AIR FLOW<br />
Where:<br />
Q = Volumetric Flow Rate<br />
P1 = Static pressure at the inlet<br />
P2 = Static pressure at the outlet<br />
r = Radius of the restriction<br />
h = (eta) absolute viscosity<br />
of the fluid<br />
L = Length of the restriction<br />
Since p, r and L are constant;<br />
Equation 1 can be rewritten as:<br />
Q = K (∆P/h) (Equation 2)<br />
Where K is a constant factor determined<br />
by the geometry of the restriction.<br />
Equation 2 shows the linear relationship<br />
between volumetric flow rate (Q)<br />
differential pressure (DP) and absolute<br />
viscosity (h) in a simpler form.<br />
All the flow meters are equipped with<br />
female NPT inlet and outlet port<br />
connections. They are position<br />
insensitive and may be mounted in any<br />
orientation that is convenient. The inlet<br />
and outlet ports are equal in size and<br />
symmetric<br />
(in-line). Normally speaking, the port<br />
sizes and dimensions for differing flow<br />
ranges are as follows:<br />
Specifications<br />
FMA-1600/FVL-1600<br />
Accuracy: ±1% full scale for air<br />
Accuracy of Gas Select Feature:<br />
2% full scale<br />
Repeatability ±0.5% full scale<br />
Response Time* (Typical): 10 ms<br />
Output: 0-5 Vdc standard<br />
Pressure Drop** (Typical):<br />
0.4 psid full Scale<br />
Operating Temperature<br />
-10 to 50°C (14 to 122°F)<br />
Zero Shift: 0.02% full scale/°C<br />
Span Shift: 0.02% full scale/°C<br />
Humidity Range: 0 - 100% noncondensing<br />
Supply Current***<br />
(Typical): 0.25 A to 1 A<br />
Excess Flow Rate: 20X full scale<br />
Common Mode<br />
Pressure (max): 125 psig<br />
Supply Voltage:<br />
Turbulent Flow<br />
7 to 30 Vdc<br />
Cable Connection:<br />
8 pin mini DIN<br />
Zero Shift:<br />
0.02% Full scale/°C<br />
Span Shift:<br />
0.02% Full scale/°C<br />
Humidity Range:<br />
0 - 100% non-condensing<br />
Supply Current***<br />
(Typical): 0.25 A to 1 A<br />
Excess Flow Rate: 20X Full scale<br />
Common Mode Pressure (max):<br />
125 psig<br />
Supply Voltage: 7 to 30 Vdc<br />
Cable Connection: 8 pin mini DIN<br />
Inlet<br />
Inlet<br />
Inlet<br />
Differential Pressure Sensor<br />
Laminar Flow<br />
Laminar Flow<br />
Laminar Flow<br />
Cavity Plate<br />
Outlet<br />
Laminar Plate<br />
Spacer<br />
Outlet<br />
Laminar Plate<br />
Spacer<br />
Outlet<br />
*Very low flow devices may have special ports to reduce dead volume.<br />
LAMINAR FLOW ELEMENT<br />
Turbulent Fl<br />
Dimensions<br />
Flow Range Height Length Depth Port Size<br />
20(S) LPM and under* 4.082" 2.375" 1.05" 1/8"NPT(F)<br />
50 (S) LPM 4.207" 2.625" 1.05" 1/4"NPT(F)<br />
100 (S) LPM 4.367" 3.5" 1.6" 1/4"NPT(F)<br />
250 to 1000 (S) LPM 5.242" 3.5" 1.6" 1/2"NPT(F)<br />