br />

Installation options<br />

Standard<br />

Use of a wetting cup<br />

Fixing<br />

with pipe clips<br />

in the inlet channel<br />

Fixing<br />

with a flange<br />

on a container<br />

with the container filled<br />

with empty container<br />

(residual liquid in the cup<br />

prevents the sensor from<br />

running dry)<br />

07.05/00388381