26.06.2014 Views

Vulkan Lokring MGLD.3000 Product Manual - Schoonover, Inc.

Vulkan Lokring MGLD.3000 Product Manual - Schoonover, Inc.

Vulkan Lokring MGLD.3000 Product Manual - Schoonover, Inc.

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>MGLD.3000</strong><br />

OPERATING INSTRUCTIONS<br />

01 <strong>MGLD.3000</strong>


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Please read these operating instructions carefully and thoroughly before installation, commissioning,<br />

and use and follow all operating- and safety instructions.<br />

Please…<br />

do not hesitate to contact us<br />

in case of questions<br />

or technical problems<br />

Bitte…<br />

wenden Sie sich bei Fragen oder<br />

technischen Problemen an uns<br />

BAL_MGLD3000_20101001_en.doc 2-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Table of contents<br />

1. Introduction..................................................................................................................................... 4<br />

2. Safety instructions ......................................................................................................................... 5<br />

2.1. Symbols, pictograms, and icons used.............................................................................................. 5<br />

2.1.1. Warning symbols and instruction symbols in the text of these operating instructions ..................... 5<br />

2.1.2. Pictograms and icons used in screen menus................................................................................... 5<br />

2.1.3. Pictograms/probe control elements used ......................................................................................... 8<br />

2.2. General safety instructions............................................................................................................... 8<br />

3. Description of the device............................................................................................................. 11<br />

3.1. General information ........................................................................................................................ 11<br />

3.2. Device components ........................................................................................................................ 11<br />

3.2.1. Base unit......................................................................................................................................... 11<br />

3.2.2. Docking unit.................................................................................................................................... 12<br />

3.2.3. Probes ............................................................................................................................................ 13<br />

3.2.4. Electro-pneumatic connection line ................................................................................................. 14<br />

3.3. Technical data ................................................................................................................................ 14<br />

3.3.1. <strong>MGLD.3000</strong> - Base unit .................................................................................................................. 14<br />

3.3.2. <strong>MGLD.3000</strong> docking unit ................................................................................................................ 15<br />

3.3.3. Probes for the <strong>MGLD.3000</strong>............................................................................................................. 16<br />

4. Installation and commissioning.................................................................................................. 17<br />

4.1. Unpacking device from packing materials...................................................................................... 17<br />

4.2. Installation and commissioning the <strong>MGLD.3000</strong>............................................................................ 17<br />

4.2.1. Stationary operation of the device with docking unit/charging batteries on the base unit ............. 17<br />

4.2.2. Connecting the probe (stationary and mobile operation of the <strong>MGLD.3000</strong>)................................. 18<br />

4.3. Operating the <strong>MGLD.3000</strong> ............................................................................................................. 21<br />

4.3.1. Basic information on the graphical user navigation........................................................................ 21<br />

4.3.2. Initial startup of the <strong>MGLD.3000</strong>..................................................................................................... 22<br />

4.3.3. Leak search mode .......................................................................................................................... 24<br />

4.3.4. Check + calibration......................................................................................................................... 29<br />

4.3.5. Settings........................................................................................................................................... 36<br />

4.3.6. Printing out and saving the log ....................................................................................................... 42<br />

4.3.7. Service............................................................................................................................................ 46<br />

4.4. Shutting off the device .................................................................................................................... 49<br />

4.5. Warning and error messages ......................................................................................................... 50<br />

5. Care and service........................................................................................................................... 53<br />

5.1. Cleaning and care .......................................................................................................................... 53<br />

5.2. Service............................................................................................................................................ 53<br />

5.3. Warranty Limitations....................................................................................................................... 54<br />

BAL_MGLD3000_20101001_en.doc 3-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

1. Introduction<br />

The Multi Gas Leak Detector <strong>MGLD.3000</strong> is designed for use by qualified personnel while closely following<br />

these operating instructions.<br />

These operating instructions provide information to users with proper training in gas leak detection on the<br />

design and function of the Multi Gas Leak Detector and also act as instructions for handling the device<br />

and its components.<br />

These operating instructions describe the condition of the Multi Gas Leak Detector as of the time of publication.<br />

All rights reserved for future technical changes and data changes that may affect the contents of this information!<br />

Conventions<br />

Handling instructions that consist of a sequence over time are combined into handling units.<br />

Safety instructions are labeled using pictograms and a signal word, and the hazard is specifically named<br />

where applicable. The meaning of the pictograms is provided in the "Safety instructions" section.<br />

Please do not hesitate to contact us with any questions or technical problems:<br />

…………<br />

BAL_MGLD3000_20101001_en.doc 4-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

2. Safety instructions<br />

To ensure your own safety, and to enable problem-free and safe operation of the <strong>MGLD.3000</strong>, please<br />

carefully read this chapter before initial operation.<br />

Please observe and closely follow all safety instructions that are listed in this manual as well as all messages<br />

and instructions that are displayed by the <strong>MGLD.3000</strong>.<br />

Damage arising as a result of improper or unauthorized use of the device, or through the use of nonoriginal<br />

spare parts, wear parts, or consumables, is not covered under warranty! Likewise, damage to<br />

wear parts is not covered by the warranty!<br />

2.1. Symbols, pictograms, and icons used<br />

2.1.1. Warning symbols and instruction symbols in the text of these operating instructions<br />

Caution!<br />

This symbol indicates a source of potential danger to the user.<br />

Caution!<br />

Hot surface!<br />

Caution!<br />

Live voltage. Shock hazard.<br />

Caution!<br />

This symbol indicates a potential hazard to the device or the device<br />

system.<br />

Disconnect power<br />

plugs!<br />

Attention!<br />

This symbol designates information on possible operating errors and<br />

proper analysis of results.<br />

2.1.2. Pictograms and icons used in screen menus<br />

Menus and menu entries<br />

Select leak search mode<br />

Analysis mode for quantitative gas leak detection (determines gas concentrations<br />

or leak rate)<br />

Dynamic search mode for qualitative leak localization<br />

BAL_MGLD3000_20101001_en.doc 5-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Multipoint mode with logging function<br />

Check + calibration<br />

active<br />

inactive<br />

Check and calibration function<br />

Check and calibration results<br />

Settings of leak search environment<br />

Settings<br />

active<br />

inactive<br />

Settings for analysis mode (threshold value)<br />

Settings for multipoint mode (number of test points, threshold values,<br />

test criteria, test identification)<br />

Alarm signal setting<br />

Gas type and leakage unit setting<br />

Language, date, and time setting<br />

Print out log and save to external data medium<br />

active<br />

inactive<br />

Select drive for storage of log data in multipoint mode, download log<br />

data<br />

Print settings for external label printer<br />

Service<br />

active<br />

inactive<br />

Query device information<br />

Set up password protection<br />

Contact information<br />

BAL_MGLD3000_20101001_en.doc 6-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Warnings and functions<br />

Warning symbol - blinks yellow in case of error<br />

No warning present<br />

Probe is being initialized and heated, Probe is calibrated<br />

Probe not connected<br />

Probe active<br />

Password protection inactive (changes can be made to device settings)<br />

Password protection active (changes to device settings are not possible)<br />

Connected to power socket<br />

Not connected to power socket - battery mode<br />

Battery charge indicator<br />

…<br />

Buttons for numeric entry<br />

Confirm<br />

Cancel<br />

Delete previous character<br />

Cancel a multipoint leak search job<br />

BAL_MGLD3000_20101001_en.doc 7-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

2.1.3. Pictograms/probe control elements used<br />

The following probe control elements can be assigned to different operating modes on the<br />

<strong>MGLD.3000</strong>. The current function of the relevant control element is explained in the relevant<br />

passage of these operating instructions. The following are typical functions:<br />

Key: Start<br />

Key: Confirm<br />

Key: Reduce sensitivity<br />

Key: <strong>Inc</strong>rease sensitivity<br />

2.2. General safety instructions<br />

Proper use<br />

The <strong>MGLD.3000</strong> may only be used for the leak detection procedures described in this manual. The device<br />

is a gas leak detector and cannot be used for measurement purposes or for stationary monitoring of ambient<br />

air.<br />

Improper use and/or operation can potentially endanger the user or other persons, or risk damage to the<br />

device. Failure to adhere to the instructions for proper use in this manual will result in reduced warranty<br />

coverage and manufacturer liability in the event that damage does occur.<br />

Local requirements<br />

Observe the safety precautions that are applicable for operation of the device (e.g., work safety precautions,<br />

accident prevention guidelines, and regulations for environmental protection).<br />

Information on potential dangers is not a substitute for mandatory work safety precautions.<br />

Personnel<br />

The <strong>MGLD.3000</strong> may only be operated by qualified personnel, who are also provided with training in<br />

proper performance of these tasks. As part of training, the contents of these operating instructions are to<br />

be explicitly communicated and instructions are to be provided on proper device operation.<br />

Use area<br />

The <strong>MGLD.3000</strong> may not be used in explosive and/or other hazardous environments. Smoking is prohibited<br />

in places where the <strong>MGLD.3000</strong> is in use.<br />

The usage area and storage area must be free of any impurities and contaminants, and in particular must<br />

be protected from the influence of chemical substances, e.g. silicone.<br />

BAL_MGLD3000_20101001_en.doc 8-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Siliceous- and sulfurous media or non-desorbing, inorganic impurities, e.g. tobacco smoke, oils, fats, and<br />

evaporating liquids, can produce contaminants which can result in changes to the sensitivity and/or selectivity<br />

qualities of the sensor. Possible consequences of this include reduced sensitivity, inaccurate concentration<br />

indicators, or a background concentration message. To detect changes in the sensor performance,<br />

a yearly check on sensors is performed by the manufacturer or at an authorized service center.<br />

Commissioning and operation<br />

The device is connected to the power socket using the supplied power connection cord without additional<br />

extensions. Use of extensions, feed lines, or connection lines for peripherals is not permitted!<br />

The device may only be repaired by customer service. Any unauthorized interference can endanger the<br />

user, jeopardize the functional reliability of the device, and cancel warranty claims.<br />

While searching for leaks, pay close attention to the path where the electro-pneumatic connection line<br />

runs between the device and the sensor. The connection line should be laid in such a way as to minimize<br />

the danger of tripping for the user of the device and other workers within the user's radius of activity.<br />

Emergency shutdown<br />

In case of emergency, shut off the <strong>MGLD.3000</strong> device power switch at the back of the housing and disconnect<br />

the power plug.<br />

Electric shock<br />

Before opening the device, always switch it off and disconnect the power plug!<br />

Interventions in the device, service- or repair works can only be performed by the manufacturer's customer<br />

service department or by authorized and trained technicians.<br />

Ventilation<br />

Make certain that the available ventilation systems on the <strong>MGLD.3000</strong> and accessory devices are functional.<br />

Concealed ventilation grids, slits, etc., can cause malfunctions or damage the device.<br />

Repairs, cleaning, and service<br />

Interventions in the device, service- or repair work may only be performed by the manufacturer's customer<br />

service department or by authorized trained technicians. Failure to follow these instructions can result in<br />

damage to the device.<br />

Any unauthorized interference can endanger the user, jeopardize the functional reliability of the device,<br />

and undermine warranty claims.<br />

An exception is made for the activities listed in the chapter "Care and service".<br />

BAL_MGLD3000_20101001_en.doc 9-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Sensitive electronic equipment<br />

Always switch off components/probes before connecting or disconnecting power. This also applies to electronic<br />

connections between components.<br />

The carbon dioxide hand probe contains a highly sensitive optical system. The system must be protected<br />

from any strong impact that could damage it. Impure intake air can result in condensation formation on<br />

optical components which may compromise the performance of the probe.<br />

Use of the touch screen requires clean fingers, lightweight gloves or pens that are specially designed for<br />

the touch screen. Physical damage (scratches) or dirt can hamper the touch screen and result in malfunctions<br />

and/or reduced functionality on the entire device.<br />

Accessories / spare parts<br />

Use only accessories and spare parts from the manufacturer.<br />

Please follow the safety guidelines contained in these operating instructions.<br />

BAL_MGLD3000_20101001_en.doc 10-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

3. Description of the device<br />

3.1. General information<br />

The <strong>MGLD.3000</strong> Multi Gas Leak Detector is a computer-controlled multi-gas detector for both mobile and<br />

stationary operation in detecting leaks of selected gases and refrigerants.<br />

The modular device consists of a portable base unit, which can be used as desired for either onsite measurement<br />

(battery mode) or in stationary mode, and a docking-/charging unit for stationary mains operation,<br />

and can be run with various active probes for gas detection (e.g. for refrigerants, hydrogen, and carbon<br />

dioxide).<br />

The device, data indicators, and system messages, etc., are configured and controlled via a graphical<br />

user interface on a graphic Touch Screen color display. In addition, an acoustic signal is emitted from an<br />

integrated mini-speaker.<br />

The device offers various operating modes for leak detection:<br />

• Analysis mode for quantitative gas leak detection (concentrations or leak rate)<br />

• Dynamic search mode for qualitative leak localization<br />

• Multipoint mode with logging function.<br />

For integration with testing environments, the <strong>MGLD.3000</strong> features various standardized interfaces<br />

(USB2.0, RS232), e.g. for barcode scanners, printers, and external storage media.<br />

3.2. Device components<br />

3.2.1. Base unit<br />

3<br />

1… Touch screen display<br />

2… Connection socket for probe<br />

3… Transport handle<br />

1<br />

2<br />

Figure 1: <strong>MGLD.3000</strong> - Base unit (front view)<br />

BAL_MGLD3000_20101001_en.doc 11-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

13 12<br />

10<br />

4 5 6<br />

7 8<br />

9<br />

Figure 2: <strong>MGLD.3000</strong> - Base unit (rear view)<br />

11<br />

10<br />

4… On-/off switch<br />

5… RS232<br />

6… VGA<br />

7… RJ45 (socket)<br />

8… USB2<br />

9… USB1<br />

10… Ventilator<br />

11... Transport handle<br />

12… Speaker jack (mini-jack)<br />

13... Mini speaker<br />

Caution<br />

The power supply of the <strong>MGLD.3000</strong> base unit for stationary operation and for charging the batteries that<br />

are integrated in the base unit may only be supplied using the included docking unit.<br />

3.2.2. Docking unit<br />

The docking unit is used for the power supply of the base unit via mains power for the stationary mode of<br />

the <strong>MGLD.3000</strong>. Defined charging of the batteries in the base unit occurs simultaneously.<br />

Caution<br />

The docking unit may only be used for docking the base unit, for the power supply to the base unit and for<br />

charging the batteries on the <strong>MGLD.3000</strong> base unit. The docking unit comes included with the<br />

<strong>MGLD.3000</strong>.<br />

1<br />

1<br />

2<br />

1… Guide pin for base unit<br />

2… Connecting jack for base unit<br />

Figure 3: <strong>MGLD.3000</strong> - Docking unit (front view)<br />

BAL_MGLD3000_20101001_en.doc 12-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

5<br />

3… Connection for power cable<br />

4… Plug connectors for control<br />

valve terminal (optional)<br />

5… USB3<br />

3<br />

4<br />

Figure 4: <strong>MGLD.3000</strong> - Docking unit (rear view)<br />

3.2.3. Probes<br />

Various active probes from the manufacturer for refrigerants, hydrogen, and carbon dioxide are available for the<br />

<strong>MGLD.3000</strong>. The probes consist of a compact aluminum housing and a sensor which is offset with a flexible extension.<br />

The probe features an LC display (4-digit), LEDs for function and trend indicators (1 green, 2 x yellow, and 2 x<br />

red), and a control panel with 4 keys for "remote control" of selected leak search functions of the <strong>MGLD.3000</strong><br />

(Figure 5: Probe for <strong>MGLD.3000</strong> (Example: refrigerant)<br />

.<br />

1… flexible sensor extension<br />

2… Sensor head<br />

3… LED function and trend indicators<br />

4… LC display<br />

5… Control panel<br />

5.1… "Start" key<br />

5.2… "Confirm" key<br />

5.3… "Reduce sensitivity" key<br />

5.4… "<strong>Inc</strong>rease sensitivity" key<br />

6… Socket for connection line<br />

Figure 5: Probe for <strong>MGLD.3000</strong> (Example: refrigerant)<br />

BAL_MGLD3000_20101001_en.doc 13-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

3.2.4. Electro-pneumatic connection line<br />

The relevant probe is connected with the base unit of the MGLD 3000 using a special electro-pneumatic connection<br />

line which is included with the <strong>MGLD.3000</strong><br />

(Figure 6: Electro-pneumatic connection line for connecting a probe to the base unit of the <strong>MGLD.3000</strong><br />

).<br />

Figure 6: Electro-pneumatic connection line for connecting a probe to the base unit of the <strong>MGLD.3000</strong><br />

3.3. Technical data<br />

3.3.1. <strong>MGLD.3000</strong> - Base unit<br />

Time from startup to operational readiness<br />

< 120 sec (varies by sensor)<br />

Response time (varies by sensor)<br />


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Charging time<br />

approx. 8 hours<br />

Permissible operating temperature 0 °C – +55 °C<br />

Permissible operating humidity<br />

20... 80 % rel. hum.<br />

Permissible storage temperature –25 °C – +70 °C<br />

Permissible storage humidity<br />

20... 80 % rel. hum.<br />

Permissible storage conditions<br />

The storage area and environment must be free of any<br />

impurities and contaminants, and in particular must be<br />

protected from the influence of chemical substances, e.g.<br />

silicone.<br />

Base unit dimensions (length x width x height)<br />

approx. 262 mm x 292 mm x 174 mm<br />

Net weight (base unit)<br />

approx. 6.250 kg (without probe and connection line)<br />

Conformity:<br />

2002/95/EC Restriction of the use of Hazardous Substances<br />

Directive (RoHS)<br />

3.3.2. <strong>MGLD.3000</strong> docking unit<br />

Interfaces 1 x USB 2.0<br />

1 x 8-pin plug connector with 4 switching outputs (max.<br />

1A/125 VAC) (optional)<br />

Permissible operating temperature 0 °C – +55 °C<br />

Permissible operating humidity<br />

20... 80% rel. hum.<br />

Permissible storage temperature -25 °C – +70 °C<br />

Permissible storage humidity<br />

20... 80% rel. hum.<br />

Power supply (docking unit)<br />

100-240 V, 50/60 Hz<br />

Power consumption<br />

max. 100W (incl. optional switching outputs)<br />

Protection class<br />

Permissible storage conditions<br />

The storage area and environment must be free of any<br />

impurities and contaminants, and in particular must be<br />

protected from the influence of chemical substances, e.g.<br />

silicone.<br />

Docking unit dimensions (length x width x height) approx. 262 mm x 292 mm x 110 mm<br />

Net weight (docking unit)<br />

approx. 2.450 kg (without connection line)<br />

Conformity:<br />

2002/95/EC Restriction of the use of Hazardous Substances<br />

Directive (RoHS)<br />

Base unit and docking unit dimensions (length x width x<br />

height)<br />

Net weight (base unit and docking unit)<br />

approx. 262 mm x 292 mm x 265 mm<br />

approx. 8.7 kg (without probe and connection line)<br />

BAL_MGLD3000_20101001_en.doc 15-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

3.3.3. Probes for the <strong>MGLD.3000</strong><br />

<strong>Product</strong> number L14004006 L14004007 L14004008<br />

<strong>Product</strong> description<br />

LOKtool Refrigerant<br />

Probe<br />

LOKtool Tracergas Probe LOKtool CO2-R744<br />

Probe<br />

Labeling on the housing<br />

REFRIGERANT TRACERGAS / H2 CO2<br />

Detectable gases R22, R134a, R290,<br />

R404A, R407C, R410A*,<br />

R600a*<br />

Lower detection threshold value 10 ppm, 0.3 g/year 1 ppm 100 ppm, 5 g/year<br />

*0.1 g/year<br />

Response time < 1 sec < 1 sec < 2 sec<br />

Dimensions (probe without flexible<br />

sensor extension)<br />

approx. 192 mm x 40<br />

mm x 28 mm<br />

approx. 192 mm x 40 mm<br />

x 28 mm<br />

approx. 180 mm x 54<br />

mm x 28 mm<br />

Dimensions (probe with flexible<br />

sensor extension)<br />

approx. 522 mm x 40<br />

mm x 28 mm<br />

approx. 522 mm x 40 mm<br />

x 28 mm<br />

approx. 510 mm x 54<br />

mm x 28 mm<br />

Dimensions (probe with flexible approx. 295 g approx. 295 g approx. 410 g<br />

sensor extension)<br />

Length of connection line approx. 6 m<br />

Net weight of connection line approx. 430 g<br />

Permissible operating temperature<br />

0 °C – +55 °C<br />

Permissible operating humidity<br />

20... 80 % rel. hum.<br />

Permissible storage temperature -25 °C – +70 °C<br />

Permissible storage humidity<br />

20... 80 % rel. hum.<br />

Permissible storage conditions The storage area and environment must be free of any impurities and contaminants,<br />

and in particular must be protected from the influence of chemical substances,<br />

e.g. silicone.<br />

H 2<br />

R744<br />

BAL_MGLD3000_20101001_en.doc 16-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

4. Installation and commissioning<br />

Caution!<br />

Follow the safety instructions provided in chapter 2.<br />

4.1. Unpacking device from packing materials<br />

When unpacking the device, examine the device and compare it with the packing list to make sure that all<br />

parts are included and undamaged.<br />

Do not discard the transport packaging. In the event that the materials need to be returned, they will need<br />

to be returned in the original packaging. Only in this way can the returned materials be protected from<br />

damage during transport.<br />

4.2. Installation and commissioning the <strong>MGLD.3000</strong><br />

4.2.1. Stationary operation of the device with docking unit/charging batteries on<br />

the base unit<br />

Caution!<br />

Exercise caution during device assembly and setup to prevent injury to hands and fingers!<br />

Place the docking unit on its feet in the designated location. A level, dry, secure, and stable surface is<br />

required. Make sure the docking unit is placed on a level and stable surface where it cannot slip or wobble.<br />

Connect the included power connection cord to the docking unit power connection.<br />

Caution!<br />

Observe the layout of both devices (front side by front side)! There must be no foreign objects between<br />

the docking unit and the base unit<br />

BAL_MGLD3000_20101001_en.doc 17-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Place the base unit with its 4 guide bushes onto the 4 guide pins of the docking unit<br />

(Figure 7: Placing the base unit on the docking unit<br />

).<br />

<br />

Figure 7: Placing the base unit on the docking unit<br />

Connect the docking unit power connection cord with the power socket.<br />

The batteries are already being charged in the base unit in this state (base unit switched off). The charging<br />

period is approx. 8 hours with empty batteries.<br />

4.2.2. Connecting the probe (stationary and mobile operation of the <strong>MGLD.3000</strong>)<br />

Caution!<br />

Only probes that are supplied by the manufacturer, as well as the included connection line, can be used<br />

for operation with the <strong>MGLD.3000</strong>.<br />

Before installation and operation, check the probe and the connection line for damage. The connection<br />

line must be free of any kinks, crushed or compressed places, cracks, or other damage and/or contamination.<br />

While operating the probe, make sure that the device is not exposed to crushing, stretching, compressing,<br />

or to any deformations and impurities.<br />

Connect the relevant probe with the provided connection socket with the plug of the supplied connection line. For<br />

maximum accuracy of positioning and in connecting the plug on the connection line with the socket on the probe, the<br />

plug features a nose and the socket has a matching opening. The plugs are also provided with an automatic stopping<br />

mechanism, which provides a firm connection between the relevant plug and socket. When connecting the connection<br />

line with the probe, make sure that you can hear and feel the plug engage with the socket of the probe.<br />

BAL_MGLD3000_20101001_en.doc 18-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

BAL_MGLD3000_20101001_en.doc 19-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

(Figure 8: Plugging the connection line plug into the sensor socket<br />

).<br />

Figure 8: Plugging the connection line plug into the sensor socket<br />

Connect the plug on the other end of the connection line of the sensor with the appropriate socket on the<br />

base unit while carefully observing the above information. Make sure that you are able to feel and hear<br />

the plug engage with the socket on the base unit.<br />

Figure 9: <strong>MGLD.3000</strong> with connected sensor<br />

The connection of sensor and base unit may only be made via the connection line when the MGLD 3000<br />

is switched off.<br />

BAL_MGLD3000_20101001_en.doc 20-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

To disconnect the connection between the relevant plug of the connection line and/or the relevant socket<br />

on the sensor or base unit, grasp the plug with your thumb and index finger on the metal ring on the exterior<br />

of the plug, carefully push it away from the socket, and then remove the plug.<br />

Figure 10: Loosening the lock and disconnecting the plug of the connection line from the socket<br />

Caution!<br />

The sensor and base unit can only be connected with (or disconnected from) the connection line when the<br />

MGLD 3000 is switched off!<br />

Force must not be use when connecting or disconnection the connection line and probe or base unit. Otherwise<br />

you may damage the plugs, connection line and/or the sockets in the probe or base unit, thus limiting<br />

device functionality and canceling warranty coverage.<br />

BAL_MGLD3000_20101001_en.doc 21-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

4.3. Operating the <strong>MGLD.3000</strong><br />

4.3.1. Basic information on the graphical user navigation<br />

Caution!<br />

The base unit of the <strong>MGLD.3000</strong> is controlled via touch screen. Use of the touch screen requires clean<br />

fingers, lightweight gloves or pens that are specially designed for the touch screen. Mechanical damage<br />

(scratches) or dirt can damage the touch screen or result in malfunctions and/or unsatisfactory performance<br />

of the entire device.<br />

The menu tab and main menu are located at the top of the screen. All operating modes can be selected<br />

and all of the settings can be made from the main menu.<br />

By pressing the pictograms in the main menu, the menu bar moves down and the relevant submenu is<br />

displayed. The individual operating modes and settings can be accessed/entered by pressing the pictograms<br />

of the relevant submenu.<br />

The submenu will remain displayed while a pictogram or any other place on the touch screen is touched,<br />

otherwise it will automatically disappear after 10 sec.<br />

When password protection is activated, only the "leak search mode" is available. All other menu items in<br />

the main menu are inactive.<br />

The menu structure is shown in the summary below:<br />

Select leak search mode<br />

Analysis mode for quantitative gas leak detection (determines gas concentrations<br />

or leak rate)<br />

Dynamic search mode for qualitative leak localization<br />

Multipoint mode with logging function<br />

Check + calibration<br />

active<br />

inactive<br />

Check and calibration function<br />

Check and calibration results<br />

Settings of leak search environment<br />

Settings<br />

active<br />

inactive<br />

Settings for analysis mode (threshold value)<br />

BAL_MGLD3000_20101001_en.doc 22-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Settings for multipoint mode (number of test points, threshold values,<br />

test criteria, test identification)<br />

Alarm signal setting<br />

Gas type and leakage unit setting<br />

Language, date, and time setting<br />

Print out log and save to external data medium<br />

active<br />

inactive<br />

Select drive for storage of log data in multipoint mode, download log<br />

data<br />

Print settings for external label printer<br />

Service<br />

active<br />

inactive<br />

Query device information<br />

Set up password protection<br />

Contact information<br />

4.3.2. Initial startup of the <strong>MGLD.3000</strong><br />

When you have installed the <strong>MGLD.3000</strong> as described in 4.2., switch on the device by briefly pressing the<br />

power switch on the rear of the base unit once. The base unit acknowledges the start with a brief signal<br />

tone.<br />

Probe<br />

The probe also acknowledges the start with a short signal tone, the green LED blinks, the word "HEAT"<br />

appears in the display indicating that the gas sensor is heating.<br />

Base unit<br />

When the system software starts, the <strong>MGLD.3000</strong> displays the start screen.<br />

BAL_MGLD3000_20101001_en.doc 23-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Start Menus Warnings / functions<br />

Select leak search<br />

mode<br />

inactive<br />

Warning symbol - blinks<br />

while sensor is heated<br />

Sensor is heating<br />

inactive<br />

inactive<br />

inactive<br />

protected with password<br />

Connected to power socket<br />

Battery charge status<br />

Yellow background<br />

Warning/error present:<br />

Sensor is heating<br />

After heating up the probe (1 - 2 min. depending on special type) the<br />

<strong>MGLD.3000</strong> is ready to operate and the probe is active.<br />

Device ready to operate<br />

Probe active<br />

The <strong>MGLD.3000</strong> provides the option to protect the settings with a user password<br />

and is protected with an initial password set at the factory.<br />

Press the "lock" pictogram to access the user password query.<br />

protected with password<br />

Password<br />

Select leak search<br />

mode<br />

inactive<br />

Warnings / functions<br />

Device ready to operate<br />

Probe active<br />

inactive<br />

inactive<br />

inactive<br />

protected with password<br />

Connected to power<br />

socket<br />

Battery charge status<br />

Entry<br />

Enter the correct initial password (123) to remove the lock. 123<br />

Delete the previous character<br />

Confirm<br />

Cancel<br />

Next, the leak search mode opens that was in use before the device was last<br />

switched off.<br />

BAL_MGLD3000_20101001_en.doc 24-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

4.3.3. Leak search mode<br />

Analysis mode Menus Warnings / functions<br />

Select leak search<br />

mode<br />

Check + calibration<br />

Device ready to operate<br />

Probe active<br />

Settings<br />

Print out log and<br />

save to external<br />

data medium<br />

Service (device<br />

status, password,<br />

contact information)<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

The leak search mode being used is displayed below the menu bar on the<br />

left, the alarm threshold setting is displayed in the center, and the selected<br />

gas type is displayed on the right .<br />

The <strong>MGLD.3000</strong> automatically detects the type of sensor that is connected<br />

at the start. The gas type that was most recently selected with<br />

this special type is saved and displayed by special type, either in red<br />

(hydrogen), blue (refrigerant), or black (CO2) (example above:<br />

5%H2).<br />

This and the indicator of the settings are adequate in the other leak<br />

search modes "dynamic search mode" and "multipoint mode".<br />

The analysis mode is provided for quantitative gas leak detection. Based on<br />

the setting, the currently detected concentration or leak rate is displayed as a<br />

number with the relevant unit (e.g., ppm for the concentration) and in graphical<br />

form as a temporal progression. As the concentration increases the frequency<br />

increases and the tone sequence of the acoustic signal becomes<br />

shorter. When the threshold boundary is exceeded, a continuous sound is<br />

emitted and the screen background and numeric data are colored red.<br />

BAL_MGLD3000_20101001_en.doc 25-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Probe<br />

• Display indicator of the currently measured concentration (value and unit<br />

ppm) or leak rate (value only).<br />

• Change to dynamic search mode<br />

Entry<br />

• Change (back) to analysis mode<br />

• As the concentration increases the tone sequence of the acoustic signal<br />

of the hand probe becomes shorter. When the threshold value is exceeded,<br />

a continuous tone is emitted. The LED trend indicator also acts<br />

as an indicator for the current concentration relative to the threshold<br />

value.<br />

Up to 22%, green LED<br />

Up to 44%, 1st yellow LED<br />

Up to 66%, 2nd yellow LED<br />

Up to 100%, 1st red LED<br />

Up to 100%, 2nd red LED<br />

Dynamic search mode Menus Warnings / functions<br />

Select leak search<br />

mode<br />

Check + calibration<br />

Device ready to operate<br />

Probe active<br />

Settings<br />

Print out log and<br />

save to external<br />

data medium<br />

Service (device<br />

status, password,<br />

contact information)<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

The dynamic search mode is provided for quick leak detection. The display of<br />

the leakage level is made qualitatively using an indicator with changing color<br />

(green, yellow or red with increasing leakage level).<br />

Using the sliding controller, the sensitivity can be set in 5 levels (min. - lowest<br />

sensitivity to max - highest sensitivity).<br />

The optical and acoustic signal outputs are not based on the absolute concentration<br />

or leak rate, but rather on the quantity of their change. A large<br />

change in concentration generally indicates a large leak.<br />

After detection of a large concentration increase and subsequent removal of<br />

the probe from the leakage, "AUTOZERO" can be displayed between the<br />

sliding controller and the indicator for the leakage level. The sensor returns to<br />

its zero point while this message is visible. During this time the response<br />

characteristics are limited particularly when small leakages are detected.<br />

BAL_MGLD3000_20101001_en.doc 26-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

The signal output in dynamic search mode is also dependent on the<br />

speed with which the sensor is approaching the leak. A very slow<br />

approach corresponds to a slow increase in concentration and accordingly<br />

a minimal change in concentration.<br />

Configuration of sensitivity<br />

• Touch screen: Touching and vertical sliding of the indicator on the sliding<br />

controller to the desired sensitivity level.<br />

• Sensor (each press of the relevant button increases or reduces the sensitivity<br />

by 1 level)<br />

When changing the sensitivity level, there is no indicator of this in the<br />

display for sensor. The configured sensitivity level can only be controlled<br />

via the touch screen.<br />

The most recent sensitivity level setting is also retained after the<br />

base unit is switched off and back on.<br />

Probe<br />

• Display indicator of the currently measured concentration (value and unit<br />

in ppm) or leak rage (value only) - the value corresponds to the value of<br />

the concentration in analysis mode.<br />

• Change to analysis mode<br />

• Change (back) to dynamic search mode<br />

• As the concentration increases the tone sequence of the acoustic signal<br />

of the hand probe becomes shorter. The LED trend indicator also acts as<br />

an indicator for the change in concentration.<br />

BAL_MGLD3000_20101001_en.doc 27-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Multipoint mode Menus Warnings / functions<br />

Select leak search<br />

mode<br />

Check + calibration<br />

Cancel a multipoint leak<br />

search<br />

Device ready to operate<br />

Settings<br />

Print out log and<br />

save to external<br />

data medium<br />

Service (device<br />

status, password,<br />

contact information)<br />

Probe active<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

Like analysis mode, multipoint mode is provided for quantitative gas leak<br />

detection. This mode however offers the added option of determining a cumulative<br />

concentration or leak rate using multiple test points and to log and print<br />

out the results of the test sequence.<br />

The maximum currently measured concentration or leak rate of the current<br />

test point and the sum of concentrations or leak rates of all test points up to<br />

the last/current test point are displayed, based on the setting, as a number<br />

with the relevant unit (e.g. g/y for the leak rate) and in graphical form as a<br />

temporal progression.<br />

see menu Settings - set gas type and leak unit<br />

see menu Print out log and save to external data storage medium<br />

- select drive for storage of log data in multipoint mode<br />

There are anywhere between 1 and 30 test points in a test sequence.<br />

The following test criteria are possible:<br />

• Maximum concentration/-leakage rate of each individual test point over<br />

point threshold<br />

or<br />

• Sum of maximum concentrations/-leakage rates of all test points above<br />

cumulative threshold.<br />

see menu Settings - Settings for multipoint mode<br />

Touch Screen indicator (e.g.):<br />

ID… Test identifier, is either issued automatically by the device at the start of<br />

a test sequence (format: JJJJMMTTHHMMSS) or can also be read with<br />

a barcode scanner as desired<br />

BAL_MGLD3000_20101001_en.doc 28-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

3/5… Number of the current test point / total number of test points<br />

and to the right of this is the maximum current measured concentration<br />

or leakage rate of the current test point (in the example above, test<br />

point no. 3 and 0.9 g/y)<br />

∑… Sum of the leakage rates or concentrations of all test points up to the<br />

last/current test point<br />

The current concentration or leakage rate of the current test point is<br />

shown in the display for the probe.<br />

• When the threshold point value is exceeded the background of the<br />

screen and the numeric indicator of the concentration or leakage rate of<br />

the test point turns red. When the cumulative threshold value is exceeded,<br />

the number indicator of the cumulative concentration or leakage<br />

rate turns red. The acoustic signal is emitted in a similar manner as in<br />

analysis mode.<br />

Probe<br />

• Start of the test sequence with the selection of the first test point and<br />

indicator for the number of the current test point and the total number of<br />

configured test points in the display (Ex.: 1 of 5).<br />

• Display indicator of the currently measured concentration (value and unit<br />

in ppm) or leakage rate (value only) of the current test point. The readiness<br />

for detection of the test point is indicated by an increasing and decreasing<br />

status button under the number indicator.<br />

• Completion of the detection of the selected test point<br />

01.05<br />

20 ppm<br />

• Selection of the next test point and indicator for the number of the current<br />

test point and the total number of configured test points in the display<br />

(Ex.: 2 of 5)<br />

• Display indicator of the currently measured concentration (value and unit<br />

in ppm) or leakage rate (value only) of the current test point<br />

• Detection of the selected test point finished<br />

02.05<br />

30 ppm<br />

• …<br />

• Selection of the last test point and indicator for the number of the current<br />

test point and the total number of configured test points in the display<br />

(Ex.: 5 of 5)<br />

• Display indicator of the currently measured concentration (value and unit<br />

in ppm) or leakage rate (value only) of the current test point<br />

• Completion of the detection of the last test point<br />

05.05<br />

30 ppm<br />

• Indicator of the results of the test sequence<br />

BAL_MGLD3000_20101001_en.doc 29-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

• When the last test sequence is finished: Start of the next test sequence<br />

• Verlassen des Mehrpunktmodus: „Zurück“<br />

A multi-point leak detection could be stopped by pressing the "BREAK"<br />

ended and the multi-point mode is been able to leave.<br />

After leaving the multi-point mode, all leak detection modes are available on<br />

the menu selection to call or set changes.<br />

For each test series, a log is created that is stored in the device or on memory<br />

stick. The print of a test report for example as a label is optionally possible.<br />

4.3.4. Check + calibration<br />

Check + calibration Menu entries Warnings / functions<br />

Select reference leak<br />

and start calibration<br />

procedure<br />

Check and calibration<br />

results<br />

Environment setting<br />

Device ready to operate<br />

Probe active<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

While using the hand probe the sensitivity of the gas sensors being used can<br />

change, even while being used properly (e.g. because of impurities).<br />

To compensate for these changes in sensitivity, a function is provided for testing<br />

the sensitivity on reference leaks and for setting a gas correction factor.<br />

BAL_MGLD3000_20101001_en.doc 30-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Before starting the functional test, the hand probe must be switched<br />

on for approx. 5 min. at a minimum and run in "clean" surrounding air<br />

(minimal or no background concentration of the gas types that are<br />

specific to the special type). These conditions are required for calculation<br />

of the zero point. If no zero point can be set, the functional test<br />

is stopped. The device must then be restarted and the functional test<br />

must be repeated after waiting 5 min.<br />

Continuous contact during the functional test must be ensured between<br />

the Teflon tip of the hand probe with the head of the test leakage.<br />

Only manufacturer-approved reference leaks can be used for the<br />

functional test.<br />

The functional test must be carried out in enclosed spaces with minimal<br />

air movement.<br />

BAL_MGLD3000_20101001_en.doc 31-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Select reference leak and start calibration<br />

procedure<br />

Warnings / functions<br />

Warning symbol - blinks<br />

during check and calibration<br />

Probe is initialized<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

First, a reference leak must be selected that is suitable for the gas type being<br />

used.<br />

The available reference leaks are pre-configured by technical support.<br />

List selection (press a list entry, scroll by pressing and simultaneously moving<br />

the finger up and down or by pressing the scroll bar, the current setting is<br />

highlighted in light blue)<br />

Start of test by pressing the "Start" button and pressing the "Start" key on<br />

the hand probe<br />

Cancel and return to the previous leak search menu<br />

"CAL" appears in the display during the entire test- and calibration procedure.<br />

Auto zero<br />

Warnings / functions<br />

Warning symbol - blinks<br />

during check and calibration<br />

Probe is initialized<br />

The present zero point in the surrounding air is set. Once this point is<br />

reached, the message "RefLeak" is displayed.<br />

Cancel and return to the previous leak search menu<br />

BAL_MGLD3000_20101001_en.doc 32-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Calibration procedure<br />

Warnings / functions<br />

Warning symbol - blinks<br />

during check and calibration<br />

Probe is initialized<br />

The sensor tip must be held at the reference leak within 5 seconds. If the<br />

reference leak is detected (through an increase in the concentration) the functional<br />

test begins, otherwise the procedure is canceled. Make sure the reference<br />

leak is opened and is filled with test gas in the specified pressure range.<br />

The functional test must be repeated.<br />

If the measured concentration value is inside the specified range (set-point +/-<br />

30%), the functional test is completed successfully and a correction factor is<br />

set to adjust the sensitivity.<br />

Cancel and return to the previous leak search menu<br />

Tracing<br />

Warnings / functions<br />

Warning symbol - blinks<br />

during check and calibration<br />

Probe is initialized<br />

The sensitivity level of the hand probe is indicated by the position of the<br />

pointer in the pointer instrument. If the sensitivity matches the set-point values,<br />

this is indicated by a pointer position in the middle green range of the<br />

scale. The functional test was passed. If the sensitivity is within the specified<br />

tolerance limits, but is smaller or higher than the set-point, this is indicated by<br />

a pointer that is in the left or right yellow part of the scale range. The functional<br />

test was passed and a correction factor was calculated.<br />

BAL_MGLD3000_20101001_en.doc 33-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

OK - Save the correction factor in the probe.<br />

Cancel and return to the previous leak search menu<br />

If the sensitivity is outside the specified tolerance limits, this is indicated by a<br />

pointer that is at either red end of the scale on the left or right. The functional<br />

test was not passed. The correction factor was set to the maximum limit.<br />

Start the new test by pressing the "Start" button<br />

Cancel and return to the previous leak search menu<br />

If the test is run several consecutive times without a passing result,<br />

the following procedure can be followed:<br />

• Make sure that the Teflon tip was being held fully in front of the<br />

gas outlet of the reference leak and that there is no air movement<br />

in the environment<br />

• Make sure that the reference leak is opened for long enough to<br />

be completely permeated by the test gas (follow manufacturer<br />

instructions)<br />

• Make sure that the pressure of the test gas is consistent with the<br />

reference leak manufacturer's specifications<br />

• Restart device and repeat test (when the device is started the<br />

sensor and the gas sensor system are initialized)<br />

• Contact technical support<br />

BAL_MGLD3000_20101001_en.doc 34-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Check and calibration results<br />

Warnings / functions<br />

Warning symbol - blinks<br />

during check and calibration<br />

Probe is initialized<br />

Date and results of the last 10 functional tests are logged for each probe.<br />

Return to the previous leak search menu<br />

Return to the previous leak search menu<br />

Environment setting<br />

In addition to the sensitivity of the hand probe, there is a second factor that must be taken into account for<br />

a successful leak search.<br />

To determine the size of a leak, it is necessary that wherever possible the entire gas quantity that flows<br />

through the leak within a specific time unit is detected by the hand probe of the leak detector and can thus<br />

be detected by the gas sensor. The transfer path from the leak to the leak detector is directly determined<br />

by the air flow in the surrounding air. There are different flows in the surrounding air in different environments<br />

(e.g., wind during a leak test conducted outdoors, a continuous air draft caused by air-conditioners<br />

in indoor spaces). The larger the flow in the surrounding air, the less gas flowing from the leak that<br />

reaches the gas sensor in the hand probe. However, to enable a leak search to be conducted in variable<br />

environmental situations, a transfer function can be selected from the leak to the gas sensor which by<br />

means of a mathematical procedure increases the measured concentration within specific limits and in this<br />

way should indirectly account for the gas portion not reaching the gas sensor.<br />

The change in selectivity is another way of detecting leaks adapt to specific environmental conditions.<br />

Setting a high selectivity allows a separation between the refrigerant gases of the detectable probe. A<br />

strong separation, however, can usually only be achieved under non-congested environment with very low<br />

background concentrations of other detectable gases from the gas sensor (eg, hydrocarbons, hydrogen).<br />

Setting a low selectivity allows detection of leaks, even under polluted ambient conditions. This setting,<br />

however, the separation between the different refrigerants is reduced.<br />

A too high setting can affect the selectivity of detection of leakage in polluted environments and<br />

the output to low concentrations or leak.<br />

BAL_MGLD3000_20101001_en.doc 35-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Environment and selectivity setting Menu entries Warnings / functions<br />

Select reference leak<br />

and start calibration<br />

procedure<br />

Check and calibration<br />

results<br />

Environment setting<br />

Device ready to operate<br />

Probe active<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

Setting the environment<br />

This dialog is used to configure the environmental function for all leak search<br />

modes.<br />

The following selection can be made:<br />

• Exterior areas (outdoors)<br />

• Hall (large air flow, e.g. from open doors)<br />

• Room (small air flow, e.g. from air-conditioners)<br />

• Lab (no air flow)<br />

The transfer functions "Outdoor", "Hall", and "Room" are only suitable<br />

for a quick leak search. For a quantitative determination of<br />

leakage level, only the "lab" transfer function is recommended.<br />

Setting the selectivity<br />

By touching and moving the slider, the selectivity in each direction is changed:<br />

• Minimum selectivity: Low selectivity in polluted environments<br />

• Maximum selectivity: high selectivity low-effected environments in<br />

The selectivity adjustment is only in the refrigerant-probe effec<br />

OK - Confirmation, apply the settings and return to the preceding existing leak<br />

detection mode.<br />

Cancel and reset to original value<br />

BAL_MGLD3000_20101001_en.doc 36-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

4.3.5. Settings<br />

Settings Menu entries Warnings / functions<br />

Limit value for analysis<br />

mode<br />

Settings for multipoint<br />

mode<br />

Volume setting for<br />

alarm signal<br />

Gas type and leakage<br />

unit setting<br />

Language setting,<br />

date- and time setting<br />

Device ready to operate<br />

Probe active<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

Settings - limit value for analysis mode Menu entries Warnings / functions<br />

Limit value for analysis<br />

mode<br />

Settings for multipoint<br />

mode<br />

Volume setting for<br />

alarm signal<br />

Gas type and leakage<br />

unit setting<br />

Language setting, dateand<br />

time setting<br />

Device ready to operate<br />

Probe active<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

This dialogue allows you to enter the limit value for the analysis mode as a<br />

numeric value.<br />

The currently configured limit value is indicated in the currently selected unit<br />

(concentration or leak rate).<br />

To change the value, enter the change using the numeric keypad.<br />

Entry, e.g. 10.5<br />

Delete the previous character<br />

Confirm, accept entry, and return to analysis mode<br />

After entering or changing the limit value, the diagram is rescaled in<br />

the diagram.<br />

Cancel and reset to original value<br />

BAL_MGLD3000_20101001_en.doc 37-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Settings - Settings for multipoint mode Menu entries Warnings / functions<br />

Limit value for analysis<br />

mode<br />

Settings for multipoint<br />

mode<br />

Volume setting for<br />

alarm signal<br />

Gas type and leakage<br />

unit setting<br />

Language setting, dateand<br />

time setting<br />

Device ready to operate<br />

Probe active<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

Number of test points<br />

This dialogue is used to enter the desired number of test points in a test sequence<br />

as a numeric value. Any value between 1 and 30 can be chosen<br />

To enter or change the value, enter the change using the numeric keypad.<br />

To open the dialog, press the display field for the number of the test points.<br />

Entry, e.g. 5<br />

Delete the previous character<br />

Confirm, accept entry, and return to the previous menu.<br />

Cancel, reset to the original value, and return to the previous menu<br />

Limit value of the test point<br />

This dialog is used to enter the upper limit value as a numeric value (concentration<br />

or leak rate according to settings) for each test point as a numeric<br />

value.<br />

see menu Settings - Set gas type and leak unit<br />

To enter or change the value, enter the change using the numeric keypad.<br />

To open the dialogue, touch the display field for the limit value for each test<br />

point (point limit).<br />

Entry, e.g. 5<br />

Delete the previous character<br />

Confirm, accept entry, and return to the previous menu.<br />

Cancel, reset to the original value, and return to the previous menu<br />

BAL_MGLD3000_20101001_en.doc 38-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Limit value for the total of all test points<br />

This dialogue is used to enter the limit value as a numeric value (concentration<br />

or leak rate according to settings) for the sum of all test points.<br />

see menu Settings - Set gas type and leak unit<br />

To enter or change the value, enter the change using the numeric keypad.<br />

To open the dialogue, touch the display field for the limit value for all test<br />

points (cumulative limit).<br />

Entry, e.g. 15<br />

Delete the previous character<br />

Confirm, accept entry, and return to the previous menu.<br />

Cancel, reset to the original value, and return to the previous menu<br />

Test criteria for multipoint tests<br />

Alternately, the following requirements can be set:<br />

• Adherence to the configured limit value for each test point<br />

(Fail on Point)<br />

Activation/deactivation:<br />

Touch the "Failed" button to the right of the display field for the limit value<br />

for each test point (Point Limit) - depending on the status, the display will<br />

change from deactivated (dark blue background) to activated (light blue<br />

background)<br />

• Adherence to the configured cumulative limit value for all test<br />

points (Fail on Total)<br />

Activation/deactivation:<br />

Touch the "Failed" button to the right of the display field for the limit value<br />

for all test points (Total Limit) - depending on the status, the display will<br />

change from deactivated (dark blue background) to activated (light blue<br />

background)<br />

• Adherence to both limit values<br />

Both buttons activated.<br />

• No test criteria identified<br />

Both buttons deactivated.<br />

Test designation (ID) setting in multipoint mode<br />

In this dialogue you can identify how the test designation (ID) should be made<br />

in the multipoint mode:<br />

• AUTOID: Automatic assignment of a time stamp as a test ID (format<br />

YYYYMMDDHHMMSS)<br />

BAL_MGLD3000_20101001_en.doc 39-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

• SCAN ID: The test ID is queried before the start of the test and is obtained<br />

via barcode scanner<br />

The test ID is entered as the name of the log file. With SCAN ID a time stamp<br />

is also appended to a scanned or entered ID which enables logging of repeat<br />

messages without overwriting the previous log file(s).<br />

The scanner must be configured so that the scanned ID ends with a<br />

.<br />

Confirm, accept the values or settings, and return to the previous leak search<br />

mode<br />

After entering or changing the limit value, the diagram is rescaled in<br />

multipoint mode.<br />

Cancel, reset to original values and return to previous leak search mode<br />

Settings - Volume setting for alarm signal Menu entries Warnings / functions<br />

Limit value for analysis<br />

mode<br />

Settings for multipoint<br />

mode<br />

Volume setting for<br />

alarm signal<br />

Gas type and leakage<br />

unit setting<br />

Language setting, dateand<br />

time setting<br />

Device ready to operate<br />

Probe active<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

The volume of all acoustic signals of the base unit can be set in 15 levels<br />

using a sliding controller (zero volume: sound off).<br />

Setting the volume<br />

Move the volume indicator by pressing the desired volume position. The base<br />

unit will emit control sounds for the volume configured.<br />

Setting signal in probe<br />

In addition to emitting an alarm signal from the base unit, a signal can also be<br />

emitted in the hand probe. The signal is activated or deactivated using the<br />

blue button.<br />

Confirm and return to previous leak search mode<br />

Cancel and reset to original value<br />

BAL_MGLD3000_20101001_en.doc 40-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Settings - Gas type and leakage unit setting Menu entries Warnings / functions<br />

Limit value for analysis<br />

mode<br />

Settings for multipoint<br />

mode<br />

Volume setting for<br />

alarm signal<br />

Gas type and leakage<br />

unit setting<br />

Language setting, dateand<br />

time setting<br />

Device ready to operate<br />

Probe active<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

This dialog allows you to preset the gas type and the unit for the indicated<br />

values (concentration or leak rate).<br />

The current settings are always highlighted in light blue.<br />

Gas types<br />

• H2: 100% hydrogen<br />

• 5% H2: 5% hydrogen in nitrogen<br />

• ~R134a: equivalence leak rate for R134a while using trace gas<br />

• ~R600a: equivalence leak rate for R600a while using trace gas<br />

The leak rate calculated here is only a guideline. A precise<br />

calculation is not possible because for a complete equivalence<br />

analysis between trace gas and R134a/R600a. Leak-specific<br />

parameters must be considered which vary for the leaks being<br />

analyzed and in general are unknown in the field.<br />

• R22, R134a, R290, R404A, R407C, R410A, R600a: Refrigerant<br />

• C02: carbon dioxide<br />

The preset gas type is indicated in the outputs of leak search modes.<br />

Setting the gas type<br />

List selection (press a list entry, scroll by pressing and simultaneously moving<br />

the finger up and down or by pressing the scroll bar, the current setting is highlighted<br />

in light blue)<br />

Unit for concentration or leak rate<br />

Concentrations are generally determined by the device (ppm) and are<br />

then converted to leak rates. The critical unit is thus the concentration<br />

data in ppm.<br />

The other units, g/y (gram/year), oz/y (ounces/year), atm ccm/s (atmospheres<br />

cubic centimeters/sec) or mbar l/s are calculated quantities<br />

and are dependent on a constant flow through the hand probe<br />

BAL_MGLD3000_20101001_en.doc 41-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Setting the unit for concentration or leak rate<br />

List selection (press a list entry, scroll by pressing and simultaneously moving<br />

the finger up and down or by pressing the scroll bar, the current setting is highlighted<br />

in light blue)<br />

Confirm, accept entry, and return to previous leak search mode<br />

Cancel and reset to original value<br />

Settings - Language setting, date- and time<br />

setting<br />

Menu entries<br />

Limit value for analysis<br />

mode<br />

Settings for multipoint<br />

mode<br />

Volume setting for<br />

alarm signal<br />

Gas type and leakage<br />

unit setting<br />

Language setting, dateand<br />

time setting<br />

Warnings / functions<br />

Device ready to operate<br />

Probe active<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

This menu is used to enter settings for the language for navigation, as well as<br />

the date and time.<br />

Date and time refer to system date- and time. These settings are used for<br />

such purposes as logging, for example.<br />

Language<br />

Once the language is selected in the list, the user navigation will then be provided<br />

in that language.<br />

Currently the following languages are available: German, English (default<br />

setting), French, Italian, Portuguese, and Spanish.<br />

List selection (press a list entry, scroll by pressing and simultaneously moving<br />

the finger up and down or by pressing the scroll bar, the current setting is<br />

highlighted in light blue), e.g.<br />

English<br />

BAL_MGLD3000_20101001_en.doc 42-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Time<br />

Touch the time indicator in the entry dialogue<br />

Delete the time displayed<br />

Enter the current time from left to right in the format HHMMSS, e.g. for<br />

15:29:00 152900<br />

Confirm and return to previous dialogue<br />

Cancel and reset to original value<br />

Date<br />

Touch the date indicator in the entry dialogue<br />

Delete the date displayed<br />

Enter the current date from left to right in the format DDMMYY, e.g. for<br />

14.06.2010 140610<br />

Confirm and return to previous dialogue<br />

Cancel and reset to original value<br />

Accept and save settings<br />

Confirm and return to previous dialog<br />

Cancel and reset to the original settings and values<br />

4.3.6. Printing out and saving the log<br />

Print out log and save to external data medium Menu entries Warnings / functions<br />

Select drive for storage<br />

of log data in<br />

multipoint mode<br />

Print settings for external<br />

label printer<br />

Device ready to operate<br />

Probe active<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

BAL_MGLD3000_20101001_en.doc 43-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

In multipoint mode the results of the completed leak searches are logged.<br />

The menu can be used to select the drive in which this data will be stored.<br />

The following options are provided:<br />

• Drive C: internal hard drive<br />

• Drive D, etc.: external storage media connected via USB (e.g., USB stick<br />

or external drive)<br />

In addition to the drive letters, the free space and total capacity of the storage<br />

medium is displayed in GB.<br />

Selection of drive for storage<br />

List selection (press a list entry, scroll by pressing and simultaneously moving<br />

the finger up and down or by pressing the scroll bar, the current setting is<br />

highlighted in light blue), e.g.<br />

Confirm and return to previous dialogue<br />

C<br />

Cancel and reset to the original settings and values<br />

Downloading data to an external storage medium<br />

All log files stored on the internal device drive can later be downloaded to an<br />

external storage medium and deleted from the internal drive.<br />

Select the drive from the list and press the "Download" button. The progress<br />

bar shows the download progress and the number of downloaded files.<br />

BAL_MGLD3000_20101001_en.doc 44-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Sample log:<br />

Makeup of the name of the log file:<br />

_.txt<br />

e.g.: 4342610640217664_20100322151954.txt<br />

1: ID: 4342610640217664<br />

2: Start: 22.03.2010 15:18:17 End: 22.03.2010 15:19:54<br />

3: <strong>MGLD.3000</strong>: 1003001/Custom_H2_01/>100200023.4101.13-DGS1610030913<br />

4: TEST: 5%H2 5.00g/y/15.00g/y<br />

5: P01/05 1.07 g/y 1<br />

6: P02/05 0.69 g/y 1<br />

7: P03/05 0.58 g/y 1<br />

8: P04/05 0.15 g/y 1<br />

9: P05/05 0.57 g/y 1<br />

10: Total 3.07 g/y 1<br />

11: Result Passed<br />

Line 1: Test ID<br />

Line 2: Test start and test end<br />

Line 3: MGLD serial number / test equipment number of customer / hand<br />

probe serial number<br />

Line 4: Test description of test gas, point limit value, total limit value<br />

Lines 5-9: Test results for each test point: Test point number, Concentration/Leak<br />

rate, Unit, Passed / Not passed<br />

Line 10: Total concentration/-leak rate, Unit, Passed / Not passed<br />

Line 11: Total test result<br />

To prevent possible loss of data, make sure system is switched off<br />

before removing removable data media!<br />

Print out log and save to external data medium<br />

Menu entries<br />

Select drive for storage<br />

of log data in multipoint<br />

mode<br />

Print settings for external<br />

label printer<br />

Warnings / functions<br />

Device ready to operate<br />

Probe active<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

In multipoint mode the test results can also be printed out on a label.<br />

The print function is optional and currently is only available for select label<br />

printers. For more information, please contact customer support.<br />

BAL_MGLD3000_20101001_en.doc 45-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Activating/deactivating print function<br />

Touch the printer button and the display will change from deactivated (dark<br />

blue background) to activated (light blue background) depending on the status<br />

of the print function<br />

Confirm and return to previous dialog<br />

Cancel and reset to original value<br />

BAL_MGLD3000_20101001_en.doc 46-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

4.3.7. Service<br />

Service Menus Warnings / functions<br />

Query device information<br />

Set up password<br />

Device ready to operate<br />

Probe active<br />

Contact information<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

Service - Query device status Menus Warnings / functions<br />

Query device information<br />

Set up password<br />

Device ready to operate<br />

Probe active<br />

Contact information<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

BAL_MGLD3000_20101001_en.doc 47-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

By selecting the "Query device status" menu item, the device information of<br />

the base unit will be indicated<br />

MGLD BASE UNIT<br />

S/N: … Manufacturer serial number<br />

HW:… Hardware version number<br />

SW:… Software version number<br />

and the currently connected probe<br />

PROBE<br />

S/N: … Serial number<br />

Custom ID: … e.g. test equipment number (can only be set up by customer<br />

support)<br />

First use: … Date of first use of the probe,<br />

Last use: … Date of last use of the probe<br />

Operation Hours:<br />

- Total: … Total hours of operation,<br />

- Last Calibration: … Hours of operation since last calibration of<br />

probe<br />

Return to previous leak search mode<br />

Return to previous leak search mode<br />

Service - Setting up password (user) Menus Warnings / functions<br />

Query device information<br />

Set up password<br />

Device ready to operate<br />

Probe active<br />

Contact information<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

The <strong>MGLD.3000</strong> comes preset with superuser password protection.<br />

This password and the related configuration options are reserved for customer<br />

support or a customer administrator.<br />

This way it is possible to set up a user password to protect the device settings<br />

and/or limit the number of users, for example.<br />

The superuser password is entered using the numeric keypad.<br />

Entry, e.g. 123<br />

For security reasons, the password is shown as a ****** character series.<br />

BAL_MGLD3000_20101001_en.doc 48-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Delete the previous character<br />

Confirm<br />

Cancel and return to previous leak search mode<br />

Once the superuser password has been entered correctly, the following dialog<br />

is displayed for setup and activation of a user password.<br />

Service - Setting up password (user) Menus Warnings / functions<br />

Query device information<br />

Set up password<br />

Device ready to operate<br />

Probe active<br />

Contact information<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

The user password can contain a sequence of digits 0… 9 of any length you<br />

like.<br />

Touch the password field display in the entry dialog (numeric keypad).<br />

Delete the time displayed password<br />

Enter the new password, e.g. 123<br />

Confirm and return to previous password dialog<br />

Cancel, reset to original values and return to previous password dialog<br />

Activating/deactivating password protection<br />

Press the button for password protection - the display will change from deactivated<br />

(dark blue background) to activated (light blue background) depending<br />

on the status<br />

Confirm, accept value/settings, and return to previous leak search mode<br />

Cancel, reset to original settings and return to previous leak search mode<br />

BAL_MGLD3000_20101001_en.doc 49-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Service - Contact information Menus Warnings / functions<br />

Query device information<br />

Set up password<br />

Device ready to operate<br />

Probe active<br />

Contact information<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

This menu displays the relevant contact information for sales and service.<br />

Return to previous leak search mode<br />

4.4. Shutting off the device<br />

The device can be shut off in any leak search mode by briefly pressing the power switch on the rear of the<br />

base unit once. The device will automatically power down in approx. 20 sec. and shut off.<br />

If the remaining battery charge level falls below 2%, the device will likewise power down and shut off<br />

automatically.<br />

The leak search mode that is active before shutdown and all settings will be saved and will be available<br />

after the unit is restarted.<br />

In the event that an automatic power down is required, the device can also be switched off manually by<br />

pressing the power switch for approx. 3 seconds.<br />

Attention<br />

While the base unit is on the docking unit and the latter is connected to the mains power, the battery will<br />

continue to charge after shutdown.<br />

BAL_MGLD3000_20101001_en.doc 50-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

4.5. Warning and error messages<br />

Warning and error messages<br />

Warnings / functions<br />

Warning symbol blinks<br />

Probe active<br />

protected with password<br />

Connected to power socket<br />

Battery charge status<br />

Yellow background<br />

Warning/error present<br />

In case of an error or a warning message, this will be indicated by a yellow<br />

background and a blinking warning symbol.<br />

To determine the type of error or warning, press the pictogram!<br />

The error code (top left), the error number (top right), and the error message<br />

will then be displayed in the following dialogue.<br />

If multiple warning- or error messages appear, these can be displayed by<br />

pressing the arrow keys.<br />

Press the "OK" button to acknowledge the messages displayed and delete<br />

them from the error memory. The error dialog will then close.<br />

When there are no warning- or error messages, the background color will be<br />

dark blue and the warning symbol will be inactive.<br />

Close the error dialogue.<br />

BAL_MGLD3000_20101001_en.doc 51-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

The following error messages can be output:<br />

Error code Error message Cause / recommendations<br />

6101 No Probe present No probe was connected to the base unit.<br />

Switch off the device. Check the connection on the electropneumatic<br />

connection line on the base unit and on the probe.<br />

Check the connection line for damage such as kinks or cracks.<br />

Switch the device back on.<br />

6201 Probe: Heating Sensor The hand probe is initialized and the sensor is heated. This warning<br />

message will disappear when the sensor has finished heating<br />

(approx. 1-2 min.). If the warning message is still displayed after<br />

2 min., restart the device.<br />

6202 Probe: Reading Calibration<br />

Data<br />

The hand probe is initialized and the calibration data is read. This<br />

message shows an error while reading the calibration data from<br />

the probe. To correct the error, restart the device.<br />

6203 Probe: Writing Calibration<br />

Data<br />

Once the test- and calibration procedure is completed, the correction<br />

factor that is calculated is saved in the probe. The message<br />

indicates that an error occurred while saving the correction<br />

factor. To ensure correct transfer of the correction factor, repeat<br />

the test- and calibration procedure.<br />

6204 Probe: Reduced Flow The current through the probe is monitored by a current sensor.<br />

6205 Probe: No Flow A current that is too low, or a complete absence of current, in the<br />

hand probe can have the following causes:<br />

• Contamination/clogging of the Teflon tip<br />

• Damage (cracks, kinks in the connection line)<br />

• Insufficient connection of the probe to the connection line<br />

After correcting the root problem, the current is re-established. It<br />

can take up to 30 sec. to determine the current.<br />

In case of error the measurement characteristics of the device<br />

can be reduced. A determination of the gas concentration is still<br />

possible, although the conversion to leak rates is incorrect.<br />

BAL_MGLD3000_20101001_en.doc 52-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

6212, 6213, Probe: Sensor<br />

6216, 6217, Probe: Calibration Data<br />

6218, 6221, Probe: Identification<br />

6222, 6223,<br />

6224<br />

6300 Base: No Communication<br />

6301 Base: Reduced Flow<br />

6302 Base: No Flow<br />

The device does not turn on.<br />

These error numbers indicate a defect in the gas sensor system<br />

or a storage error in the hand probe. If the error occurs again<br />

after restarting the device, contact technical support.<br />

There is a problem with the internal device communication between<br />

the individual components. If the error occurs again after<br />

restarting the device, contact technical support.<br />

The current in the base unit is monitored by a current sensor. A<br />

current that is too low, or a complete absence of current, in the<br />

main device can have the following causes:<br />

• Contamination/clogging of the Teflon tip in the hand probe<br />

• Damage (kinks in the connection line)<br />

• Closed inline filter in the main device<br />

After correcting the root problem, the current is re-established.<br />

The determination of the current can take up to 30 sec.<br />

In case of error, the measurement characteristics of the device<br />

can be reduced. A determination of the gas concentration is still<br />

possible, although the conversion to leak rates is incorrect.<br />

The battery is dead and the device is not connected to the power<br />

supply.<br />

Check for a proper power connection. Check for proper positioning<br />

on the docking unit.<br />

BAL_MGLD3000_20101001_en.doc 53-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

5. Care and service<br />

Please follow the general safety instructions in 2.2 of these operating instructions.<br />

Interventions in the device, service or repair works can only be performed by the manufacturer's customer<br />

service department or by authorized professional technicians.<br />

5.1. Cleaning and care<br />

Remove the power plugs from the socket before cleaning.<br />

External cleaning of the <strong>MGLD.3000</strong> and its components may only be performed using a soft,<br />

lightly moistened, non-dripping cloth. Chemical cleaning agents, abrasives, or solvents may not be used.<br />

The contacts between the docking unit and the base unit, between the device and connection line,<br />

probe and connection line, and the interfaces with the rear of the device may only be cleaned with a dry,<br />

dust-free brush.<br />

During longer periods between use, the device and hand probes must be wrapped in their original packaging<br />

materials and placed in storage. Especially with probes, the storage area must be absolutely free of<br />

any impurities and contaminants, and must be specially protected from the influence of chemical substances,<br />

e.g. silicone.<br />

5.2. Service<br />

Before startup, regularly check the electro-pneumatic connection line for kinks, crushing or compression<br />

damage, cracks and other damage and/or impurities. Replace damaged connection lines immediately to<br />

ensure proper leak search results and to protect the device from contamination.<br />

Also regularly check the Teflon tip and the ventilation inlets/outlets on the rear of the device before startup<br />

for dirt buildup and contamination and remove these.<br />

Generally a yearly check of the hand probes by the manufacturer's customer service or by manufacturerauthorized<br />

technicians / an authorized service center is recommended.<br />

We also recommend yearly replacement of all filters by the manufacturer customer service department or<br />

by manufacturer-authorized specialists / an authorized service center.<br />

BAL_MGLD3000_20101001_en.doc 54-54


Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

5.3. Warranty Limitations<br />

In the following parts of the claim due to technological features or normal wear and tear is restricted:<br />

Battery<br />

6 months<br />

Membrane pump 8000h membrane, 20000h motor 1)<br />

Gas sensor (Ref./H2-probe) 24 months 2)<br />

Photoacoust. chamber (CO2-probe) 12 months 3)<br />

Plugs<br />

500 connections<br />

(Basis unit / Docking station)<br />

Connetctors (hose)<br />

1000 connections<br />

Touch panel 24 months 4)<br />

1)<br />

2)<br />

3)<br />

4)<br />

approximate values of the manufacturer to pollution-free intake<br />

) under normal use. Silicon-containing and sulfur-containing media and not desorbing inorganic<br />

contaminants, for example, tobacco smoke, oils, fats and liquids can be evaporated at the gas sensor<br />

lead to toxicity, which may lead to a change of Sensitivity and / or selectivity of the sensor lead.<br />

Possible consequences are a reduced sensitivity, distorted concentration or ad a background<br />

concentration. For the early detection of changes in sensor behavior for an annual review of the<br />

handheld probe by an authorized service center is recommended. The service followed the<br />

manufacturer must be adhered to-ten.<br />

in normal use. Vibrations that the highly sensitive optical system damage gene can be avoided.<br />

Contaminated intake air can cause hardware to perform optical components that affect the function of<br />

the probe. The service followed the manufacturer must be observed.<br />

If correct operation. Operate only with clean hands, soft gloves or suitable for touch panel pins. can<br />

mechanical damage (scratches) or contamination from the touch panel damage and lead to<br />

malfunction or to a restricted operation of the entire unit. The service followed the manufacturer must<br />

be observed.<br />

BAL_MGLD3000_20101001_en.doc 55-54

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!