13 Table of Anomaly Signals<br />
In case of malfunctioning of the system, the controller switches off the<br />
regulation output and displays the type of anomaly.<br />
For example the controller will signal the breakage of any connected<br />
thermocouple by displaying<br />
notifications, see the table below.<br />
# Cause What to do<br />
E-01 Error in E ² PROM cell<br />
programming<br />
E-02 Cold junction sensor fault or room<br />
temperature outside of allowed<br />
limits.<br />
E-04 Incorrect configuration data.<br />
Possible loss of calibration values.<br />
E-05 Thermocouple open or<br />
temperature outside of limits.<br />
(flashing) on display. For other<br />
Call Assistance<br />
Call Assistance<br />
Check if the configuration parameters<br />
are correct.<br />
Check the connection with the<br />
sensors and their integrity.<br />
48