AIR FLOW<br />

Where:<br />

Q = Volumetric Flow Rate<br />

P1 = Static pressure at the inlet<br />

P2 = Static pressure at the outlet<br />

r = Radius of the restriction<br />

h = (eta) absolute viscosity<br />

of the fluid<br />

L = Length of the restriction<br />

Since p, r and L are constant;<br />

Equation 1 can be rewritten as:<br />

Q = K (∆P/h) (Equation 2)<br />

Where K is a constant factor determined<br />

by the geometry of the restriction.<br />

Equation 2 shows the linear relationship<br />

between volumetric flow rate (Q)<br />

differential pressure (DP) and absolute<br />

viscosity (h) in a simpler form.<br />

All the flow meters are equipped with<br />

female NPT inlet and outlet port<br />

connections. They are position<br />

insensitive and may be mounted in any<br />

orientation that is convenient. The inlet<br />

and outlet ports are equal in size and<br />

symmetric<br />

(in-line). Normally speaking, the port<br />

sizes and dimensions for differing flow<br />

ranges are as follows:<br />

Specifications<br />

FMA-1600/FVL-1600<br />

Accuracy: ±1% full scale for air<br />

Accuracy of Gas Select Feature:<br />

2% full scale<br />

Repeatability ±0.5% full scale<br />

Response Time* (Typical): 10 ms<br />

Output: 0-5 Vdc standard<br />

Pressure Drop** (Typical):<br />

0.4 psid full Scale<br />

Operating Temperature<br />

-10 to 50°C (14 to 122°F)<br />

Zero Shift: 0.02% full scale/°C<br />

Span Shift: 0.02% full scale/°C<br />

Humidity Range: 0 - 100% noncondensing<br />

Supply Current***<br />

(Typical): 0.25 A to 1 A<br />

Excess Flow Rate: 20X full scale<br />

Common Mode<br />

Pressure (max): 125 psig<br />

Supply Voltage:<br />

Turbulent Flow<br />

7 to 30 Vdc<br />

Cable Connection:<br />

8 pin mini DIN<br />

Zero Shift:<br />

0.02% Full scale/°C<br />

Span Shift:<br />

0.02% Full scale/°C<br />

Humidity Range:<br />

0 - 100% non-condensing<br />

Supply Current***<br />

(Typical): 0.25 A to 1 A<br />

Excess Flow Rate: 20X Full scale<br />

Common Mode Pressure (max):<br />

125 psig<br />

Supply Voltage: 7 to 30 Vdc<br />

Cable Connection: 8 pin mini DIN<br />

Inlet<br />

Inlet<br />

Inlet<br />

Differential Pressure Sensor<br />

Laminar Flow<br />

Laminar Flow<br />

Laminar Flow<br />

Cavity Plate<br />

Outlet<br />

Laminar Plate<br />

Spacer<br />

Outlet<br />

Laminar Plate<br />

Spacer<br />

Outlet<br />

*Very low flow devices may have special ports to reduce dead volume.<br />

LAMINAR FLOW ELEMENT<br />

Turbulent Fl<br />

Dimensions<br />

Flow Range Height Length Depth Port Size<br />

20(S) LPM and under* 4.082" 2.375" 1.05" 1/8"NPT(F)<br />

50 (S) LPM 4.207" 2.625" 1.05" 1/4"NPT(F)<br />

100 (S) LPM 4.367" 3.5" 1.6" 1/4"NPT(F)<br />

250 to 1000 (S) LPM 5.242" 3.5" 1.6" 1/2"NPT(F)<br />

Order Online<br />

omega.com ®<br />

Over 100 000 Products Available!<br />

For the complete selection of Air Flow products,<br />

refer to The Green Book ®<br />