Maintenance, cleaning, disposal<br />
15 Calibrate the instrument (see 4.4 CALIBRATION). The<br />
instrument is now ready for operation again.<br />
5.2 Cleaning<br />
5.2.1 Cleaning the measuring instrument<br />
Occasionally wipe the outside of the measuring instrument<br />
with a damp, lint-free cloth. Disinfect the housing with isopropanol<br />
as required.<br />
Warning<br />
Avoid contact with acetone or similar detergents that contain<br />
solvents as these can damage the housing.<br />
Remove any splashes immediately.<br />
5.2.2 Cleaning the cells<br />
Cells used for measuring turbidity must always be clean.<br />
Therefore, clean them regularly:<br />
1 Clean cells inside and out with hydrochloric acid or<br />
laboratory soap<br />
2 Rinse out several times with distilled water<br />
3 Let them dry in the air<br />
4 Only hold cells by the top or by the light protective<br />
cap so that the optical path is not impaired<br />
Note<br />
Scratches in the glass change the optical characteristics of<br />
the cell and falsify the measured value. For this reason, never<br />
use scratched cells!<br />
40