28.10.2014 Views

Bioquell Q10 - CapellaScience

Bioquell Q10 - CapellaScience

Bioquell Q10 - CapellaScience

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Bioquell</strong> | Q-10<br />

• Hydrogen peroxide vapour (HPV) bio-decontamination<br />

technology capable of 6-log sporicidal inactivation<br />

• Easy to operate, fast bio-decontamination cycles<br />

using unique parametric control system<br />

• Mobile system capable of bio-decontaminating<br />

area volumes up to 250m 3<br />

• Unique combination catalytic aerator and<br />

distribution system<br />

• Excellent material compatibility including sensitive<br />

electronics


<strong>Bioquell</strong> Q-10<br />

The <strong>Bioquell</strong> Q-10 is a hydrogen peroxide vapour (HPV)<br />

bio-decontamination system which can be used to<br />

bio-decontaminate rooms and laboratories in a non-regulated<br />

environment. The <strong>Bioquell</strong> Q-10 has unique parametric control<br />

for ease of operation and fast bio-decontamination cycles.<br />

The associated <strong>Bioquell</strong> R-10 aeration and distribution system assists<br />

vapour distribution during bio-decontamination and reduces cycle<br />

times using a high efficiency catalytic aeration process.<br />

Features in brief<br />

• Intuitive software with PLC controls<br />

• Integrated 4.5 ltr H 2<br />

O 2<br />

reservoir for continuous room gassing<br />

• Siemens ® control systems with colour touch screen<br />

• Compact and highly portable design<br />

• Easily cleanable external surface<br />

• Volt-free contacts allowing remote start/stop functionality<br />

• Integral thermal paper printer<br />

• Lightweight detachable control panel for outside room operation<br />

• Cycle edit and store facility for new applications<br />

• Effective against a wide range of micro-organisms<br />

• Hand held hydrogen peroxide gas sensor supplied as standard<br />

• Utilises 30-35% w/w <strong>Bioquell</strong> hydrogen peroxide solution<br />

<strong>Bioquell</strong> hydrogen peroxide solution<br />

• <strong>Bioquell</strong> hydrogen peroxide is specified<br />

and approved for use with <strong>Bioquell</strong><br />

HPV generators<br />

• Approved for use as a biocide by regulatory<br />

bodies (BPD, EPA)<br />

• Selected for optimal biocidal effectiveness<br />

086/10


<strong>Bioquell</strong> Q-10 HPV generator<br />

Full 3-D surface sterilisation: Computational<br />

fluid dynamics used to fine-tune nozzle design<br />

and optimise vapour distribution.<br />

Safety: HPV monitoring device<br />

provided as standard, housed<br />

neatly on the control panel.<br />

Repeatable cycles:<br />

Intelligent software<br />

monitors key<br />

environmental factors<br />

and optimises each<br />

sterilisation cycle.<br />

Simple to operate: Intuitive<br />

icon-driven menu with colour<br />

touch screen.<br />

Self sterilisation: Integral fans on<br />

both units ensure all internal parts<br />

are sterilised, negating the risk of<br />

cross-contamination.<br />

Ergonomic design:<br />

Detachable control panel<br />

is lightweight and free<br />

standing.<br />

<strong>Bioquell</strong> R-10<br />

aerator<br />

Multiple and complex room<br />

sterilisation: Additional<br />

distribution ensures surface<br />

sterilisation of topographically<br />

complex rooms.<br />

User friendly: Unit filling<br />

mechanism is concealed<br />

behind the detachable control<br />

panel. Bottles of <strong>Bioquell</strong><br />

hydrogen peroxide ensure safe<br />

handling of sterilant.<br />

Rapid sterilisation cycles:<br />

Fast aeration via the R-10,<br />

which catalytically converts<br />

hydrogen peroxide vapour to<br />

safe and residue-free water<br />

vapour and oxygen.


<strong>Bioquell</strong> Q-10<br />

External materials of construction<br />

Temperature Operating 15°C to 30°C<br />

Storage 0°C to 35°C<br />

Relative humidity Operating 5-95% RH<br />

Storage<br />

95% RH (non-condensing)<br />

Power requirement<br />

Power supply<br />

Power consumption 230V 2760W<br />

120V<br />

1680W<br />

100V<br />

1500W<br />

Operating position<br />

<strong>Bioquell</strong> R-10<br />

Polypropylene (PP), acrylonitrile butadiene styrene (ABS), acrylonitrile<br />

butadiene styrene / polycarbonate mix (ABS/PC), aluminium (epoxy<br />

coated), stainless steel<br />

AC 230V, single phase, 50/60Hz, 12 amps<br />

AC 120V, single phase, 50/60Hz, 14 amps<br />

AC 100V, single phase, 50/60Hz, 15 amps<br />

Installation category II<br />

Upright only<br />

Dimensions (WxHxD) With Control Panel 565 x 1250 x 665mm<br />

Without Control Panel 565 x 1250 x 620mm<br />

Weight<br />

Maximum H 2<br />

O 2<br />

liquid volume<br />

H 2<br />

O 2<br />

liquid concentration<br />

Control system<br />

65 kg, including control panel (empty H 2<br />

0 2<br />

tank)<br />

4.5 litres<br />

30-35% w/w. <strong>Bioquell</strong> supply only<br />

Siemens ® S7 200PLC<br />

Airflow rate Vaporisation Injection flow 45 m 3 /hr<br />

Distribution<br />

≥800 m 3 /hr<br />

Temperature of supply vapour at outlet<br />

50-70°C<br />

H 2<br />

O 2<br />

liquid injection rate range 230V Approximately 4 - 20 g/min<br />

100V/120V<br />

Approximately 4 - 10 g/min<br />

Enclosure decontamination capacity<br />

Data printer<br />

Human interface<br />

Power cable length<br />

Control panel cable length 7.6m<br />

Locking brake<br />

External connections<br />

External materials of construction<br />

Nominally 250m 3 for rooms (subject to configuration and loading)<br />

Thermal print record<br />

Colour touch screen<br />

5m<br />

(2x) Castor lock<br />

Temperature Operating 15°C to 30°C<br />

Storage 0°C to 35°C<br />

Power requirement<br />

Power supply<br />

Power consumption 230V @ 50Hz 690W<br />

120V @ 50Hz 550W<br />

100V @ 50Hz 690W<br />

Operating position<br />

• Mains power supply • Cycle status signal digital (volt-free) output<br />

(requires voltage from external equipment) • Start/stop (run) signal<br />

24VDC input (requires volt-free connection at external equipment)<br />

• (2x) Aeration run 24VDC to external aeration units RS485 15-way<br />

sub D connector for communications (on control panel)<br />

Polypropylene (PP), acrylonitrile butadiene styrene / polycarbonate<br />

mix (ABS/PC), aluminium alloy (epoxy coated), stainless steel<br />

AC 230V, single phase, 50/60 Hz, 3.0 amps<br />

AC 120V, single phase, 50/60 Hz, 4.6 amps<br />

AC 100V, single phase, 50/60 Hz, 6.9 amps<br />

Installation category II<br />

Upright only<br />

Dimensions (WxHxD) Unit only 615 x 1200 x 580mm<br />

Weight<br />

Aeration method<br />

40 kg<br />

Airflow rate Distribution ≥800 m 3 /hr<br />

Aeration<br />

≈3000 m 3 /hr<br />

Power cable length<br />

Control cable length<br />

Locking brake<br />

External connections<br />

Patented catalyst technology<br />

5m<br />

5m<br />

(2x) Castor lock<br />

• Mains power supply • Start/stop (run) signal 24VDC input (requires<br />

volt-free connection at external equipment) • 24VDC to external aeration<br />

units (serial connection of units)<br />

086/10


Hydrogen peroxide vapour (HPV) technology<br />

HPV concentration<br />

1<br />

2<br />

3<br />

Inactivation of micro-organisms<br />

t=0*<br />

GASSING DWELL AERATION<br />

TIME<br />

* Conditioning phase not shown (vaporiser reaches temperature)<br />

Note: “features” on graph change e.g. depending on size of<br />

chamber (& related temperature effect)<br />

4<br />

5<br />

Schematic of <strong>Bioquell</strong> HPV concentration/time graph<br />

1. HPV introduced – initial rapid increase in HPV concentration<br />

2. HPV saturation/“dew point” achieved – onset of rapid bio-decontamination<br />

3. HPV gassing plateau – sustained micro-condensation effecting bio-deactivation<br />

4. Dwell period (contact time)<br />

5. Aeration – removal of HPV from the atmosphere typically by catalytic<br />

conversion to water vapour and oxygen, leaving no residues<br />

Schematic of dew point and <strong>Bioquell</strong> kill dynamics<br />

1. Injection of HPV into the enclosure starts (t=0)<br />

2. Only slight decline in biological indicator population prior to dew point<br />

3. Rapid micro-biological kill occurs immediately after saturation/“dew point” is achieved –<br />

process optimised to achieve reliable 6-log sporicidal reduction on all exposed surfaces<br />

4. Onset of micro-condensation correlates with rapid bio-deactivation<br />

5. Bio-deactivation achieved via micro-condensation (invisible layer) of hydrogen<br />

peroxide (2-6µm)<br />

* Typical BI: Tyvek ® pouched 6-log<br />

Geobacillus stearothermophilus spores<br />

Key benefits of HPV<br />

• Excellent material compatibility<br />

including sensitive electronics<br />

• Environmentally friendly –<br />

“residue-free” (breaks down to<br />

water vapour and oxygen)<br />

• Effective against a wide range<br />

of micro-organisms<br />

• High level disinfection achieved<br />

(6-log sporicidal reduction)<br />

HPV applications<br />

• Facility/area bio-decontamination<br />

in non-regulated environments<br />

• Food manufacturing or processing<br />

• Laboratories<br />

• Biomedical<br />

• Biosafety/security<br />

• Healthcare facilities


Other unique bio-decontamination<br />

solutions from <strong>Bioquell</strong><br />

<strong>Bioquell</strong> RBDS rapid facility bio-decontamination service<br />

A unique bio-decontamination service from <strong>Bioquell</strong><br />

for zones, rooms and buildings:<br />

• “Infinitely scalable” – no room or building too large<br />

• Emergency, scheduled, contract or new-build<br />

• Rapid deployment by highly trained <strong>Bioquell</strong> technicians<br />

<strong>Bioquell</strong> Clarus C HPV multi-function generator<br />

Large scale bio-decontamination capabilities designed for<br />

GMP manufacturing compliance:<br />

• Room bio-decontamination up to 350m 3<br />

• Large isolators and chambers<br />

• Permanent, semi-permanent and fixed solutions available<br />

<strong>Bioquell</strong> HPV enabled material transfer solutions<br />

Bio-decontamination chambers:<br />

• Material transfer incorporating HPV bio-decontamination<br />

at ambient temperature<br />

• Rapid decontamination leading to fast material transfer<br />

• Inflatable door seals for reliable, safe operation<br />

Disclaimer: <strong>Bioquell</strong> UK Ltd or its affiliates, distributors, agents or licensees<br />

(together “<strong>Bioquell</strong>”) reserves the right to change the specification of the<br />

<strong>Bioquell</strong> Q-10 and R-10 without notice. In addition, <strong>Bioquell</strong> makes no<br />

claims as to the guarantee of sterilisation using these products; however,<br />

<strong>Bioquell</strong> recommends that customers ensure that the requisite level of<br />

bio-decontamination is achieved using standard biological indicators such as 6-log<br />

Geobacillus stearothermophilus spores; and the <strong>Bioquell</strong> Q-10 and R-10, subject to<br />

appropriate cycle development, are designed to be able to provide such levels of<br />

bio-deactivation.<br />

Your local distributor:<br />

Clarus / <strong>Bioquell</strong> / RBDS are registered trade marks of <strong>Bioquell</strong> UK Ltd.<br />

© <strong>Bioquell</strong> UK Ltd (2010). All rights reserved.<br />

For further information on <strong>Bioquell</strong>’s technology:<br />

E: info@bioquell.com<br />

W: www.bioquell.com<br />

<strong>Bioquell</strong> UK Ltd<br />

T: +44 (0)1264 835 835<br />

<strong>Bioquell</strong> Ireland<br />

T: +353 (0)61 603 622<br />

<strong>Bioquell</strong> Inc<br />

T: +1 (215) 682 0225<br />

<strong>Bioquell</strong> SAS<br />

T: +33 (0)1 43 78 15 94<br />

<strong>Bioquell</strong> Asia Pacific Pte Ltd<br />

T: +65 6592 5145<br />

086/10