What to do if ...<br />

“LoBat” display<br />

flashes<br />

Cause<br />

– The backup batteries<br />

are almost empty<br />

Remedy<br />

– Change the batteries (see<br />

5.1.2 CHANGING THE<br />

BATTERIES)<br />

“Cal” display<br />

flashes<br />

Cause<br />

– The calibration interval<br />

set up has expired<br />

Remedy<br />

– Calibrate the instrument<br />

(see 4.3 CALIBRATION)<br />

Measured values<br />

that are obviously<br />

too high<br />

Cause<br />

Remedy<br />

– Cell contaminated – Clean the cell<br />

– Cell scratched – Replace the cell<br />

– Cell misted up – Regulate the temperature<br />

of the sample before<br />

measuring<br />

– Air bubbles in the cell – Remove any air bubbles<br />

– Measuring instrument<br />

needs to be calibrated<br />

– Calibrate the measuring<br />

instrument<br />

The sample does<br />

not run out of the<br />

flow-through<br />

vessel<br />

Cause<br />

– The end of the outlet<br />

hose is placed too high<br />

– There is a foreign body in<br />

the flow-through vessel<br />

Remedy<br />

– Place the end of the hose<br />

lower than the measuring<br />

instrument<br />

– Remove the foreign body<br />

44