Measurement technology | METPOINT ®<br />
METPOINT ® FLM<br />
Scope of supply<br />
››<br />
Data display DD109 incl.<br />
››<br />
Data logger for 1 million measured values<br />
››<br />
Sensor FS109, 185 m/s, 220 mm insertion depth<br />
››<br />
Connecting cable ca. 5 m<br />
For METPOINT ® FLM<br />
Stationary Volume flow measuring<br />
<strong>Technical</strong> Data Display DD109<br />
Dimensions (mm)<br />
Operating temperature<br />
Power supply<br />
Sensor inputs<br />
Keyboard<br />
Display<br />
Alarm output<br />
Analogue output<br />
Integrated <strong>data</strong> logger<br />
118 x 115 x 93 (width | height | depth)<br />
0 … +50°C<br />
100 .… 240 VAC / 50-60 Hz / 10 VA<br />
2 inputs for dew point and consumption probes<br />
(optional 2 analogue inputs)<br />
4 buttons<br />
Graphic display, 160 x 100 Pixel<br />
2 relays, 230 VAC, 3A, potential-free, changer<br />
Connetion of the 4...20 mA signals of the dew point<br />
and consumption sensors (max. load < 500 Ohm)<br />
up to 1,000,000 values<br />
<strong>Technical</strong> Data Flow sensor FS109<br />
Dimensions (mm)<br />
220 mm insertion depth<br />
Operating temperature<br />
Power supply<br />
Measuring principle<br />
Measuring medium<br />
Air moisture of the<br />
measuring medium<br />
Operating pressure<br />
Material sensor tube<br />
and screw joint<br />
Analogue output<br />
Pulse output<br />
62 x 415 x 74,5 (width | height | depth)<br />
–30 ... +140 °C sensor tube, –30 ... +80 °C housing<br />
12 bis 30 VDC, Versorgung über das optionale<br />
Steckernetzteil oder das DD 109<br />
Calorimetric measurement<br />
Air, gases<br />
Max. 90 % r.h. (no water drops)<br />
Up to 50 bar [g]<br />
Stainless steel 1,4301<br />
4 ... 20 mA (load < 500 Ohm), accuracy 0.06 mA<br />
1 pulse per m3, active signal max. current l = 10 mA<br />
Flow Sensor FS109 variants<br />
For METPOINT ® FLM Insertion depth 120 mm * Insertion depth 160 mm * Insertion depth 220 mm * Insertion depth 300 mm * Insertion depth 400 mm *<br />
* Connecting cable not included in the scope of delivery<br />
131
Measurement technology | METPOINT ®<br />
METPOINT ® FLM compact<br />
<strong>Technical</strong> Data Flow sensor FS 109<br />
Sensor<br />
Pt45, Pt1000<br />
Operating temperature 0 … +50 °C<br />
Air moisture of the<br />
measuring medium<br />
Operating pressure<br />
Measuring unit/<br />
measuring section<br />
Power supply<br />
Power input<br />
Analogue output<br />
Pulse output<br />
Accuracy<br />
Max. 90% r.h. (no water drops)<br />
Up to 16 bar [g]<br />
Material: stainless steel 1,4301<br />
12 to 30 VDC, supply via the optional AC adapter<br />
or the DD 109<br />
Max. 80 mA at 24 VDC<br />
4... 20 mA (load < 500 Ohm), accuracy 0.06 mA<br />
1pulse per m3 or per l, pulse output potential-free,<br />
switching performance max. 30 VDC, 20 mA<br />
± 1,5% v.M. ± 0,05% v.E.<br />
Volume flow Sensor FS211 Variants<br />
62 74,5<br />
H1<br />
H<br />
A<br />
SW 36<br />
A<br />
R<br />
R<br />
L1<br />
L<br />
Measuring section DN 8 ¼" DN15 ¼" DN20 ¼" DN25 1" DN32 1½" DN40 1¼" DN50 2"<br />
Measuring range 0.8 ... 90 l/min 0.2 ... 90 m³/min 0.3 ... 170 m³/min 0.5 ... 290 m³/min 0.7 ... 480 m³/min 1 ... 550 m³/min 2 ... 900 m³/min<br />
PN (bar [g]) 16 16 16 16 16 16 16<br />
AD/ID (mm) 13.7/8.5 21.3/16.1 26.9/21.7 33.7/27.3 42.4/36.0 48.3/41.9 60.3/53.1<br />
L (mm) 194 300 475 475 475 475 475<br />
L1 (mm) 137 210 275 275 275 275 275<br />
H (mm) 176 176,4 179.2 182.6 186.9 189.9 195.9<br />
H1 (mm) 165.7 165.7 165.7 165.7 165.7 165.7 165.7<br />
A (mm) 15 20 20 25 25 25 30<br />
135